510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

GE Brightspeed Delight CT Scanner System

K082816510(k) TraditionalCleared Oct 8, 200813-day reviewClass II21 CFR 892.1750Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · JAK System, X-Ray, Tomography, Computed

Cleared on Oct 8, 2008 as substantially equivalent to K052855 (GE Healthcare, 2005) after a 13-day review. Its family tree reaches back 3 generations to 2001. 6 of the 22 devices in its family, counting itself, have a Class I recall. 16 later devices descend from it. It has been recalled 9 times.

Decided
Oct 8, 2008
Review
13 days
Class
Class II
Product code
JAK
Recalls
9
Predicates
1

Its longest line of descent runs back 3 generations to K013561 (2001). 5 ancestors in all, the oldest K000300 (2000). 16 later devices descend from it, 2 citing it directly. 6 of the 22 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 2 other ancestors2009: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared2016: 1 descendant cleared, 1 with a Class I recall12025: 3 descendants cleared2018: 2 descendants cleared, 1 with a Class I recall12011: 2 descendants cleared, 2 with a Class I recall22017: 1 descendant cleared2012: 1 descendant cleared, 1 with a Class I recall12021: 1 descendant cleared2019: 1 descendant cleared2023: 1 descendant cleared2026: 1 descendant clearedK013561 · Lightspeed 4.0 CT Scanner System · GE Medical Systems · 2001 · recalledK030420 · Lightspeed 5.0 Computed Tomography System · GE Medical Systems · 2003 · recalledK052855 · Starspeed Series CT Scanner Systems · GE Healthcare · 2005 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 5 descendants, 3 with a Class I recall33 generations forward: 5 descendants, 2 with a Class I recall24 generations forward: 2 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant K052855 · Starspeed Series CT Scanner Systems · GE Healthcare · 2005 · recalledK030420 · Lightspeed 5.0 Computed Tomography System · GE Medical Systems · 2003 · recalledK013561 · Lightspeed 4.0 CT Scanner System · GE Medical Systems · 2001 · recalled0+5
  1. Root · 3 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
  2. 2 generations backK030420 Lightspeed 5.0 Computed Tomography SystemGE Medical Systems · 2003 · Recalled
  3. 1 generation backK052855 Starspeed Series CT Scanner SystemsGE Healthcare · 2005 · Recalled
  4. This deviceK082816 GE Brightspeed Delight CT Scanner SystemGe Medical Systems, LLC · 2008
  5. Descendant · cites it directly · 13 of its ownK093514 Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Recalled (Class I)
  6. Descendant · 2 generations on · 6 of its ownK153402 Discovery NM/CT 670 CZTGE Healthcare · 2016 · Recalled (Class I)
  7. Descendant · 3 generations on · 3 of its ownK182484 VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018
  8. Descendant · 4 generations on · 2 of its ownK190457 VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2019
  9. Descendant · 2 generations on · 2 of its ownK111445 Discovery NM 630GE Healthcare · 2011 · Recalled (Class I)

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2009: 12011: 22012: 12013: 12016: 12017: 12018: 22019: 12021: 12023: 12025: 32026: 1
2009high 3 (2025) · low 0 (2010)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 16 · See all
  • K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · cites it directly13descendants
  • K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on6descendants
  • K182484VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018 · 3 generations on3descendants
  • K190457VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2019 · 4 generations on2descendants
  • K111445Discovery NM 630GE Healthcare · 2011 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K013561Lightspeed 4.0 CT Scanner SystemGE Medical Systems20015
−2K030420Lightspeed 5.0 Computed Tomography SystemGE Medical Systems20034
−1K052855Starspeed Series CT Scanner SystemsGE Healthcare20055
0 (this device)K082816GE Brightspeed Delight CT Scanner SystemGe Medical Systems, LLC2008—
+1K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa2009Class I
+2K153402Discovery NM/CT 670 CZTGE Healthcare2016Class I
+3K182484VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd.2018—
+4K190457VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd.2019—
+2K111445Discovery NM 630GE Healthcare2011Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200911
201122
201211
201310
201611
201710
201821
201910
202110
202310
202530
202610
FIG. 1 — Lineage of K082816, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052855Starspeed Series CT Scanner Systems
Summary page 2
…similar intended uses, such as the previous StarSpeed CT Scanners. Predicate Device(s): GE StarSpeed Series CT Scanner System (K052855) DEVICE DESCRIPTION The GE BrightSpeed Delight Series CT Scanner System (aka GE BrightSpeed Elite, BrightSpeed Elite Select, Br…
GE Healthcare20055

What descends from it? (16)

See all 16

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093514Discovery Model NM/CT 670 Class IGe Medical Systems F.I. Haifa200913
K123596Optima CT520Ge Hangwei Medical Systems Co., Ltd.20131

Has anything in its family been recalled? (6 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 24, 2018Class IIZ-2586-2018Component change controlSome CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Apr 27, 2018Class IIZ-2873-2018Employee errorGE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
Apr 27, 2018Class IIZ-2872-2018Employee errorGE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
May 19, 2017Class IIZ-2188-2017Software designIt was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
May 19, 2017Class IIZ-2186-2017Software designIt was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
Oct 9, 2013Class IIZ-0840-2014Software designThere is an issue with the Manual Film Composer feature on some CT products. There is an opportunity, while following a specific workflow, to create a film image with one patient's images and another patient's name in the footer. This is…
Jan 19, 2011Not classifiedZ-2214-2011Software designGE Healthcare has become aware of a potential set of circumstances that could cause X-ray continuation during an unexpected table stop on certain BrightSpeed systems (BrightSpeed Excel/Edge/Elite Select and BrightSpeed Elite Computed…
Sep 3, 2010Not classifiedZ-0369-2011Software designGE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems.
Sep 3, 2010Not classifiedZ-0368-2011Software designGE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 12 Class I recalls
  • K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K173816NM/CT 850, NM/CT 860GE Healthcare · 2018 · 2 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
  • K111445Discovery NM 630GE Healthcare · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K113350Brivo NM 615GE Healthcare · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
  • K121019Optima NM/CT 640GE Healthcare · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 7 years and passes through 3 companies; a new generation was cleared every 2.6 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • Of its descendants, K093514 (Ge Medical Systems F.I. Haifa, 2009) has gone furthest, with 13 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The GE BrightSpeed Delight Series Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, patient for all ages, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient, and equipment…

Read the full statement (147 words)

The GE BrightSpeed Delight Series Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, patient for all ages, including Axial, Cine, Helical, Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.

The GE BrightSpeed Delight Series CT Scanner System is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications.

The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082816: GE Brightspeed Delight CT Scanner System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082816

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026