510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

O-Arm Imaging System

K092564510(k) AbbreviatedCleared Jun 17, 2010301-day reviewClass II21 CFR 892.1650Recalls 18

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Navigation, Inc. (Littleton) · OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Cleared on Jun 17, 2010 as substantially equivalent to K050996 (Breakaway Imaging, LLC, 2005) and 1 other predicate after a 301-day review. Its family tree reaches back 4 generations to 2000. 2 of the 23 devices in its family, counting itself, have a Class I recall. 16 later devices descend from it. It has been recalled 18 times.

Decided
Jun 17, 2010
Review
301 days
Class
Class II
Product code
OXO
Recalls
18
Predicates
2

Its longest line of descent runs back 4 generations to K003266 (2000). 6 ancestors in all. 16 later devices descend from it, 2 citing it directly. 2 of the 23 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors, 1 with a Class I recall12015: 1 descendant cleared2021: 2 descendants cleared, 1 with a Class I recall12017: 1 descendant cleared2020: 1 descendant cleared2025: 3 descendants cleared2024: 6 descendants cleared2022: 1 descendant cleared2026: 1 descendant clearedK003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 2000K032280 · Siremobil Iso-C 3D · Siemens AG · 2003K050996 · O-Arm Imaging System · Breakaway Imaging, LLC · 2005 · recalledK060344 · O-Arm Imaging System · Breakaway Imaging, LLC · 2006 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 3 generations back: 2 other ancestors, 1 with a Class I recall1 1 generation forward: 2 descendants2 generations forward: 2 descendants, 1 with a Class I recall13 generations forward: 8 descendants4 generations forward: 3 descendants5 generations forward: 1 descendant K060344 · O-Arm Imaging System · Breakaway Imaging, LLC · 2006 · recalledK050996 · O-Arm Imaging System · Breakaway Imaging, LLC · 2005 · recalledK032280 · Siremobil Iso-C 3D · Siemens AG · 2003K003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 20000+5
  1. Root · 4 generations backK003266 Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc. · 2000
  2. 3 generations backK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. 2 generations backK050996 O-Arm Imaging SystemBreakaway Imaging, LLC · 2005 · Recalled
  4. 1 generation backK060344 O-Arm Imaging SystemBreakaway Imaging, LLC · 2006 · Recalled
  5. This deviceK092564 O-Arm Imaging SystemMedtronic Navigation, Inc. (Littleton) · 2010
  6. Descendant · cites it directly · 15 of its ownK151000 O-Arm O2 Imaging SystemMedtronic, Inc. · 2015 · Recalled
  7. Descendant · 2 generations on · 14 of its ownK173664 Medtronic O-arm O2 Imaging SystemMedtronic, Inc. · 2017 · Recalled
  8. Descendant · 3 generations on · 13 of its ownK200074 O-arm O2 Imaging SystemMedtronic, Inc. · 2020
  9. Descendant · cites it directly · 10 of its ownK210912 Excelsius3DGlobus Medical, Inc. · 2021
  10. Descendant · 2 generations on · 9 of its ownK211616 ExcelsiusHubGlobus Medical, Inc. · 2021 · Recalled (Class I)

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2015: 12017: 12020: 12021: 22022: 12024: 62025: 32026: 1
2015high 6 (2024) · low 0 (2016)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 16 · See all
  • K151000O-Arm O2 Imaging SystemMedtronic, Inc. · 2015 · cites it directly15descendants
  • K173664Medtronic O-arm O2 Imaging SystemMedtronic, Inc. · 2017 · 2 generations on14descendants
  • K200074O-arm O2 Imaging SystemMedtronic, Inc. · 2020 · 3 generations on13descendants
  • K210912Excelsius3DGlobus Medical, Inc. · 2021 · cites it directly10descendants
  • K211616ExcelsiusHubGlobus Medical, Inc. · 2021 · 2 generations on9descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K003266Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc.2000—
−3K032280Siremobil Iso-C 3DSiemens AG2003—
−2K050996O-Arm Imaging SystemBreakaway Imaging, LLC20054
−1K060344O-Arm Imaging SystemBreakaway Imaging, LLC20061
0 (this device)K092564O-Arm Imaging SystemMedtronic Navigation, Inc. (Littleton)2010—
+1K151000O-Arm O2 Imaging SystemMedtronic, Inc.20157
+2K173664Medtronic O-arm O2 Imaging SystemMedtronic, Inc.20172
+3K200074O-arm O2 Imaging SystemMedtronic, Inc.2020—
+1K210912Excelsius3DGlobus Medical, Inc.2021—
+2K211616ExcelsiusHubGlobus Medical, Inc.2021Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200121

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201710
202010
202121
202210
202460
202530
202610
FIG. 1 — Lineage of K092564, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OXO device

  1. #1 K253296 Digital C-ARM with 3Dscore 15.1 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050996O-Arm Imaging System
Summary page 1
…tem Mobile x-ray system Name: System, X-ray, mobile Regulation: 21 CFR 892.1720 Product cddt;{ LH N ¢ o0 O-arm® Imaging System, which was cleared to market in 510(k)s K050996 and K060344, The O-arm® Imaging System is a mobile x-ray system which provides 3D imaging as well as 2D fluoroscopic imaging. T…
Breakaway Imaging, LLC20054
*K060344O-Arm Imaging System
Summary page 1
…-ray system Name: System, X-ray, mobile Regulation: 21 CFR 892.1720 Product cddt;{ LH N ¢ o0 O-arm® Imaging System, which was cleared to market in 510(k)s K050996 and K060344, The O-arm® Imaging System is a mobile x-ray system which provides 3D imaging as well as 2D fluoroscopic imaging. The system co…
Breakaway Imaging, LLC20061

Also cleared against K050996: 2 other devices →

What descends from it? (16)

See all 16

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151000O-Arm O2 Imaging SystemMedtronic, Inc.201515
K210912Excelsius3DGlobus Medical, Inc.202110

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2019Class IIZ-1932-2019Device DesignThe 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in…
Aug 30, 2018Class IIZ-0271-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0273-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0280-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0275-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0277-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0272-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0278-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0279-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0274-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Aug 30, 2018Class IIZ-0276-2019Software designSoftware anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version…
Jan 25, 2018Class IIZ-0928-2018Software designSoftware update ("SW Update 3.2.1") is being implemented to address the following issues: Inability to power the system after shutdown, System stays in standalone mode, Image reconstruction, System/Pendant bootup, Dose display/report…
Mar 17, 2016Class IIZ-2082-2016Radiation Control for Health and Safety ActIt has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
Jul 25, 2013Class IIZ-0973-2014Radiation Control for Health and Safety ActReview of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in…
Jan 31, 2013Class IIZ-0926-2013Component design/selectionPotential failure of the braking system that controls the O-arm Imaging System gantry movement in the vertical direction
Jul 11, 2012Class IIZ-2244-2012Component change controlA fault in a high voltage relay in the 0-arm¿ Imaging System might cause the relay to overheat and burn.
Oct 15, 2010Not classifiedZ-0447-2011Equipment maintenanceSafety Notice: Door Cable Periodic Maintenance Schedule and Cable Wear for the O-ARM Imaging System Breakage of the door cable during opening and closing may occur and cause physical harm or delay of therapy
Oct 12, 2010Class IIZ-1849-2012Software designSoftware Version 3.1.1 does not meet navigational accuracy.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 3 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall
  • K211616ExcelsiusHubGlobus Medical, Inc. · 2021 · 2 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 10 years and passes through 4 companies; a new generation was cleared every 2.8 years on median.
  • 4 of the 6 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K151000 (Medtronic, Inc., 2015) has gone furthest, with 15 descendants of its own.
  • 2 other devices also cite its predicate K050996.

How does its category compare?

All OXO devices

OXO Image-Intensified Fluoroscopic X-Ray System, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · OXO

4in 2025

2010: 12012: 42013: 22014: 22015: 12016: 32017: 42018: 12019: 32021: 22024: 22025: 42026 (partial year): 2
2007high 4 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OXO

97in 2025

2012: 48 days median2016: 128 days median2017: 57 days median2019: 213 days median2025: 97 days median
2007high 213 (2019) · low 48 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OXO

4in 2024

2012: 42013: 32014: 42016: 32017: 42018: 112024: 42026 (partial year): 1
2007high 11 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OXO

118in 2025

2012: 52013: 412014: 892015: 972016: 3,8202017: 1,6972018: 1,6472019: 1,2972020: 8162021: 5542022: 2952023: 2752024: 1922025: 118
2007high 3,820 (2016) · low 0 (2007)2025

MAUDE reports naming product code OXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—00
20110—00
201244845
20132—341
20142—489
20151—097
2016312833,820
201745741,697
20181—111,647
2019321301,297
20200—0816
20212—0554
20220—0295
20230—0275
20242—4192
20254970118
20262—1—

What is it indicated for?

The O-arm® Imaging System is a portable x-ray system designed for 2D fluoroscopic and 3D imaging for high contrast objects and anatomic structures.

The O-arm® Imaging System is compatible with certain Image Guided Surgery Systems. The O-arm® Imaging System may be used with medical charged particle radiation therapy systems for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial landmarks.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K092564: O-Arm Imaging System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K092564

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026