Neonatal Bilirubin on Rapidlab Models 1245 and 1265

K073537510(k) TraditionalCleared Jun 27, 2008193-day reviewClass I21 CFR 862.1113Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions Diagnostics · MQM Bilirubin (Total And Unbound) In The Neonate Test System

Cleared on Jun 27, 2008 as substantially equivalent to K043218 (Radiometer Medical Aps, 2005) and 1 other predicate after a 193-day review. Its family tree reaches back 6 generations to 1986. 1 later device descends from it. It has been recalled 7 times.

Decided
Jun 27, 2008
Review
193 days
Class
Class I
Product code
MQM
Recalls
7
Predicates
2

Its longest line of descent runs back 6 generations to K861087 (1986). 21 ancestors in all, the oldest K790608 (1979). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1999: 1 other ancestor1998: 2 other ancestors1992: 2 other ancestors1984: 1 other ancestor1979: 1 other ancestor2000: 1 other ancestor1997: 1 other ancestor1996: 3 other ancestors1995: 1 other ancestor1987: 1 other ancestor1986: 1 other ancestor2011: 1 descendant clearedK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 1986K961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalledK041874 · ABL800 Flex · Radiometer Medical Aps · 2004 · recalledK043218 · ABL800 Flex with Flexq Module, Model ABL8XX · Radiometer Medical Aps · 2005 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 4 other ancestors3 generations back: 3 other ancestors4 generations back: 5 other ancestors5 generations back: 1 other ancestor6 generations back: 1 other ancestor 1 generation forward: 1 descendant K043218 · ABL800 Flex with Flexq Module, Model ABL8XX · Radiometer Medical Aps · 2005 · recalledK041874 · ABL800 Flex · Radiometer Medical Aps · 2004 · recalledK992859 · ABL700 Series Analyzer with Autocheck Module · Radiometer America, Inc. · 2000 · recalledK972673 · Avl 9181 Electrolyte Analyzer · Avl Scientific Corp. · 1997 · recalledK961458 · Avl 9180 Electrolyte Analyzer · Avl Scientific Corp. · 1996 · recalledK861087 · Avl Model 983 Electrolyte Analyzer · Avl Scientific Corp. · 1986−50
  1. Root · 6 generations backK861087 Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp. · 1986
  2. 5 generations backK961458 Avl 9180 Electrolyte AnalyzerAvl Scientific Corp. · 1996 · Recalled
  3. 4 generations backK972673 Avl 9181 Electrolyte AnalyzerAvl Scientific Corp. · 1997 · Recalled
  4. 3 generations backK992859 ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc. · 2000 · Recalled
  5. 2 generations backK041874 ABL800 FlexRadiometer Medical Aps · 2004 · Recalled
  6. 1 generation backK043218 ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps · 2005 · Recalled
  7. This deviceK073537 Neonatal Bilirubin on Rapidlab Models 1245 and 1265Siemens Medical Solutions Diagnostics · 2008
  8. Descendant · cites it directly · 0 of its ownK110277 Rapidpoint 405 Neonatal BilirubinSiemens Healthcare Diagnostics · 2011 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K110277Rapidpoint 405 Neonatal BilirubinSiemens Healthcare Diagnostics · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K861087Avl Model 983 Electrolyte AnalyzerAvl Scientific Corp.1986—
−5K961458Avl 9180 Electrolyte AnalyzerAvl Scientific Corp.19961
−4K972673Avl 9181 Electrolyte AnalyzerAvl Scientific Corp.19977
−3K992859ABL700 Series Analyzer with Autocheck ModuleRadiometer America, Inc.20004
−2K041874ABL800 FlexRadiometer Medical Aps200422
−1K043218ABL800 Flex with Flexq Module, Model ABL8XXRadiometer Medical Aps20053
0 (this device)K073537Neonatal Bilirubin on Rapidlab Models 1245 and 1265Siemens Medical Solutions Diagnostics2008—
+1K110277Rapidpoint 405 Neonatal BilirubinSiemens Healthcare Diagnostics20115

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198410
198610
198710
199220
199510
199630
199710
199820
199910
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
FIG. 1 — Lineage of K073537, 1986–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MQM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K043218ABL800 Flex with Flexq Module, Model ABL8XX
Summary page 1
…ABL800 FLEX ABL735 Common name Blood gas analyzer Blood gas analyzer 510(k) Number K043218 K991417 Manufacturer Radiometer Medical ApS Radiometer America Device D…
Radiometer Medical Aps20053
*K991417Abl 735 Analyzer
Summary page 1
…ABL735 Common name Blood gas analyzer Blood gas analyzer 510(k) Number K043218 K991417 Manufacturer Radiometer Medical ApS Radiometer America Device Description Neonatal Bilirubin (nBili…
Radiometer America, Inc.19992

Also cleared against K043218: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K110277Rapidpoint 405 Neonatal BilirubinSiemens Healthcare Diagnostics20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 11, 2016Class IIZ-2802-2016Software designThere is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system
Aug 11, 2016Class IIZ-2803-2016Software designThere is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned Off for the system
May 6, 2015Class IIZ-1962-2015Software designD50 and D51 diagnostic error codes are not functional.
Dec 30, 2014Class IIZ-1254-2015Software designNeonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
Dec 30, 2014Class IIZ-1255-2015Software designNeonatal bilirubin (nBili) parameter may have increased variability when the nBili concentration is greater than 12 mg/dl and the tHb concentration exceeds the upper reportable range of greater than 25 g/dl.
Sep 17, 2012Class IIZ-0107-2014Software Manufacturing/Software DeploymentRAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
Sep 17, 2012Class IIZ-0108-2014Software Manufacturing/Software DeploymentRAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 3.1 years on median.
  • 12 of the 21 ancestors have a recall of any class.
  • Of its descendants, K110277 (Siemens Healthcare Diagnostics, 2011) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K043218.

How does its category compare?

All MQM devices

MQM Bilirubin (Total And Unbound) In The Neonate Test System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · MQM

1in 2017

2008: 12011: 12014: 22017: 1
2007high 2 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MQM

Too few MQM 510(k)s in any year since 2007 to give a median.

Recalls per year · MQM

No recall names MQM since 2007.

Adverse event reports per year · MQM

2in 2025

2007: 52008: 32010: 42011: 32012: 42014: 12015: 12016: 12017: 12018: 52019: 62020: 12021: 12023: 142024: 172025: 2
2007high 17 (2024) · low 0 (2009)2025

MAUDE reports naming product code MQM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—05
20081—03
20090—00
20100—04
20111—03
20120—04
20130—00
20142—01
20150—01
20160—01
20171—01
20180—05
20190—06
20200—01
20210—01
20220—00
20230—014
20240—017
20250—02
20260—0—

What is it indicated for?

The neonatal bilirubin test intended use on the Rapidlab 1245 and Rapidlab 1265 analyzers is an in vitro diagnostic test for the determination of total neonatal bilirubin (nBili) concentration in the whole blood of newborn infants. Measurement of nBili aids in assessing the risk of kernicterus.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K073537: Neonatal Bilirubin on Rapidlab Models 1245 and 1265.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K073537

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026