510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Terumo HX2 Temperature Management System

K071521510(k) TraditionalCleared Nov 21, 2007170-day reviewClass II21 CFR 870.4250Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Terumo Cardiovascular Systems Corp. · DWC Controller, Temperature, Cardiopulmonary Bypass

Cleared on Nov 21, 2007 as substantially equivalent to K883603 (3M Health Care, Sarns, 1988) after a 170-day review. Its family tree reaches back 1 generation to 1988. It has been recalled 7 times.

Decided
Nov 21, 2007
Review
170 days
Class
Class II
Product code
DWC
Recalls
7
Predicates
1

Its longest line of descent runs back 1 generation to K883603 (1988). No later device cites it as a predicate yet.

K883603 · Sarns Temp. Control and Monitor System (TCM II) · 3M Health Care, Sarns · 1988 · recalled19851990199520002005
  1. Root · 1 generation backK883603 Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns · 1988 · Recalled
  2. This deviceK071521 Terumo HX2 Temperature Management SystemTerumo Cardiovascular Systems Corp. · 2007
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K883603Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns19884
0 (this device)K071521Terumo HX2 Temperature Management SystemTerumo Cardiovascular Systems Corp.2007—
FIG. 1 — Lineage of K071521, 1988–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWC device

  1. #1 K220635 Heater-Cooler System 3Tscore 5.3 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K883603Sarns Temp. Control and Monitor System (TCM II)
Summary page 1
…troller (21 CFR 870.4250) Legally Marketed Predicate Device: T Sarns M Dual Cooler/Heater (Preamendment) Sarns TM Temperature Control and Monitor System (TCM II)(K883603) Indications for Use: The Terumo® HX2TM Temperature Management System is indicated for use to supply temperature regulated water…
3M Health Care, Sarns19884
*—Sarns TM Dual Cooler/Heater (Preamendment)
Named without a submission number
—

Also cleared against K883603: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 16, 2021Class IIZ-1789-2021Device DesignTerumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best…
Mar 16, 2018Class IIZ-1416-2018Labeling designUpdate to cooler-heater cleaning instructions.
Jun 8, 2016Class IIZ-2207-2016Nonconforming Material/ComponentTerumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning…
Feb 6, 2015Class IIZ-1116-2015Device DesignMalfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing…
Oct 16, 2012Class IIZ-0278-2013Nonconforming Material/ComponentBased on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may…
Oct 16, 2012Class IIZ-0279-2013Nonconforming Material/ComponentBased on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may…
Jun 11, 2010Not classifiedZ-2060-2010Device DesignThe HX2 Temperature Management System mixing valve may fail resulting in an unusable water channel and then the machine would go into standby mode.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • 1 other device also cites its predicate K883603.

How does its category compare?

All DWC devices

DWC Controller, Temperature, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2007, this device’s year, is highlighted.

FDA decisions per year · DWC

2in 2022

2007: 12010: 12012: 12013: 12014: 12020: 12022: 2
2007high 2 (2022) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWC

Too few DWC 510(k)s in any year since 2007 to give a median.

Recalls per year · DWC

5in 2023

2010: 12012: 22014: 12015: 12016: 12017: 12018: 32021: 52023: 5
2007high 5 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWC

670in 2025

2007: 32009: 22010: 2432011: 2812012: 3512013: 2442014: 3782015: 5752016: 7872017: 7162018: 5642019: 6972020: 5482021: 6222022: 6052023: 9282024: 5552025: 670
2007high 928 (2023) · low 0 (2008)2025

MAUDE reports naming product code DWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20080—00
20090—02
20101—1243
20110—0281
20121—2351
20131—0244
20141—1378
20150—1575
20160—1787
20170—1716
20180—3564
20190—0697
20201—0548
20210—5622
20222—0605
20230—5928
20240—0555
20250—0670
20260—0—

What is it indicated for?

The Terumo® HX2TM Temperature Management System is indicated for use to supply temperature regulated water to heat exchangers during open heart surgery.

Operation 100 to 400 C, less than 75% relative humidity, non-condensing.

Storage Ventilated area, -300 to 54 0C (-22 to 130 0F), less than 95% relative humidity, non-condensing.

Transporting Units may be transported up or down an incline of up to 15 degrees without tipping.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K071521: Terumo HX2 Temperature Management System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K071521

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026