510(k) · Cleared
Recalled · Class II · 2022Resert XL HLD High Level Disinfectant
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
STERIS Corporation · MED Sterilant, Medical Devices
Cleared on Sep 2, 2008 as substantially equivalent to K970230 (Reckitt Benckiser, Inc., 1997) after a 200-day review. Its family tree reaches back 2 generations to 1994. 1 later device descends from it. It has been recalled once.
- Decided
- Sep 2, 2008
- Review
- 200 days
- Class
- Class II
- Product code
- MED
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 2 generations to K924334 (1994). 1 later device descends from it, 1 citing it directly.
- Root · 2 generations backK924334 Cidex Formula 7* Long-Life Activated DialdehydeJohnson & Johnson Medical, Inc. · 1994
- 1 generation backK970230 SporoxReckitt Benckiser, Inc. · 1997
- This deviceK080420 Resert XL HLD High Level DisinfectantSTERIS Corporation · 2008
- Descendant · cites it directly · 0 of its ownK091022 Modification to Resert XL HLD High Level DisinfectantSTERIS Corporation · 2010
Most influential descendants
Showing 1 of 1 · See all- K091022Modification to Resert XL HLD High Level DisinfectantSTERIS Corporation · 2010 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K924334 | Cidex Formula 7* Long-Life Activated Dialdehyde | Johnson & Johnson Medical, Inc. | 1994 | — |
| −1 | K970230 | Sporox | Reckitt Benckiser, Inc. | 1997 | — |
| 0 (this device) | K080420 | Resert XL HLD High Level Disinfectant | STERIS Corporation | 2008 | — |
| +1 | K091022 | Modification to Resert XL HLD High Level Disinfectant | STERIS Corporation | 2010 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2010 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MED device
- #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K970230 | SporoxSummary page 2…ification Number: 21 CFR 880.6885 Product Code: MED. 2. Predicate Device SPOROX® Sterilizing and Disinfecting Solution (K970230) manufactured by Rickitt & Colman, Inc. 3. Description of Device Resert XL HLD High Level Disinfectant is a nom… | 1997 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K091022 | Modification to Resert XL HLD High Level Disinfectant | 2010 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 25, 2022 | Class II | Z-1605-2022 | Environmental control | Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 10.8 years on median.
- None of the 2 ancestors has a recall.
- Of its descendants, K091022 (STERIS Corporation, 2010) has gone furthest, with 0 descendants of its own.
- 2 other devices also cite its predicate K970230.
How does its category compare?
All MED devicesMED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · MED
2in 2023
2026 to Oct 25, outlined.
Median 510(k) review, days · MED
84in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MED
2in 2011
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MED
9in 2025
MAUDE reports naming product code MED, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 3 | 34 |
| 2008 | 1 | — | 2 | 79 |
| 2009 | 0 | — | 0 | 158 |
| 2010 | 6 | 281 | 0 | 89 |
| 2011 | 0 | — | 2 | 106 |
| 2012 | 3 | 111 | 0 | 337 |
| 2013 | 1 | — | 0 | 140 |
| 2014 | 1 | — | 0 | 105 |
| 2015 | 0 | — | 0 | 110 |
| 2016 | 1 | — | 0 | 135 |
| 2017 | 1 | — | 0 | 74 |
| 2018 | 2 | — | 0 | 42 |
| 2019 | 4 | 84 | 0 | 35 |
| 2020 | 0 | — | 0 | 32 |
| 2021 | 2 | — | 0 | 31 |
| 2022 | 1 | — | 0 | 16 |
| 2023 | 2 | — | 0 | 9 |
| 2024 | 0 | — | 0 | 8 |
| 2025 | 0 | — | 0 | 9 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
Resert XL HLD is a ready to use liquid chemical germicide. The product is a 2.0% nominal hydrogen peroxide solution. The minimum recommended concentration (MRC) is 1.5%. Resert XL HLD is intended to be used by health care practitioners in clinical settings as a ready to use liquid chemical disinfectant for the high level disinfection of semi-critical medical devices for which alternative methods of terminal reprocessing are not suitable or available.
High Level Disinfectant: Resert XL HLD High Level Disinfectant is a high level disinfectant when used or reused undiluted for a maximum of 21 days at a minimum temperature of 20°C (68°F) for a minimum immersion time of 8 minutes.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K080420 · printed Nov 3, 2026
Cite this page
510(k) Report. “K080420: Resert XL HLD High Level Disinfectant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K080420
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026