510(k) Report

510(k) · Cleared

Cleared

Digital C-ARM with 3D

K253296510(k) TraditionalCleared Jun 5, 2026249-day reviewClass II21 CFR 892.1650

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Allengers Medical Systems Limited · OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Cleared on Jun 5, 2026 as substantially equivalent to K200218 (Allengers Medical Systems Limited, 2020) and 2 other predicates after a 249-day review. Its family tree reaches back 13 generations to 1980. 3 of the 103 devices in its family, counting itself, have a Class I recall.

Decided
Jun 5, 2026
Review
249 days
Class
Class II
Product code
OXO
Recalls
None
Predicates
3

Its longest line of descent runs back 13 generations to K801623 (1980). 102 ancestors in all. No later device cites it as a predicate yet. 3 of the 103 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2025: 1 other ancestor2020: 2 other ancestors2023: 1 other ancestor2018: 1 other ancestor2017: 3 other ancestors2016: 3 other ancestors2013: 2 other ancestors2014: 1 other ancestor2009: 2 other ancestors2005: 1 other ancestor2004: 9 other ancestors2015: 2 other ancestors2008: 2 other ancestors2006: 3 other ancestors2002: 5 other ancestors, 1 with a Class I recall11997: 4 other ancestors1996: 5 other ancestors2012: 2 other ancestors2003: 2 other ancestors, 1 with a Class I recall12001: 5 other ancestors, 1 with a Class I recall12000: 4 other ancestors1999: 2 other ancestors1993: 4 other ancestors1991: 7 other ancestors1986: 1 other ancestor1995: 11 other ancestors1989: 1 other ancestor1994: 1 other ancestor1992: 1 other ancestor1988: 1 other ancestorK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK123529 · Artis Q and Q.Zen- Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2013 · recalledK163286 · ARTIS pheno · Siemens Medical Solution USA, Inc. · 2017 · recalledK181560 · Cios Alpha · Siemens Medical Systems USA, Inc. · 2018 · recalled×2: 2 generations cleared close together, K203504 (2020), K210054 (2021) 19801990200020102020
Other ancestors, by generation 1 generation back: 2 other ancestors2 generations back: 5 other ancestors3 generations back: 7 other ancestors4 generations back: 11 other ancestors5 generations back: 14 other ancestors, 2 with a Class I recall26 generations back: 4 other ancestors7 generations back: 7 other ancestors, 1 with a Class I recall18 generations back: 8 other ancestors9 generations back: 7 other ancestors10 generations back: 11 other ancestors11 generations back: 9 other ancestors12 generations back: 4 other ancestors K210054 · Cios Spin · Siemens Medical Systems USA, Inc. · 2021 · recalledK203504 · Cios Flow · Siemens Medical Solutions USA, Inc. · 2020 · recalledK181560 · Cios Alpha · Siemens Medical Systems USA, Inc. · 2018 · recalledK163286 · ARTIS pheno · Siemens Medical Solution USA, Inc. · 2017 · recalledK123529 · Artis Q and Q.Zen- Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2013 · recalledK090745 · Artis Zee / Zeegos with SW VC14 · Siemens Medical Solutions USA, Inc. · 2009 · recalledK073290 · Artis Zee and Zeego · Siemens Medical Solutions USA, Inc. · 2008 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−10−50
  1. Root · 13 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 9 generations between
  3. 3 generations backK181560 Cios AlphaSiemens Medical Systems USA, Inc. · 2018 · Recalled
  4. 2 generations backK203504 Cios FlowSiemens Medical Solutions USA, Inc. · 2020 · Recalled
  5. 1 generation backK210054 Cios SpinSiemens Medical Systems USA, Inc. · 2021 · Recalled
  6. This deviceK253296 Digital C-ARM with 3DAllengers Medical Systems Limited · 2026
Show all 13 generations
  1. Root · 13 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 12 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 11 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 10 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 9 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 8 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 7 generations backK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. 6 generations backK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  9. 5 generations backK123529 Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2013 · Recalled
  10. 4 generations backK163286 ARTIS phenoSiemens Medical Solution USA, Inc. · 2017 · Recalled
  11. 3 generations backK181560 Cios AlphaSiemens Medical Systems USA, Inc. · 2018 · Recalled
  12. 2 generations backK203504 Cios FlowSiemens Medical Solutions USA, Inc. · 2020 · Recalled
  13. 1 generation backK210054 Cios SpinSiemens Medical Systems USA, Inc. · 2021 · Recalled
  14. This deviceK253296 Digital C-ARM with 3DAllengers Medical Systems Limited · 2026

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−13K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−12K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−11K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−10K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−9K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−8K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−7K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
−6K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
−5K123529Artis Q and Q.Zen- Modular Angiographic SystemSiemens Medical Solutions USA, Inc.201331
−4K163286ARTIS phenoSiemens Medical Solution USA, Inc.201714
−3K181560Cios AlphaSiemens Medical Systems USA, Inc.20187
−2K203504Cios FlowSiemens Medical Solutions USA, Inc.20201
−1K210054Cios SpinSiemens Medical Systems USA, Inc.20211
0 (this device)K253296Digital C-ARM with 3DAllengers Medical Systems Limited2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198810
198910
199170
199210
199340
199410
1995110
199650
199740
199920
200040
200151
200251
200321
200490
200510
200630
200820
200920
201220
201320
201410
201520
201630
201730
201810
202020
202310
202510
FIG. 1 — Lineage of K253296, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OXO device

  1. #1 K250587 Orthoscan TAU Mini C-Armscore 15.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K200218Digiscan FDX
Summary page 5
…Product Code: OXO, JAA, OWB 3. Predicate Device: Device (trade) name: Digiscan FDX 510(K) Number : K200218 Clearance Date : August 7, 2022 Classification Name: Image-Intensifie…
Allengers Medical Systems Limited2020—
*K210054Cios Spin
Summary page 5
…XO, JAA, OWB 4. Reference Device : Reference Device 1 Device (trade) name: Cios Spin 510(K) Number : K210054 Clearance Date : February 05, 2021 Classification Name: Image-Intensi…
Siemens Medical Systems USA, Inc.20211
*K243303Ziehm Vision RFD 3D
Summary page 6
…gers Medical Systems Limited Reference Device 2 Device (trade) name: Ziehm Vision RFD 3D 510(K) Number : K243303 Clearance Date : January 21, 2025 Classification Name: Image-Intensifie…
Ziehm Imaging GmbH2025—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K021049Oec 9800 Plus Digital Mobile Imaging SystemGe Oec Medical Systems · 2002 · 5 generations back · ◆ in the 2002 ancestors bar of FIG. 11 Class I recall
  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 7 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 46 years and passes through 8 companies; a new generation was cleared every 2.0 years on median.
  • 51 of the 102 ancestors have a recall of any class; 3 have a Class I recall.

How does its category compare?

All OXO devices

OXO Image-Intensified Fluoroscopic X-Ray System, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · OXO

4in 2025

2010: 12012: 42013: 22014: 22015: 12016: 32017: 42018: 12019: 32021: 22024: 22025: 42026 (partial year): 2
2007high 4 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OXO

97in 2025

2012: 48 days median2016: 128 days median2017: 57 days median2019: 213 days median2025: 97 days median
2007high 213 (2019) · low 48 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OXO

4in 2024

2012: 42013: 32014: 42016: 32017: 42018: 112024: 42026 (partial year): 1
2007high 11 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OXO

118in 2025

2012: 52013: 412014: 892015: 972016: 3,8202017: 1,6972018: 1,6472019: 1,2972020: 8162021: 5542022: 2952023: 2752024: 1922025: 118
2007high 3,820 (2016) · low 0 (2007)2025

MAUDE reports naming product code OXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—00
20110—00
201244845
20132—341
20142—489
20151—097
2016312833,820
201745741,697
20181—111,647
2019321301,297
20200—0816
20212—0554
20220—0295
20230—0275
20242—4192
20254970118
20262—1—

What is it indicated for?

Digiscan V30 3D is X-Ray based fluoroscopy equipment. It is intended for use in visualizing complex anatomical structures and procedures such as 3D Reconstruction, endoscopy Studies (ERCP), urology, Gastro, orthopedic, neurology, Fertility studies (HSG), cholangiography, vascular angiography with and without subtraction, critical care and emergency room procedures. This equipment produces real time images of anatomy using x-rays.

Exclusion:

Device is not meant for Mammography.

Contraindications of device:

Exposure of X-ray should be avoided during pregnancy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K253296: Digital C-ARM with 3D.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K253296

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded Oct 30, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026