510(k) · Cleared
Recalled · 2009TSX-101A/a and TSX-101A/C Aquilion Multislice CT Scanners
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Toshiba America Medical Systems, In.C · JAK System, X-Ray, Tomography, Computed
Cleared on Nov 10, 2003 as substantially equivalent to K031469 (Toshiba America Medical Systems, In.C, 2003) after a 14-day review. Its family tree reaches back 1 generation to 2003. 25 later devices descend from it. It has been recalled 2 times.
- Decided
- Nov 10, 2003
- Review
- 14 days
- Class
- Class II
- Product code
- JAK
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 1 generation to K031469 (2003). 25 later devices descend from it, 1 citing it directly.
- Root · 1 generation backK031469 Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7Toshiba America Medical Systems, In.C · 2003 · Recalled
- This deviceK033418 TSX-101A/a and TSX-101A/C Aquilion Multislice CT ScannersToshiba America Medical Systems, In.C · 2003
- Descendant · cites it directly · 24 of its ownK051833 TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT…Toshiba America Medical Systems, In.C · 2005
- Descendant · 2 generations on · 17 of its ownK080211 Model TSX-101A/H/I Aquilion 32/64 SP CT SystemToshiba America Medical Systems, In.C · 2008 · Recalled
- Descendant · 3 generations on · 15 of its ownK101888 Scenaria Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2011 · Recalled
- Descendant · 4 generations on · 14 of its ownK123509 Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2013
- Descendant · 5 generations on · 11 of its ownK161748 SUPRIA w/guideShot OptionHitachi Medical Systems America, Inc. · 2016
Showing 5 of 25 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 25 · See all- K051833TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT ScannerToshiba America Medical Systems, In.C · 2005 · cites it directly24descendants
- K080211Model TSX-101A/H/I Aquilion 32/64 SP CT SystemToshiba America Medical Systems, In.C · 2008 · 2 generations on17descendants
- K101888Scenaria Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2011 · 3 generations on15descendants
- K123509Scenaria Phase 2 Whole-Body X-Ray CT SystemHitachi Medical Systems America, Inc. · 2013 · 4 generations on14descendants
- K161748SUPRIA w/guideShot OptionHitachi Medical Systems America, Inc. · 2016 · 5 generations on11descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K031469 | Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7 | Toshiba America Medical Systems, In.C | 2003 | 1 |
| 0 (this device) | K033418 | TSX-101A/a and TSX-101A/C Aquilion Multislice CT Scanners | Toshiba America Medical Systems, In.C | 2003 | — |
| +1 | K051833 | TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT Scanner | Toshiba America Medical Systems, In.C | 2005 | — |
| +2 | K080211 | Model TSX-101A/H/I Aquilion 32/64 SP CT System | Toshiba America Medical Systems, In.C | 2008 | 8 |
| +3 | K101888 | Scenaria Whole-Body X-Ray CT System | Hitachi Medical Systems America, Inc. | 2011 | 3 |
| +4 | K123509 | Scenaria Phase 2 Whole-Body X-Ray CT System | Hitachi Medical Systems America, Inc. | 2013 | — |
| +5 | K161748 | SUPRIA w/guideShot Option | Hitachi Medical Systems America, Inc. | 2016 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
| 2008 | 2 | 0 |
| 2011 | 1 | 0 |
| 2012 | 2 | 0 |
| 2013 | 2 | 0 |
| 2015 | 3 | 0 |
| 2016 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 1 | 0 |
| 2019 | 2 | 0 |
| 2020 | 2 | 0 |
| 2022 | 1 | 0 |
| 2023 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAK device
- #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K031469 | Aquilion Super 4 Multislice CT Scanner, Model TSX-101A/7Summary page 1…ated accessories. Materials and construction are similar to the Aquilion CT Scanner Multi-slice 16 (CGS31A) [KO220501 and the TSX- l O l N 7 Aquilion Super 4 CT Scanner [K031469], and are compliant with IEC60601-1 and associated collateral and particular standards, and applicable sections of 21 CFR Subchap… | 2003 |
| * | — | TSX-101N8 Aquilion 16 CT Named without a submission number | |
| * | — | TSXC-101A/8 CT System Named without a submission number | |
| † | K022050 | CGS-31A Multislice Upgrade Kit for Aquilion CT Scanner. TSX-101A (Aquilion Advance Multi) | 2002 |
What descends from it? (25)
See all 251 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K051833 | TSX-101A/H Aquilion 64 SP CT Scanner; TSX-101A/I Aquilion 32 SP CT Scanner | 2005 | 24 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 8, 2009 | Not classified | Z-1164-2010 | Component design/selection | Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large… |
| Jan 8, 2009 | Not classified | Z-1165-2010 | Component design/selection | Toshiba America Medical Systems Inc (TAMS) has initiated a field correction on the Aquilion 32 CT Scanners due to the potential for the bolus tracking software (SureStart) to cause delay of the helical scan, up to 10 seconds, when large… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Its one ancestor has a recall.
- Of its descendants, K051833 (Toshiba America Medical Systems, In.C, 2005) has gone furthest, with 24 descendants of its own.
How does its category compare?
All JAK devicesJAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JAK
27in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAK
141in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAK
17in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAK
219in 2025
MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 14 | 27 | 11 | 76 |
| 2008 | 26 | 42 | 16 | 120 |
| 2009 | 15 | 31 | 29 | 105 |
| 2010 | 7 | 92 | 51 | 119 |
| 2011 | 16 | 155 | 34 | 205 |
| 2012 | 28 | 121 | 23 | 192 |
| 2013 | 17 | 132 | 11 | 414 |
| 2014 | 30 | 200 | 10 | 308 |
| 2015 | 30 | 140 | 28 | 454 |
| 2016 | 25 | 129 | 44 | 365 |
| 2017 | 29 | 96 | 33 | 338 |
| 2018 | 24 | 114 | 10 | 543 |
| 2019 | 28 | 94 | 5 | 233 |
| 2020 | 30 | 112 | 8 | 177 |
| 2021 | 20 | 145 | 15 | 261 |
| 2022 | 30 | 131 | 9 | 199 |
| 2023 | 27 | 139 | 9 | 310 |
| 2024 | 26 | 141 | 20 | 212 |
| 2025 | 27 | 141 | 17 | 219 |
| 2026 | 21 | 167 | 23 | — |
What is it indicated for?
X-ray imaging of whole body - Computerized Tomography
Including:
Axial
Volumetric (Helical)
CT Fluoroscopy
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K033418 · printed Nov 5, 2026
Cite this page
510(k) Report. “K033418: TSX-101A/a and TSX-101A/C Aquilion Multislice CT Scanners.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K033418