510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Giraffe OmniBed Carestation CS1

K152814510(k) SpecialCleared Mar 17, 2016171-day reviewClass II21 CFR 880.5400Recalls 8 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · FMZ Incubator, Neonatal

Cleared on Mar 17, 2016 as substantially equivalent to K101788 (Ohmeda Medical, 2010) and 3 other predicates after a 171-day review. Its family tree reaches back 3 generations to 1997. 4 of the 10 devices in its family, counting itself, have a Class I recall. 3 later devices descend from it. It has been recalled 8 times, including 3 Class I recalls.

Decided
Mar 17, 2016
Review
171 days
Class
Class II
Product code
FMZ
Recalls
8 (3 Class I)
Predicates
4

Its longest line of descent runs back 3 generations to K954204 (1997). 6 ancestors in all. 3 later devices descend from it, 1 citing it directly. 4 of the 10 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2010: 1 other ancestor, 1 with a Class I recall12000: 2 other ancestors2022: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared2026: 1 descendant clearedK954204 · Drager Incubator 8000 IC · Drager, Inc. · 1997K020543 · Ohmeda Medical Giraffe Omnibed · Ohmeda Medical · 2002 · Class I recallK071175 · Ohmeda Medical Giraffe Omnibed · GE Healthcare · 20071995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 1 descendant3 generations forward: 1 descendant K071175 · Ohmeda Medical Giraffe Omnibed · GE Healthcare · 2007K020543 · Ohmeda Medical Giraffe Omnibed · Ohmeda Medical · 2002 · Class I recallK954204 · Drager Incubator 8000 IC · Drager, Inc. · 19970
  1. Root · 3 generations backK954204 Drager Incubator 8000 ICDrager, Inc. · 1997
  2. 2 generations backK020543 Ohmeda Medical Giraffe OmnibedOhmeda Medical · 2002 · Recalled (Class I)
  3. 1 generation backK071175 Ohmeda Medical Giraffe OmnibedGE Healthcare · 2007
  4. This deviceK152814 Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016
  5. Descendant · cites it directly · 2 of its ownK213553 Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · Recalled (Class I)
  6. Descendant · 2 generations on · 1 of its ownK251663 Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025
  7. Descendant · 3 generations on · 0 of its ownK260326 Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation…Datex-Ohmeda, Inc. · 2026

Class I recalls in this family

Most influential descendants

Showing 3 of 3 · See all
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · cites it directly2descendants
  • K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025 · 2 generations on1descendant
  • K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2026 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K954204Drager Incubator 8000 ICDrager, Inc.1997—
−2K020543Ohmeda Medical Giraffe OmnibedOhmeda Medical2002Class I
−1K071175Ohmeda Medical Giraffe OmnibedGE Healthcare2007—
0 (this device)K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company2016—
+1K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc.2022Class I
+2K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc.2025—
+3K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
201011

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202211
202510
202610
FIG. 1 — Lineage of K152814, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMZ device

  1. #1 K251663 Giraffe OmniBed Carestation (CS1); Giraffe…score 14.4 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101788Giraffe Omnibed
Summary page 4
…sification Names: Incubator, Neonatal: (880.5400) Regulatory Class II Product Code: FMZ, FMT Predicate Device(s): Giraffe OmniBed, K101788, K071175 K020543, K993407 Device Description: The Giraffe OmniBed Carestation is an updated version of the cleared…
Ohmeda Medical2010Class I
*K071175Ohmeda Medical Giraffe Omnibed
Summary page 4
…n Names: Incubator, Neonatal: (880.5400) Regulatory Class II Product Code: FMZ, FMT Predicate Device(s): Giraffe OmniBed, K101788, K071175 K020543, K993407 Device Description: The Giraffe OmniBed Carestation is an updated version of the cleared…
GE Healthcare2007—
*K020543Ohmeda Medical Giraffe Omnibed
Summary page 4
…Incubator, Neonatal: (880.5400) Regulatory Class II Product Code: FMZ, FMT Predicate Device(s): Giraffe OmniBed, K101788, K071175 K020543, K993407 Device Description: The Giraffe OmniBed Carestation is an updated version of the cleared…
Ohmeda Medical2002Class I
*K993407Ohmeda Medical Omnibed
Summary page 4
…r, Neonatal: (880.5400) Regulatory Class II Product Code: FMZ, FMT Predicate Device(s): Giraffe OmniBed, K101788, K071175 K020543, K993407 Device Description: The Giraffe OmniBed Carestation is an updated version of the cleared…
Ohmeda Medical20004

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213553Giraffe Incubator Carestation CS1 Class IDatex Ohmeda, Inc.20222

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2026Class IIZ-2477-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
May 1, 2026Class IIZ-2476-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
Sep 3, 2024Class IZ-0048-2025Use errorGE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors…
Jun 6, 2023Class IZ-0140-2025Component design/selectionPrior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were…
May 19, 2023Class IIZ-2100-2023Nonconforming Material/ComponentGE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation…
Oct 11, 2019Class IZ-0539-2020Device DesignGiraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Jul 26, 2019Class IIZ-0281-2020Nonconforming Material/ComponentFasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.
Sep 25, 2018Class IIZ-0479-2019Process designPotential electric leakage affecting the user.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K101788Giraffe OmnibedOhmeda Medical · 2010 · Direct predicate · ◆ in the 2010 ancestors bar of FIG. 13 Class I recalls
  • K020543Ohmeda Medical Giraffe OmnibedOhmeda Medical · 2002 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · Cites it directly · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 19 years and passes through 4 companies; a new generation was cleared every 5.2 years on median.
  • 3 of the 6 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K213553 (Datex Ohmeda, Inc., 2022) has gone furthest, with 2 descendants of its own.

How does its category compare?

All FMZ devices

FMZ Incubator, Neonatal: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · FMZ

2in 2025

2007: 32008: 12010: 42011: 12014: 12015: 12016: 12017: 12018: 22019: 22022: 12025: 22026 (partial year): 2
2007high 4 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMZ

60in 2010

2007: 81 days median2010: 60 days median
2007high 81 (2007) · low 60 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMZ

2in 2024

2008: 42011: 52013: 1, 1 Class I2016: 12018: 32019: 12023: 8, 8 Class I2024: 2, 2 Class I2026 (partial year): 3
2007high 8 (2023) · low 0 (2007)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · FMZ

250in 2025

2007: 142008: 102009: 82010: 112011: 102012: 122013: 112014: 32015: 92016: 122017: 42018: 182019: 242020: 82021: 152022: 272023: 192024: 1182025: 250
2007high 250 (2025) · low 3 (2014)2025

MAUDE reports naming product code FMZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007381014
20081—410
20090—08
2010460011
20111—510
20120—012
20130—1 (1)11
20141—03
20151—09
20161—112
20171—04
20182—318
20192—124
20200—08
20210—015
20221—027
20230—8 (8)19
20240—2 (2)118
20252—0250
20262—3—

What is it indicated for?

The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user’s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed…

Read the full statement (138 words)

The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user’s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.

Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment.

This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152814: Giraffe OmniBed Carestation CS1.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152814

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026