Dca Vantage, Model 5075

K071466510(k) SpecialCleared Oct 9, 2007133-day reviewClass II21 CFR 864.7470Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions Diagnostics · LCP Assay, Glycosylated Hemoglobin

Cleared on Oct 9, 2007 as substantially equivalent to K951361 (Heraeus Kulzer, Inc., 1995) and 1 other predicate after a 133-day review. Its family tree reaches back 1 generation to 1995. 8 later devices descend from it. It has been recalled once.

Decided
Oct 9, 2007
Review
133 days
Class
Class II
Product code
LCP
Recalls
1
Predicates
2

Its longest line of descent runs back 1 generation to K951361 (1995). 2 ancestors in all. 8 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor2017: 1 descendant cleared2016: 1 descendant cleared2014: 3 descendants cleared2010: 1 descendant cleared2009: 1 descendant cleared2012: 1 descendant clearedK951361 · Dca 2000 System for Hemoglobin A1C · Heraeus Kulzer, Inc. · 1995 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 5 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant K951361 · Dca 2000 System for Hemoglobin A1C · Heraeus Kulzer, Inc. · 1995 · recalled0
  1. Root · 1 generation backK951361 Dca 2000 System for Hemoglobin A1CHeraeus Kulzer, Inc. · 1995 · Recalled
  2. This deviceK071466 Dca Vantage, Model 5075Siemens Medical Solutions Diagnostics · 2007
  3. Descendant · cites it directly · 2 of its ownK082275 Clover A1C/Glycosylated Hemoglobin Monitor & Blood Glucose Monitor…Infopia Co, Ltd. · 2010
  4. Descendant · cites it directly · 1 of its ownK130014 A1C Gear SystemSakae Corporation · 2014
  5. Descendant · 2 generations on · 1 of its ownK121366 Hemocue HBA1C 501Infopia Co, Ltd. · 2012
  6. Descendant · cites it directly · 0 of its ownK163633 cobas HbA1c Test, cobas b 101 systemRoche Diagnostics Operations · 2017
  7. Descendant · cites it directly · 0 of its ownK142664 URight Hemoglobin A1c system, FORA A1c SystemTaidoc Technology Corporation · 2016

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2009: 12010: 12012: 12014: 32016: 12017: 1
2009high 3 (2014) · low 0 (2011)2026

Most influential descendants

Showing 5 of 8 · See all
  • K082275Clover A1C/Glycosylated Hemoglobin Monitor & Blood Glucose Monitor, Model Igm-0023Infopia Co, Ltd. · 2010 · cites it directly2descendants
  • K130014A1C Gear SystemSakae Corporation · 2014 · cites it directly1descendant
  • K121366Hemocue HBA1C 501Infopia Co, Ltd. · 2012 · 2 generations on1descendant
  • K163633cobas HbA1c Test, cobas b 101 systemRoche Diagnostics Operations · 2017 · cites it directly0descendants
  • K142664URight Hemoglobin A1c system, FORA A1c SystemTaidoc Technology Corporation · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K951361Dca 2000 System for Hemoglobin A1CHeraeus Kulzer, Inc.19951
0 (this device)K071466Dca Vantage, Model 5075Siemens Medical Solutions Diagnostics2007—
+1K082275Clover A1C/Glycosylated Hemoglobin Monitor & Blood Glucose Monitor, Model…Infopia Co, Ltd.2010—
+1K130014A1C Gear SystemSakae Corporation2014—
+2K121366Hemocue HBA1C 501Infopia Co, Ltd.2012—
+1K163633cobas HbA1c Test, cobas b 101 systemRoche Diagnostics Operations2017—
+1K142664URight Hemoglobin A1c system, FORA A1c SystemTaidoc Technology Corporation2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201010
201210
201430
201610
201710
FIG. 1 — Lineage of K071466, 1995–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LCP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCP device

  1. #1 K221326 Nova Allegro HbA1c Assay, Nova Allegro Analyzerscore 9.6 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K951361Dca 2000 System for Hemoglobin A1C
Summary page 1
…645) Title 21 CFR Creatinine Test System (862.1225) 510(k) Numbers * DCA 2000 System for Hemoglobin Alc (K951361) - 1.C PI a DCA 2000+ Microalbumin / Creatinine Assay (…
Heraeus Kulzer, Inc.19951
*K963142Dca 2000 + Microalbumin/Creatinine Assay
Summary page 1
…2000 System for Hemoglobin Alc (K951361) - 1.C PI a DCA 2000+ Microalbumin / Creatinine Assay (K963142)) ) (!- Manufacturer Bayer HealthCare LLC (currently Siemens Medical S…
Bayer Corp.1996—

Also cleared against K951361: 4 other devices →

What descends from it? (8)

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K082275Clover A1C/Glycosylated Hemoglobin Monitor & Blood Glucose Monitor, Model Igm-0023Infopia Co, Ltd.20102
K130014A1C Gear SystemSakae Corporation20141
K163633cobas HbA1c Test, cobas b 101 systemRoche Diagnostics Operations20170
K142664URight Hemoglobin A1c system, FORA A1c SystemTaidoc Technology Corporation20160
K091711S- Test Hemoglobin Alc (HBA1C), Model RC0023Alfa Wassermann Diagnostic Technologies, Inc.20090

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 30, 2020Class IIZ-2747-2020Labeling designIf DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K082275 (Infopia Co, Ltd., 2010) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K951361.

How does its category compare?

All LCP devices

LCP Assay, Glycosylated Hemoglobin: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · LCP

1in 2024

2007: 72008: 52009: 62010: 22011: 102012: 92013: 22014: 82015: 32016: 22017: 52018: 22019: 42020: 12023: 22024: 1
2007high 10 (2011) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCP

269in 2019

2007: 133 days median2008: 206 days median2009: 158 days median2011: 176 days median2012: 143 days median2014: 127 days median2015: 239 days median2017: 218 days median2019: 269 days median
2007high 269 (2019) · low 127 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LCP

1in 2025

2007: 12008: 32009: 62010: 52011: 512013: 42014: 22015: 22016: 62017: 12018: 12022: 12024: 12025: 1
2007high 51 (2011) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LCP

22in 2025

2007: 52008: 112009: 152010: 42011: 272012: 512013: 1932014: 1822015: 642016: 292017: 1492018: 1,9112019: 1142020: 132021: 72022: 82023: 272024: 392025: 22
2007high 1,911 (2018) · low 4 (2010)2025

MAUDE reports naming product code LCP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007713315
20085206311
20096158615
20102—54
2011101765127
20129143051
20132—4193
201481272182
20153239264
20162—629
201752181149
20182—11,911
201942690114
20201—013
20210—07
20220—18
20232—027
20241—139
20250—122
20260—0—

What is it indicated for?

The DCA VantageTM is a semi-automated, benchtop system. It is designed to quantitatively measure the percent Hemoglobin Alc in blood and low concentrations of albumin in urine (microalbuminuria), creatinine in urine, and the albumin/creatinine ratio in urine. The measurement of hemoglobin Alc concentration is recommended for monitoring the long- term glycemic control of persons with diabetes. Testing for microalbuminuria (low concentration of albumin in…

Read the full statement (162 words)

The DCA VantageTM is a semi-automated, benchtop system. It is designed to quantitatively

measure the percent Hemoglobin Alc in blood and low concentrations of albumin in urine

(microalbuminuria), creatinine in urine, and the albumin/creatinine ratio in urine.

The measurement of hemoglobin Alc concentration is recommended for monitoring the long-

term glycemic control of persons with diabetes.

Testing for microalbuminuria (low concentration of albumin in urine) is recommended in patients

with insulin-dependent diabetes mellitus (IDDM) as well as patients with non-insulin dependent

diabetes mellitus (NIDDM). It is intended for use in both screening for, and monitoring

treatment of, microalbuminuria.

Measurement of creatinine is used as a calculation basis to adjust the albumin/creatinine ratio

result for varying urine concentrations.

The reporting of albumin/creatinine ratio is recommended for the early detection of kidney

diseases.

The DCA Vantage system is for use in laboratories such as: physician office laboratories,

clinics, and hospitals.

Tests performed using the DCA Vantage TM are intended for in vitro diagnostic use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071466: Dca Vantage, Model 5075.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K071466

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026