510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Heater Cooler Unit, Model Hcu 30

K031544510(k) TraditionalCleared Aug 7, 200383-day reviewClass II21 CFR 870.4250Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jostra AB · DWC Controller, Temperature, Cardiopulmonary Bypass

Cleared on Aug 7, 2003 as substantially equivalent to K883603 (3M Health Care, Sarns, 1988) after an 83-day review. Its family tree reaches back 1 generation to 1988. 7 later devices descend from it. It has been recalled 4 times.

Decided
Aug 7, 2003
Review
83 days
Class
Class II
Product code
DWC
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K883603 (1988). 7 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2012: 1 descendant cleared2006: 1 descendant cleared2013: 1 descendant cleared2020: 1 descendant cleared2014: 1 descendant cleared2022: 2 descendants clearedK883603 · Sarns Temp. Control and Monitor System (TCM II) · 3M Health Care, Sarns · 1988 · recalled1990200020102020
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant K883603 · Sarns Temp. Control and Monitor System (TCM II) · 3M Health Care, Sarns · 1988 · recalled0
  1. Root · 1 generation backK883603 Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns · 1988 · Recalled
  2. This deviceK031544 Heater Cooler Unit, Model Hcu 30Jostra AB · 2003
  3. Descendant · cites it directly · 4 of its ownK052601 Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH · 2006 · Recalled
  4. Descendant · 2 generations on · 1 of its ownK191402 Heater-Cooler System 3TLivanova Deutschland, GmbH · 2020 · Recalled
  5. Descendant · 3 generations on · 0 of its ownK220635 Heater-Cooler System 3TLivanova Deutschland, GmbH · 2022
  6. Descendant · 2 generations on · 0 of its ownK212657 Quantum Heater-CoolerSpectrum Medical, Ltd. · 2022
  7. Descendant · 2 generations on · 0 of its ownK140012 Sorin FlexthermSorin Group Deutschland GmbH · 2014

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2006: 12012: 12013: 12014: 12020: 12022: 2
2006high 2 (2022) · low 0 (2007)2026

Most influential descendants

Showing 5 of 7 · See all
  • K052601Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH · 2006 · cites it directly4descendants
  • K191402Heater-Cooler System 3TLivanova Deutschland, GmbH · 2020 · 2 generations on1descendant
  • K220635Heater-Cooler System 3TLivanova Deutschland, GmbH · 2022 · 3 generations on0descendants
  • K212657Quantum Heater-CoolerSpectrum Medical, Ltd. · 2022 · 2 generations on0descendants
  • K140012Sorin FlexthermSorin Group Deutschland GmbH · 2014 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K883603Sarns Temp. Control and Monitor System (TCM II)3M Health Care, Sarns19884
0 (this device)K031544Heater Cooler Unit, Model Hcu 30Jostra AB2003—
+1K052601Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH200618
+2K191402Heater-Cooler System 3TLivanova Deutschland, GmbH20201
+3K220635Heater-Cooler System 3TLivanova Deutschland, GmbH2022—
+2K212657Quantum Heater-CoolerSpectrum Medical, Ltd.2022—
+2K140012Sorin FlexthermSorin Group Deutschland GmbH2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
201210
201310
201410
202010
202220
FIG. 1 — Lineage of K031544, 1988–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWC device

  1. #1 K220635 Heater-Cooler System 3Tscore 5.3 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K883603Sarns Temp. Control and Monitor System (TCM II)
Summary page 1
…mes: Controller, Temperature, Cardiopulmonary Bypass Predicate Device: Sarns temperature control and monitor system (TCM 11) K883603 Device Description: The Jostra Heater-Cooler Unit HCU 30 supplies temperature-controlled water for cardioplegia heat exchanger…
3M Health Care, Sarns19884

Also cleared against K883603: 1 other device →

What descends from it? (7)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K052601Stockert Heater-Cooler System 3TSorin Group Deutschland GmbH20064
K130300Heater-Cooler Unit Hcu 40Maquet Cardiopulmonary, AG20130
K122813Hemotherm Model 400CE Cooler/Heater SystemCincinnati Sub-Zero Products, Inc.20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 14, 2021Class IIZ-2140-2021Device DesignThere is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
Sep 27, 2016Class IIZ-0348-2017Device DesignThere is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
Nov 13, 2015Class IIZ-0589-2016Device DesignFailure of the HEATER-COOLER UNIT (HCU 30) Type 2 power supply board. A drop of more than 10% voltage can cause overheating and subsequent failing of the R56 thermistor of the power supply board. Board failure typically occurs when the…
Nov 10, 2006Not classifiedZ-0350-2007OtherThe HCU 30 system potentially will not be have the power capacity to run the compressor when the heaters are on during operation. This may lead to a situation where the user may not be able to cool the patient and/or cardioplegia…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K052601 (Sorin Group Deutschland GmbH, 2006) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K883603.

How does its category compare?

All DWC devices

DWC Controller, Temperature, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DWC

2in 2022

2007: 12010: 12012: 12013: 12014: 12020: 12022: 2
2007high 2 (2022) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWC

Too few DWC 510(k)s in any year since 2007 to give a median.

Recalls per year · DWC

5in 2023

2010: 12012: 22014: 12015: 12016: 12017: 12018: 32021: 52023: 5
2007high 5 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWC

670in 2025

2007: 32009: 22010: 2432011: 2812012: 3512013: 2442014: 3782015: 5752016: 7872017: 7162018: 5642019: 6972020: 5482021: 6222022: 6052023: 9282024: 5552025: 670
2007high 928 (2023) · low 0 (2008)2025

MAUDE reports naming product code DWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—03
20080—00
20090—02
20101—1243
20110—0281
20121—2351
20131—0244
20141—1378
20150—1575
20160—1787
20170—1716
20180—3564
20190—0697
20201—0548
20210—5622
20222—0605
20230—5928
20240—0555
20250—0670
20260—0—

What is it indicated for?

The Jostra HCU 30 is intended to circulate water through heat exchange circuits to warm or cool a patient during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031544: Heater Cooler Unit, Model Hcu 30.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K031544

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026