510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Artiste Solution with Sys_VB50 Update

K103606510(k) TraditionalCleared Apr 15, 2011127-day reviewClass II21 CFR 892.5050Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · IYE Accelerator, Linear, Medical

Cleared on Apr 15, 2011 as substantially equivalent to K102671 (Siemens Medical Solutions USA, Inc., 2010) and 4 other predicates after a 127-day review. Its family tree reaches back 12 generations to 1992. 4 later devices descend from it. It has been recalled 7 times.

Decided
Apr 15, 2011
Review
127 days
Class
Class II
Product code
IYE
Recalls
7
Predicates
5

Its longest line of descent runs back 12 generations to K914698 (1992). 88 ancestors in all, the oldest K841997 (1984). 4 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2010: 1 other ancestor2006: 7 other ancestors2008: 1 other ancestor2007: 1 other ancestor2005: 2 other ancestors2004: 7 other ancestors1999: 2 other ancestors1994: 4 other ancestors2003: 3 other ancestors2002: 4 other ancestors2001: 7 other ancestors1997: 3 other ancestors1996: 5 other ancestors1995: 3 other ancestors1990: 3 other ancestors1988: 1 other ancestor1986: 1 other ancestor2000: 4 other ancestors1992: 2 other ancestors1991: 3 other ancestors1984: 1 other ancestor1998: 5 other ancestors1993: 2 other ancestors1989: 3 other ancestors1987: 1 other ancestor2014: 2 descendants cleared2012: 1 descendant cleared2018: 1 descendant clearedK914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K031764 · Oncor Avant-Garde with Coherence Workspaces · Siemens Medical Solutions USA, Inc. · 2003 · recalledK060226 · Oncor Expression · Siemens Medical Solutions USA, Inc. · 2006 · recalledK072485 · Artiste MV SA · Siemens Medical Solutions USA, Inc. · 2007 · recalled×2: 2 generations cleared close together, K082775 (2009), K090683 (2009) K102671 · Syngo RT Therapist Connect Workspace, V4.2 · Siemens Medical Solutions USA, Inc. · 201019801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 10 other ancestors3 generations back: 18 other ancestors4 generations back: 14 other ancestors5 generations back: 10 other ancestors6 generations back: 15 other ancestors7 generations back: 6 other ancestors8 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 1 descendant K102671 · Syngo RT Therapist Connect Workspace, V4.2 · Siemens Medical Solutions USA, Inc. · 2010K090683 · Syngo RT Therapist Connect · Siemens Medical Solutions USA, Inc. · 2009 · recalledK082775 · Prescision Option · Siemens Medical Solutions USA, Inc. · 2009K072485 · Artiste MV SA · Siemens Medical Solutions USA, Inc. · 2007 · recalledK060226 · Oncor Expression · Siemens Medical Solutions USA, Inc. · 2006 · recalledK031764 · Oncor Avant-Garde with Coherence Workspaces · Siemens Medical Solutions USA, Inc. · 2003 · recalledK022036 · Siemens Virtual Simulation (Vsim) · Siemens Medical Solutions USA, Inc. · 2002K013112 · Focal Sim · Computerized Medical Systems, Inc. · 2001 · recalledK993923 · Adac PINNACLE3TM Radiation Therapy Planning System · Adac Laboratories · 2000K990833 · Dose Calc, Model VS. 1.02 · Integrity Medical Software, Inc. · 1999K980904 · Mucheck Monitor Unit Validation Program · Medical Dosimetry Services, Inc. · 1998K914698 · Setup Computer with Software V-1261,62 and 63 · K.S., Inc. · 1992−10−50
  1. Root · 12 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 8 generations between
  3. 3 generations backK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  4. 2 generations backK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · Recalled
  5. 1 generation backK102671 Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010
  6. This deviceK103606 Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011
  7. Descendant · cites it directly · 3 of its ownK121295 Artistetm Solution with Sys VcioSiemens Medical Solutions USA, Inc. · 2012 · Recalled
  8. Descendant · cites it directly · 2 of its ownK132935 Artiste Solution with Sys_VC10C and Syngo RT Therapist Workspace…Siemens Medical Solutions USA, Inc. · 2014
  9. Descendant · 2 generations on · 1 of its ownK142434 Artiste Solution, Sys_VC10C, Phase 2 Update with the Syngo RT…Siemens Medical Solutions USA, Inc. · 2014 · Recalled
  10. Descendant · 3 generations on · 0 of its ownK182050 Linear Accelerator Control Console Software v13.1.010 Update, syngo…Siemens Medical Solutions USA, Inc. · 2018
Show all 12 generations
  1. Root · 12 generations backK914698 Setup Computer with Software V-1261,62 and 63K.S., Inc. · 1992
  2. 11 generations backK980904 Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc. · 1998
  3. 10 generations backK990833 Dose Calc, Model VS. 1.02Integrity Medical Software, Inc. · 1999
  4. 9 generations backK993923 Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories · 2000
  5. 8 generations backK013112 Focal SimComputerized Medical Systems, Inc. · 2001 · Recalled
  6. 7 generations backK022036 Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc. · 2002
  7. 6 generations backK031764 Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  8. 5 generations backK060226 Oncor ExpressionSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  9. 4 generations backK072485 Artiste MV SASiemens Medical Solutions USA, Inc. · 2007 · Recalled
  10. 3 generations backK082775 Prescision OptionSiemens Medical Solutions USA, Inc. · 2009
  11. 2 generations backK090683 Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc. · 2009 · Recalled
  12. 1 generation backK102671 Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc. · 2010
  13. This deviceK103606 Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc. · 2011

Most influential descendants

Showing 4 of 4 · See all
  • K121295Artistetm Solution with Sys VcioSiemens Medical Solutions USA, Inc. · 2012 · cites it directly3descendants
  • K132935Artiste Solution with Sys_VC10C and Syngo RT Therapist Workspace, V4.3.1 Upgrade Kit…Siemens Medical Solutions USA, Inc. · 2014 · cites it directly2descendants
  • K142434Artiste Solution, Sys_VC10C, Phase 2 Update with the Syngo RT Therapist & RT Oncologist…Siemens Medical Solutions USA, Inc. · 2014 · 2 generations on1descendant
  • K182050Linear Accelerator Control Console Software v13.1.010 Update, syngo Radiation Therapy…Siemens Medical Solutions USA, Inc. · 2018 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K914698Setup Computer with Software V-1261,62 and 63K.S., Inc.1992—
−11K980904Mucheck Monitor Unit Validation ProgramMedical Dosimetry Services, Inc.1998—
−10K990833Dose Calc, Model VS. 1.02Integrity Medical Software, Inc.1999—
−9K993923Adac PINNACLE3TM Radiation Therapy Planning SystemAdac Laboratories2000—
−8K013112Focal SimComputerized Medical Systems, Inc.200110
−7K022036Siemens Virtual Simulation (Vsim)Siemens Medical Solutions USA, Inc.2002—
−6K031764Oncor Avant-Garde with Coherence WorkspacesSiemens Medical Solutions USA, Inc.200317
−5K060226Oncor ExpressionSiemens Medical Solutions USA, Inc.200672
−4K072485Artiste MV SASiemens Medical Solutions USA, Inc.200724
−3K082775Prescision OptionSiemens Medical Solutions USA, Inc.2009—
−2K090683Syngo RT Therapist ConnectSiemens Medical Solutions USA, Inc.20094
−1K102671Syngo RT Therapist Connect Workspace, V4.2Siemens Medical Solutions USA, Inc.2010—
0 (this device)K103606Artiste Solution with Sys_VB50 UpdateSiemens Medical Solutions USA, Inc.2011—
+1K121295Artistetm Solution with Sys VcioSiemens Medical Solutions USA, Inc.20125
+1K132935Artiste Solution with Sys_VC10C and Syngo RT Therapist Workspace, V4.3.1…Siemens Medical Solutions USA, Inc.2014—
+2K142434Artiste Solution, Sys_VC10C, Phase 2 Update with the Syngo RT Therapist & RT…Siemens Medical Solutions USA, Inc.20145
+3K182050Linear Accelerator Control Console Software v13.1.010 Update, syngo Radiation…Siemens Medical Solutions USA, Inc.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198610
198710
198810
198930
199030
199130
199220
199320
199440
199530
199650
199730
199850
199920
200040
200170
200240
200330
200470
200520
200670
200710
200810
201010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201420
201810
FIG. 1 — Lineage of K103606, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K102671Syngo RT Therapist Connect Workspace, V4.2
Summary page 1
…ion: 892.5050 Product Code: IYE Substantial Equivalence Claimed To: %~ syngo®& RT Therapist K102671 /syngb® RT Therapist Connect v4.2 software update Connect for RT Oct. 12, SYS_-VB50 and Control Console 12, an…
Siemens Medical Solutions USA, Inc.2010—
*K072485Artiste MV SA
Summary page 1
…4.2 2010 licensed features,-In-Line kVieWTM and Multiple including Control XTM . Console 12. ART1STETM Solution K072485/I ARTISTETM Solution with the syngo®)RT Therapist (aka ARTISTE MV) Dec. 27, Connect v4.2 with software upda…
Siemens Medical Solutions USA, Inc.200724
*K060633Coherence Oncologist Workspace, Model 2.0 and Physicist Workspace, Model 1.0
Summary page 5
…features in the Dosimetrist v2.7 (KIOI 1119) *Image and plan review and approval and Adaptive Targeting in the Oncologist v2.O (K060633) Summary: In sumnmary, it is SIEMENS' belief that the ARTISTE with the Sys VB50 update does not introduce any new potential saf…
Siemens Medical Solutions USA, Inc.20061
*K101119Syngo Dosimetrist Workspace V2.7 Model: 1056864
Summary page 2
…May 01, Targeting and 3D) portai image approval. Adaptive Targeting and 2006 2D portal image approval. syngo®)Dosimetrist K101119/ syngo®&RT Oncologist v4.2 with Basic, Advance Workspace v2.7 June 16, and Model based Segmentatio…
Siemens Medical Solutions USA, Inc.20101
*K082775Prescision Option
Summary page 5
…Therapist Assist workspaces, electron cone accessories and a collision avoidance system, *and to the stereotactic option FreScision (K082775) for high dose rate, unflat energies, *additionally to the Advanced Segmentation and Model ba…
Siemens Medical Solutions USA, Inc.2009—

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121295Artistetm Solution with Sys VcioSiemens Medical Solutions USA, Inc.20123
K132935Artiste Solution with Sys_VC10C and Syngo RT Therapist Workspace, V4.3.1 Upgrade Kit…Siemens Medical Solutions USA, Inc.20142

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 18, 2020Class IIZ-1677-2020Use errorPotential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
Jul 26, 2016Class IIZ-2876-2016Software designSoftware patch installation to address several safety issues.
Aug 5, 2015Class IIZ-2812-2015Software design (manufacturing process)A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.
Apr 4, 2013Class IIZ-1182-2013Software designSoftware update to fix multiple safety related issues.
Jun 25, 2012Class IIZ-2073-2012Software designPotential safety issue when using the IMMEDIATE RESUMPTION feature after a Control Console interlock during the delivery of flattening-filter-free (Multiple-X) beams or segments. In a very rare scenario, the RT Therapist will send again…
Apr 3, 2012Not classifiedZ-1486-2012Software designSIEMENS has received reports where a dropping down of the 550 TxT" Treatment Table was observed while the table was being moved downwards during operation by means of the hand control.
Nov 2, 2011Not classifiedZ-0462-2012Software designThe software release will bring current Installed Base system to the forward production version. It is a result of the final system testing addressing several safety issues that are not critical for the safe use of the device and multiple…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 6 companies; a new generation was cleared every 1.1 years on median.
  • 33 of the 88 ancestors have a recall of any class.
  • Of its descendants, K121295 (Siemens Medical Solutions USA, Inc., 2012) has gone furthest, with 3 descendants of its own.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT) and precision stereotactic applications. The stereotactic applications may…

Read the full statement (217 words)

The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.

The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT) and precision stereotactic applications. The stereotactic applications may include single-session radiosurgery, fractionated stereotactic radiation therapy, or fractionated stereotactic intensity modulated radiation therapy for lesions, tumors and conditions anywhere in the head and body where radiation therapy is indicated.

The syngo® Suite for Oncology Workspaces:

The syngo® workspaces includes a number of syngo® based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording.

The syngo® RT Therapist Connect Workspace v4.2, contains software applications that permits patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording. The syngo® RT Therapist Connect Workspace v4.2, can be interfaced with third party devices conforming to the DICOM Standard.

The syngo® RT Oncologist Workspace permits localization, contouring, segmentation, image calibration, and review of treatment plan parameters. In addition, it includes tools and administrative functions to aid in the diagnosis, staging, and prescription of radiation therapy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103606: Artiste Solution with Sys_VB50 Update.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103606

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026