510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

ST Aia Pack HBA1C

K100304510(k) TraditionalCleared Aug 27, 2010205-day reviewClass II21 CFR 864.7470Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tosoh Bioscience, Inc. · LCP Assay, Glycosylated Hemoglobin

Cleared on Aug 27, 2010 as substantially equivalent to K011434 (Tosoh Medics, Inc., 2001) after a 205-day review. Its family tree reaches back 2 generations to 1997. It has been recalled 4 times.

Decided
Aug 27, 2010
Review
205 days
Class
Class II
Product code
LCP
Recalls
4
Predicates
1

Its longest line of descent runs back 2 generations to K972265 (1997). No later device cites it as a predicate yet.

  1. Root · 2 generations backK972265 A1C 2.2 Plus - Automated Glycohemoglobin AnalyzerTosoh Medics, Inc. · 1997
  2. 1 generation backK011434 G7 Automated HPLC AnalyzerTosoh Medics, Inc. · 2001 · Recalled
  3. This deviceK100304 ST Aia Pack HBA1CTosoh Bioscience, Inc. · 2010
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K972265A1C 2.2 Plus - Automated Glycohemoglobin AnalyzerTosoh Medics, Inc.1997—
−1K011434G7 Automated HPLC AnalyzerTosoh Medics, Inc.20012
0 (this device)K100304ST Aia Pack HBA1CTosoh Bioscience, Inc.2010—
FIG. 1 — Lineage of K100304, 1997–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LCP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCP device

  1. #1 K221326 Nova Allegro HbA1c Assay, Nova Allegro Analyzerscore 9.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011434G7 Automated HPLC Analyzer
Summary page 2
…tion Name: Class 11 LCP 21 CFR 864.7470 Glycosylated Hemoglobin Assay Predicate Device: K01 1434 Tosoh Automated Glycohemoglobin Analyzer HLC-723G7…
Tosoh Medics, Inc.20012

Also cleared against K011434: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 6, 2018Class IIZ-2745-2018Software designHbAE is known to interfere with the HbA1c assay on the current version of software, Ver. 5.23. Customers should exercise caution when reviewing chromatograms and ensure that Flag 43 is enabled on their device to avoid reporting invalid…
Mar 5, 2018Class IIZ-1810-2018Nonconforming Material/ComponentThe assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results.
Mar 5, 2018Class IIZ-1812-2018Nonconforming Material/ComponentThe assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results.
Mar 5, 2018Class IIZ-1809-2018Nonconforming Material/ComponentThe assay can potentially generate erroneously elevated or erroneously decreased HbA1c patient results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 2 companies; a new generation was cleared every 8.9 years on median.
  • 1 of the 2 ancestors has a recall of any class.
  • 4 other devices also cite its predicate K011434.

How does its category compare?

All LCP devices

LCP Assay, Glycosylated Hemoglobin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · LCP

1in 2024

2007: 72008: 52009: 62010: 22011: 102012: 92013: 22014: 82015: 32016: 22017: 52018: 22019: 42020: 12023: 22024: 1
2007high 10 (2011) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCP

269in 2019

2007: 133 days median2008: 206 days median2009: 158 days median2011: 176 days median2012: 143 days median2014: 127 days median2015: 239 days median2017: 218 days median2019: 269 days median
2007high 269 (2019) · low 127 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LCP

1in 2025

2007: 12008: 32009: 62010: 52011: 512013: 42014: 22015: 22016: 62017: 12018: 12022: 12024: 12025: 1
2007high 51 (2011) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LCP

22in 2025

2007: 52008: 112009: 152010: 42011: 272012: 512013: 1932014: 1822015: 642016: 292017: 1492018: 1,9112019: 1142020: 132021: 72022: 82023: 272024: 392025: 22
2007high 1,911 (2018) · low 4 (2010)2025

MAUDE reports naming product code LCP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007713315
20085206311
20096158615
20102—54
2011101765127
20129143051
20132—4193
201481272182
20153239264
20162—629
201752181149
20182—11,911
201942690114
20201—013
20210—07
20220—18
20232—027
20241—139
20250—122
20260—0—

What is it indicated for?

ST AlA-PACK HbAlc is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement for percent concentration of Hemoglobin A1c (HbAlc) in EDTA whole blood on Tosoh AlA System Analyzer. HbAlc measurement is used in the management and treatment of diabetes. The ST AlA-PACK HbAlc CALIBRATOR SET is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of ST AlA-PACK HbAlc Assay. The AlA-PACK HbAlc CONTROL SET is intended for IN VITRO…

Read the full statement (120 words)

ST AlA-PACK HbAlc is designed for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement for percent concentration of Hemoglobin A1c (HbAlc) in EDTA whole blood on Tosoh AlA System Analyzer. HbAlc measurement is used in the management and treatment of diabetes.

The ST AlA-PACK HbAlc CALIBRATOR SET is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of ST AlA-PACK HbAlc Assay.

The AlA-PACK HbAlc CONTROL SET is intended for IN VITRO DIAGNOSTIC USE ONLY for performing quality control procedures with the ST AlA-PACK HbAlc Assay.

The ST AlA-PACK HbAlc PRETREATMENT SOLUTION is intended for IN VITRO DIAGNOSTIC USE ONLY for the pretreatment of the patient samples or AlA-PACK HbAlc CONTROL SET for the ST AlA-PACK HbA1lc assay.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100304: ST Aia Pack HBA1C.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K100304

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026