510(k) Report

510(k) · Cleared

Cleared

Aesculap Sterilcontainer S

K093493510(k) TraditionalCleared Jul 27, 2010259-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aesculap, Inc. · FRG Wrap, Sterilization

Cleared on Jul 27, 2010 as substantially equivalent to K040865 (Aesculap, 2004) and 1 other predicate after a 259-day review. Its family tree reaches back 1 generation to 2004. 4 later devices descend from it.

Decided
Jul 27, 2010
Review
259 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K040865 (2004). 2 ancestors in all. 4 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2008: 1 other ancestor2015: 1 descendant cleared2018: 1 descendant cleared2011: 1 descendant cleared2017: 1 descendant clearedK040865 · Aesculap Sterrad 100S Compatable Sterilcontainer · Aesculap · 2004200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 1 descendant K040865 · Aesculap Sterrad 100S Compatable Sterilcontainer · Aesculap · 20040
  1. Root · 1 generation backK040865 Aesculap Sterrad 100S Compatable SterilcontainerAesculap · 2004
  2. This deviceK093493 Aesculap Sterilcontainer SAesculap, Inc. · 2010
  3. Descendant · cites it directly · 1 of its ownK110682 Steritite Container System & Meditray ProductsCase Medical, Inc. · 2011
  4. Descendant · cites it directly · 0 of its ownK182032 SterilContainer S SystemAesculap, Inc. · 2018
  5. Descendant · 2 generations on · 0 of its ownK161415 SteriTite Containers System & MediTray ProductsCase Medical, Inc. · 2017
  6. Descendant · cites it directly · 0 of its ownK142970 SterilContainer S SystemAesculap, Inc. · 2015

Most influential descendants

Showing 4 of 4 · See all
  • K110682Steritite Container System & Meditray ProductsCase Medical, Inc. · 2011 · cites it directly1descendant
  • K182032SterilContainer S SystemAesculap, Inc. · 2018 · cites it directly0descendants
  • K161415SteriTite Containers System & MediTray ProductsCase Medical, Inc. · 2017 · 2 generations on0descendants
  • K142970SterilContainer S SystemAesculap, Inc. · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K040865Aesculap Sterrad 100S Compatable SterilcontainerAesculap2004—
0 (this device)K093493Aesculap Sterilcontainer SAesculap, Inc.2010—
+1K110682Steritite Container System & Meditray ProductsCase Medical, Inc.2011—
+1K182032SterilContainer S SystemAesculap, Inc.2018—
+2K161415SteriTite Containers System & MediTray ProductsCase Medical, Inc.2017—
+1K142970SterilContainer S SystemAesculap, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201510
201710
201810
FIG. 1 — Lineage of K093493, 2004–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040865Aesculap Sterrad 100S Compatable Sterilcontainer
Summary page 1
…G SUBSTANTIAL EQUIVALENCE Aesculap®, Inc. believes that the SterilContainer S is substantially equivalent to: * Aesculap STERRAD 1005SCompatible Sterilcontainer (K040865). * SteriTite Universal Container System with MediTray Products fbrSTERRAD 200, STERRAID NX, Ozone 125L, Sterili…
Aesculap2004—
*K080558Steritite Container System & Meditray Products
Summary page 1
…Tite Universal Container System with MediTray Products fbrSTERRAD 200, STERRAID NX, Ozone 125L, Sterilization, and Prevacuum Steam Flash Sterilization (K080558). DEVICE DESCRIPTION The Aesculap Sterilcontainer S is designed as a container system that will all ow for sterilization and sto…
Case Medical, Inc.20081

Also cleared against K040865: 1 other device →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K110682Steritite Container System & Meditray ProductsCase Medical, Inc.20111
K182032SterilContainer S SystemAesculap, Inc.20180
K142970SterilContainer S SystemAesculap, Inc.20150

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K110682 (Case Medical, Inc., 2011) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K040865.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

The Aesculap Sterilcontainer S is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and disposable polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This container has been validated with stainless steel lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed…

Read the full statement (217 words)

The Aesculap Sterilcontainer S is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and disposable polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This container has been validated with stainless steel lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used, This container system is compatible for use with the STERRAD 200, STERRAD NX (Standard cycle and Advanced cycle), and STERRAD 100ONX (Standard cycle and Flex cycle). The Sterilcontainer S System includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

The Steriicontainer S is recommended for surface and lumens:

STERRAD 200, stainless steel lumens Ž:3mm I.D. x!•400mm L

STERRAD NX standard cycle, stainless steel lumens Ž 2mm I.D. x S400mm L

STERRAD NX advanced cycle, stainless steel lumens Ž 1mm l.D. x S 500mm L

STERRADI1 ONX standard cycle, stainless steel lumens Ž0.7mm l.D. x 5500mm L

STERRAD 1 ONX flex cycle, porous lumens (flexible endoscope) Žlmm1.0. x S850mm L

30 days of shelf life testing has been conducted for the Sterilicontainer S.

The attached tables identify which products may be sterilized in which STERRAD sterilization cycle.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093493: Aesculap Sterilcontainer S.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093493

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026