510(k) · Cleared
Recalled · Class II · 2018GE Brightspeed Delight CT Scanner System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ge Medical Systems, LLC · JAK System, X-Ray, Tomography, Computed
Cleared on Oct 8, 2008 as substantially equivalent to K052855 (GE Healthcare, 2005) after a 13-day review. Its family tree reaches back 3 generations to 2001. 6 of the 22 devices in its family, counting itself, have a Class I recall. 16 later devices descend from it. It has been recalled 9 times.
- Decided
- Oct 8, 2008
- Review
- 13 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 9
- Predicates
- 1
Its longest line of descent runs back 3 generations to K013561 (2001). 5 ancestors in all, the oldest K000300 (2000). 16 later devices descend from it, 2 citing it directly. 6 of the 22 devices in this family, counting this one, have a Class I recall.
- Root · 3 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
- 2 generations backK030420 Lightspeed 5.0 Computed Tomography SystemGE Medical Systems · 2003 · Recalled
- 1 generation backK052855 Starspeed Series CT Scanner SystemsGE Healthcare · 2005 · Recalled
- This deviceK082816 GE Brightspeed Delight CT Scanner SystemGe Medical Systems, LLC · 2008
- Descendant · cites it directly · 13 of its ownK093514 Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Recalled (Class I)
- Descendant · 2 generations on · 6 of its ownK153402 Discovery NM/CT 670 CZTGE Healthcare · 2016 · Recalled (Class I)
- Descendant · 3 generations on · 3 of its ownK182484 VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018
- Descendant · 4 generations on · 2 of its ownK190457 VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2019
- Descendant · 2 generations on · 2 of its ownK111445 Discovery NM 630GE Healthcare · 2011 · Recalled (Class I)
Showing 5 of 16 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 16 · See all- K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · cites it directly13descendants
- K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on6descendants
- K182484VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018 · 3 generations on3descendants
- K190457VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2019 · 4 generations on2descendants
- K111445Discovery NM 630GE Healthcare · 2011 · 2 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K013561 | Lightspeed 4.0 CT Scanner System | GE Medical Systems | 2001 | 5 |
| −2 | K030420 | Lightspeed 5.0 Computed Tomography System | GE Medical Systems | 2003 | 4 |
| −1 | K052855 | Starspeed Series CT Scanner Systems | GE Healthcare | 2005 | 5 |
| 0 (this device) | K082816 | GE Brightspeed Delight CT Scanner System | Ge Medical Systems, LLC | 2008 | — |
| +1 | K093514 | Discovery Model NM/CT 670 | Ge Medical Systems F.I. Haifa | 2009 | Class I |
| +2 | K153402 | Discovery NM/CT 670 CZT | GE Healthcare | 2016 | Class I |
| +3 | K182484 | VERITON CT whole body SPECT/CT system | Spectrum Dynamics Medical, Ltd. | 2018 | — |
| +4 | K190457 | VERITON CT whole body SPECT/CT system | Spectrum Dynamics Medical, Ltd. | 2019 | — |
| +2 | K111445 | Discovery NM 630 | GE Healthcare | 2011 | Class I |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2000 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2009 | 1 | 1 |
| 2011 | 2 | 2 |
| 2012 | 1 | 1 |
| 2013 | 1 | 0 |
| 2016 | 1 | 1 |
| 2017 | 1 | 0 |
| 2018 | 2 | 1 |
| 2019 | 1 | 0 |
| 2021 | 1 | 0 |
| 2023 | 1 | 0 |
| 2025 | 3 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K052855 | Starspeed Series CT Scanner SystemsSummary page 2…similar intended uses, such as the previous StarSpeed CT Scanners. Predicate Device(s): GE StarSpeed Series CT Scanner System (K052855) DEVICE DESCRIPTION The GE BrightSpeed Delight Series CT Scanner System (aka GE BrightSpeed Elite, BrightSpeed Elite Select, Br… | 2005 |
What descends from it? (16)
See all 162 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (6 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 24, 2018 | Class II | Z-2586-2018 | Component change control | Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply). |
| Apr 27, 2018 | Class II | Z-2873-2018 | Employee error | GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry. |
| Apr 27, 2018 | Class II | Z-2872-2018 | Employee error | GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry. |
| May 19, 2017 | Class II | Z-2188-2017 | Software design | It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed. |
| May 19, 2017 | Class II | Z-2186-2017 | Software design | It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed. |
| Oct 9, 2013 | Class II | Z-0840-2014 | Software design | There is an issue with the Manual Film Composer feature on some CT products. There is an opportunity, while following a specific workflow, to create a film image with one patient's images and another patient's name in the footer. This is… |
| Jan 19, 2011 | Not classified | Z-2214-2011 | Software design | GE Healthcare has become aware of a potential set of circumstances that could cause X-ray continuation during an unexpected table stop on certain BrightSpeed systems (BrightSpeed Excel/Edge/Elite Select and BrightSpeed Elite Computed… |
| Sep 3, 2010 | Not classified | Z-0369-2011 | Software design | GE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems. |
| Sep 3, 2010 | Not classified | Z-0368-2011 | Software design | GE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems. |
The 5 latest of 9 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 12 Class I recalls
- K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
- K173816NM/CT 850, NM/CT 860GE Healthcare · 2018 · 2 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
- K111445Discovery NM 630GE Healthcare · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
- K113350Brivo NM 615GE Healthcare · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
- K121019Optima NM/CT 640GE Healthcare · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls
Nearest 5 of 6; all are on the web page.
What does this lineage say?
- The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 2.6 years on median.
- 3 of the 5 ancestors have a recall of any class.
- Of its descendants, K093514 (Ge Medical Systems F.I. Haifa, 2009) has gone furthest, with 13 descendants of its own.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
The GE BrightSpeed Delight Series Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, patient for all ages, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient, and equipment…
Read the full statement (147 words)
The GE BrightSpeed Delight Series Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, patient for all ages, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.
The GE BrightSpeed Delight Series CT Scanner System is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K082816 · printed Nov 6, 2026
Cite this page
510(k) Report. “K082816: GE Brightspeed Delight CT Scanner System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082816