510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Resert XL HLD High Level Disinfectant

K080420510(k) TraditionalCleared Sep 2, 2008200-day reviewClass II21 CFR 880.6885Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · MED Sterilant, Medical Devices

Cleared on Sep 2, 2008 as substantially equivalent to K970230 (Reckitt Benckiser, Inc., 1997) after a 200-day review. Its family tree reaches back 2 generations to 1994. 1 later device descends from it. It has been recalled once.

Decided
Sep 2, 2008
Review
200 days
Class
Class II
Product code
MED
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K924334 (1994). 1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2010: 1 descendant clearedK924334 · Cidex Formula 7* Long-Life Activated Dialdehyde · Johnson & Johnson Medical, Inc. · 1994K970230 · Sporox · Reckitt Benckiser, Inc. · 19971990200020102020
Descendants, by generation 1 generation forward: 1 descendant K970230 · Sporox · Reckitt Benckiser, Inc. · 1997K924334 · Cidex Formula 7* Long-Life Activated Dialdehyde · Johnson & Johnson Medical, Inc. · 19940
  1. Root · 2 generations backK924334 Cidex Formula 7* Long-Life Activated DialdehydeJohnson & Johnson Medical, Inc. · 1994
  2. 1 generation backK970230 SporoxReckitt Benckiser, Inc. · 1997
  3. This deviceK080420 Resert XL HLD High Level DisinfectantSTERIS Corporation · 2008
  4. Descendant · cites it directly · 0 of its ownK091022 Modification to Resert XL HLD High Level DisinfectantSTERIS Corporation · 2010

Most influential descendants

Showing 1 of 1 · See all
  • K091022Modification to Resert XL HLD High Level DisinfectantSTERIS Corporation · 2010 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K924334Cidex Formula 7* Long-Life Activated DialdehydeJohnson & Johnson Medical, Inc.1994—
−1K970230SporoxReckitt Benckiser, Inc.1997—
0 (this device)K080420Resert XL HLD High Level DisinfectantSTERIS Corporation2008—
+1K091022Modification to Resert XL HLD High Level DisinfectantSTERIS Corporation2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
FIG. 1 — Lineage of K080420, 1994–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MED device

  1. #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K970230Sporox
Summary page 2
…ification Number: 21 CFR 880.6885 Product Code: MED. 2. Predicate Device SPOROX® Sterilizing and Disinfecting Solution (K970230) manufactured by Rickitt & Colman, Inc. 3. Description of Device Resert XL HLD High Level Disinfectant is a nom…
Reckitt Benckiser, Inc.1997—

Also cleared against K970230: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K091022Modification to Resert XL HLD High Level DisinfectantSTERIS Corporation20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 25, 2022Class IIZ-1605-2022Environmental controlFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 10.8 years on median.
  • None of the 2 ancestors has a recall.
  • Of its descendants, K091022 (STERIS Corporation, 2010) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K970230.

How does its category compare?

All MED devices

MED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · MED

2in 2023

2007: 22008: 12010: 62012: 32013: 12014: 12016: 12017: 12018: 22019: 42021: 22022: 12023: 22026 (partial year): 2
2007high 6 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MED

84in 2019

2010: 281 days median2012: 111 days median2019: 84 days median
2007high 281 (2010) · low 84 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MED

2in 2011

2007: 32008: 22011: 2
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MED

9in 2025

2007: 342008: 792009: 1582010: 892011: 1062012: 3372013: 1402014: 1052015: 1102016: 1352017: 742018: 422019: 352020: 322021: 312022: 162023: 92024: 82025: 9
2007high 337 (2012) · low 8 (2024)2025

MAUDE reports naming product code MED, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—334
20081—279
20090—0158
20106281089
20110—2106
201231110337
20131—0140
20141—0105
20150—0110
20161—0135
20171—074
20182—042
2019484035
20200—032
20212—031
20221—016
20232—09
20240—08
20250—09
20262—0—

What is it indicated for?

Resert XL HLD is a ready to use liquid chemical germicide. The product is a 2.0% nominal hydrogen peroxide solution. The minimum recommended concentration (MRC) is 1.5%. Resert XL HLD is intended to be used by health care practitioners in clinical settings as a ready to use liquid chemical disinfectant for the high level disinfection of semi-critical medical devices for which alternative methods of terminal reprocessing are not suitable or available.

High Level Disinfectant: Resert XL HLD High Level Disinfectant is a high level disinfectant when used or reused undiluted for a maximum of 21 days at a minimum temperature of 20°C (68°F) for a minimum immersion time of 8 minutes.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K080420: Resert XL HLD High Level Disinfectant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K080420

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026