510(k) Report

510(k) · Cleared

Recalled · 2012

Model TSX-301B Aquilion CT System

K083282510(k) TraditionalCleared Dec 12, 200835-day reviewClass II21 CFR 892.1750Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshiba America Medical Systems, In.C · JAK System, X-Ray, Tomography, Computed

Cleared on Dec 12, 2008 as substantially equivalent to K072606 (Toshiba America Medical Systems, In.C, 2007) and 1 other predicate after a 35-day review. Its family tree reaches back 3 generations to 2003. 1 later device descends from it. It has been recalled 8 times.

Decided
Dec 12, 2008
Review
35 days
Class
Class II
Product code
JAK
Recalls
8
Predicates
2

Its longest line of descent runs back 3 generations to K031469 (2003). 5 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 3 generations backK031469 Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7Toshiba America Medical Systems, In.C · 2003 · Recalled
  2. 2 generations backK033418 TSX-101A/a and TSX-101A/C Aquilion Multislice CT ScannersToshiba America Medical Systems, In.C · 2003 · Recalled
  3. 1 generation backK051833 TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT…Toshiba America Medical Systems, In.C · 2005
  4. This deviceK083282 Model TSX-301B Aquilion CT SystemToshiba America Medical Systems, In.C · 2008
  5. Descendant · cites it directly · 0 of its ownK120546 Aquilion PremiumToshiba Medical Systems Corporation, Japan · 2012 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K120546Aquilion PremiumToshiba Medical Systems Corporation, Japan · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K031469Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7Toshiba America Medical Systems, In.C20031
−2K033418TSX-101A/a and TSX-101A/C Aquilion Multislice CT ScannersToshiba America Medical Systems, In.C20032
−1K051833TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT ScannerToshiba America Medical Systems, In.C2005—
0 (this device)K083282Model TSX-301B Aquilion CT SystemToshiba America Medical Systems, In.C2008—
+1K120546Aquilion PremiumToshiba Medical Systems Corporation, Japan20122

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200610
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
FIG. 1 — Lineage of K083282, 2003–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072606Aquilion One TSX-301A/2 Computed Tomography System
Summary page 1
…Subchapter J, Federal Diagnostic X-ray Equipment Standard Predicate Device(s): TSX-301A12, Aquilion ONE CT System, k072606 TSX-lI lA/h, /I Aquilion 32/64 SP CT System, kOS 1833…
Toshiba America Medical Systems, In.C20074
*K051833TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT ScannerToshiba America Medical Systems, In.C2005—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K120546Aquilion PremiumToshiba Medical Systems Corporation, Japan20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 3, 2012Not classifiedZ-1034-2012Software designThe recall was initiated because Toshiba has confirmed that their is a potential software issue for the TSX-301A Aquilon One and the TSX-301B Aquilon One.
Jan 3, 2012Not classifiedZ-1035-2012Software designThe recall was initiated because Toshiba has confirmed that their is a potential software issue for the TSX-301A Aquilon One and the TSX-301B Aquilon One.
Jun 22, 2011Not classifiedZ-3195-2011Software designToshiba America Medical Systems is initiating a field correction on their TSX-301B: AQUILION Premium; System, X-Ray, Tomography, Computed, TSX-301A: AQUILION One; System, X-Ray, Tomography, Computed, TSX-20IA: AQUlLlON LB; System, X-Ray…
Jan 7, 2011Not classifiedZ-2895-2011Software designToshiba America Medical Systems (TAMS) is initiating a field correction for the TSX-301A and TSX-301B: Aquilion 64 System, X-Ray, Tomography because communication between reconstruction units in the system console may be broken due to…
Jan 7, 2011Not classifiedZ-2896-2011Software designToshiba America Medical Systems (TAMS) is initiating a field correction for the TSX-301A and TSX-301B: Aquilion 64 System, X-Ray, Tomography because communication between reconstruction units in the system console may be broken due to…
Sep 7, 2010Not classifiedZ-1257-2011Software designIncorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is approximately 10%…
Jul 22, 2010Not classifiedZ-1665-2011Software designToshiba America Medical Systems (TAMS) has found that communication may be cut off in the Aquilion ONE/ Premium Systems, owing to the failure of a part used in the console.
Jan 7, 2010Not classifiedZ-0907-2010Device DesignWhen W-Volume is performed in Prospective CTA or Target CTA studies with ECG gating, unnecessary intervals may be set before starting the volume scan. As a result, scanning cannot be performed at the appropriate timing of constant…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 1 company; a new generation was cleared every 1.7 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • Of its descendants, K120546 (Toshiba Medical Systems Corporation, Japan, 2012) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head.

The Aquilion has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K083282: Model TSX-301B Aquilion CT System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K083282

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026