510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

G8 Automated HPLC Analyzer, Model HLC-723 G8

K071132510(k) TraditionalCleared Sep 16, 2008512-day reviewClass II21 CFR 864.7470Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tosoh Bioscience, Inc. · LCP Assay, Glycosylated Hemoglobin

Cleared on Sep 16, 2008 as substantially equivalent to K011434 (Tosoh Medics, Inc., 2001) after a 512-day review. Its family tree reaches back 2 generations to 1997. 4 later devices descend from it. It has been recalled once.

Decided
Sep 16, 2008
Review
512 days
Class
Class II
Product code
LCP
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K972265 (1997). 4 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2012: 1 descendant cleared2017: 2 descendants cleared2014: 1 descendant clearedK972265 · A1C 2.2 Plus - Automated Glycohemoglobin Analyzer · Tosoh Medics, Inc. · 1997K011434 · G7 Automated HPLC Analyzer · Tosoh Medics, Inc. · 2001 · recalled1995200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants K011434 · G7 Automated HPLC Analyzer · Tosoh Medics, Inc. · 2001 · recalledK972265 · A1C 2.2 Plus - Automated Glycohemoglobin Analyzer · Tosoh Medics, Inc. · 19970
  1. Root · 2 generations backK972265 A1C 2.2 Plus - Automated Glycohemoglobin AnalyzerTosoh Medics, Inc. · 1997
  2. 1 generation backK011434 G7 Automated HPLC AnalyzerTosoh Medics, Inc. · 2001 · Recalled
  3. This deviceK071132 G8 Automated HPLC Analyzer, Model HLC-723 G8Tosoh Bioscience, Inc. · 2008
  4. Descendant · cites it directly · 2 of its ownK122101 Capillarys Hemoglobin HBA1C Capillarys 2 Flex Piercing Capillarys…Sebia · 2012
  5. Descendant · 2 generations on · 1 of its ownK133344 Minicap HB A1C, Minicap Flex-Piercing, HB A1C Capillarys Controls…Sebia · 2014
  6. Descendant · cites it directly · 0 of its ownK162822 ADAMS A1c HA-8180V, CALIBRATOR 80Arkray Factory, Inc. · 2017
  7. Descendant · 2 generations on · 0 of its ownK162281 CAPI 3 Hb A1c, MULTI-SYSTEM Hb A1C CAPILLARY CONTROLS (2)Sebia · 2017

Most influential descendants

Showing 4 of 4 · See all
  • K122101Capillarys Hemoglobin HBA1C Capillarys 2 Flex Piercing Capillarys HBA1C Controls…Sebia · 2012 · cites it directly2descendants
  • K133344Minicap HB A1C, Minicap Flex-Piercing, HB A1C Capillarys Controls, HBA1C Capillarys…Sebia · 2014 · 2 generations on1descendant
  • K162822ADAMS A1c HA-8180V, CALIBRATOR 80Arkray Factory, Inc. · 2017 · cites it directly0descendants
  • K162281CAPI 3 Hb A1c, MULTI-SYSTEM Hb A1C CAPILLARY CONTROLS (2)Sebia · 2017 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K972265A1C 2.2 Plus - Automated Glycohemoglobin AnalyzerTosoh Medics, Inc.1997—
−1K011434G7 Automated HPLC AnalyzerTosoh Medics, Inc.20012
0 (this device)K071132G8 Automated HPLC Analyzer, Model HLC-723 G8Tosoh Bioscience, Inc.2008—
+1K122101Capillarys Hemoglobin HBA1C Capillarys 2 Flex Piercing Capillarys HBA1C…Sebia2012—
+2K133344Minicap HB A1C, Minicap Flex-Piercing, HB A1C Capillarys Controls, HBA1C…Sebia2014—
+1K162822ADAMS A1c HA-8180V, CALIBRATOR 80Arkray Factory, Inc.2017—
+2K162281CAPI 3 Hb A1c, MULTI-SYSTEM Hb A1C CAPILLARY CONTROLS (2)Sebia2017—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201410
201720
FIG. 1 — Lineage of K071132, 1997–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LCP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCP device

  1. #1 K221326 Nova Allegro HbA1c Assay, Nova Allegro Analyzerscore 9.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011434G7 Automated HPLC AnalyzerTosoh Medics, Inc.20012

Also cleared against K011434: 4 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122101Capillarys Hemoglobin HBA1C Capillarys 2 Flex Piercing Capillarys HBA1C Controls…Sebia20122
K162822ADAMS A1c HA-8180V, CALIBRATOR 80Arkray Factory, Inc.20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 7, 2016Class IIZ-1225-2017Software designThe device defect associated with this recall may result in HbA1c values with a negative bias in patients with the hemoglobin variants C, D, E and S. The negative bias occurs at the higher levels of HbA1c, e.g. 9%, but not at the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 2 companies; a new generation was cleared every 7.0 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K122101 (Sebia, 2012) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K011434.

How does its category compare?

All LCP devices

LCP Assay, Glycosylated Hemoglobin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · LCP

1in 2024

2007: 72008: 52009: 62010: 22011: 102012: 92013: 22014: 82015: 32016: 22017: 52018: 22019: 42020: 12023: 22024: 1
2007high 10 (2011) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCP

269in 2019

2007: 133 days median2008: 206 days median2009: 158 days median2011: 176 days median2012: 143 days median2014: 127 days median2015: 239 days median2017: 218 days median2019: 269 days median
2007high 269 (2019) · low 127 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LCP

1in 2025

2007: 12008: 32009: 62010: 52011: 512013: 42014: 22015: 22016: 62017: 12018: 12022: 12024: 12025: 1
2007high 51 (2011) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LCP

22in 2025

2007: 52008: 112009: 152010: 42011: 272012: 512013: 1932014: 1822015: 642016: 292017: 1492018: 1,9112019: 1142020: 132021: 72022: 82023: 272024: 392025: 22
2007high 1,911 (2018) · low 4 (2010)2025

MAUDE reports naming product code LCP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007713315
20085206311
20096158615
20102—54
2011101765127
20129143051
20132—4193
201481272182
20153239264
20162—629
201752181149
20182—11,911
201942690114
20201—013
20210—07
20220—18
20232—027
20241—139
20250—122
20260—0—

What is it indicated for?

The G8 Automated Glycohemoglobin Analyzer HLC-723G8 is intended for INVITRO DIAGNOSTIC USE for the measurement of hemoglobin Al c (HbAl c) in whole blood specimens. Alc measurements are used in the clinical management of diabetes to assess the long-term efficacy of diabetic control.

The AlC Calibrator Set is a reference agent designed for calibrating Tosoh Automated Glycohemoglobin Analyzer HLC-723G8.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071132: G8 Automated HPLC Analyzer, Model HLC-723 G8.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K071132

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026