510(k) · Cleared

Recalled · Class II · 2014

Spacelabs Medical Patient Monitors

K102422510(k) SpecialCleared Dec 7, 2010104-day reviewClass II21 CFR 870.1025Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spacelabs Medical, Inc. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Dec 7, 2010 as substantially equivalent to K972282 (Spacelabs Medical, Inc., 1997) after a 104-day review. Its family tree reaches back 2 generations to 1984. 3 later devices descend from it. It has been recalled 3 times.

Decided
Dec 7, 2010
Review
104 days
Class
Class II
Product code
MHX
Recalls
3
Predicates
1

Its longest line of descent runs back 2 generations to K842616 (1984). 3 ancestors in all. 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1993: 1 other ancestor2012: 1 descendant cleared2017: 1 descendant cleared2011: 1 descendant clearedK842616 · Spacelabs Alpha PC Patient Computer · Spacelabs, Inc. · 1984 · recalledK972282 · Spacelabs Medical Pcis Patient Monitors · Spacelabs Medical, Inc. · 1997 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 3 descendants K972282 · Spacelabs Medical Pcis Patient Monitors · Spacelabs Medical, Inc. · 1997 · recalledK842616 · Spacelabs Alpha PC Patient Computer · Spacelabs, Inc. · 1984 · recalled0
  1. Root · 2 generations backK842616 Spacelabs Alpha PC Patient ComputerSpacelabs, Inc. · 1984 · Recalled
  2. 1 generation backK972282 Spacelabs Medical Pcis Patient MonitorsSpacelabs Medical, Inc. · 1997 · Recalled
  3. This deviceK102422 Spacelabs Medical Patient MonitorsSpacelabs Medical, Inc. · 2010
  4. Descendant · cites it directly · 0 of its ownK160667 Anybattery C2 Rechargeable BatteryAnybattery, Inc. · 2017
  5. Descendant · cites it directly · 0 of its ownK120616 Spacelabs Healthcare Qube Compact Monitor (91390Spacelabs Medical, Inc. · 2012 · Recalled
  6. Descendant · cites it directly · 0 of its ownK112962 Spacelabs Healthcare Xprezzon Beside Monitor (9913930)Spacelabs Medical, Inc. · 2011 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K160667Anybattery C2 Rechargeable BatteryAnybattery, Inc. · 2017 · cites it directly0descendants
  • K120616Spacelabs Healthcare Qube Compact Monitor (91390Spacelabs Medical, Inc. · 2012 · cites it directly0descendants
  • K112962Spacelabs Healthcare Xprezzon Beside Monitor (9913930)Spacelabs Medical, Inc. · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K842616Spacelabs Alpha PC Patient ComputerSpacelabs, Inc.19842
−1K972282Spacelabs Medical Pcis Patient MonitorsSpacelabs Medical, Inc.19974
0 (this device)K102422Spacelabs Medical Patient MonitorsSpacelabs Medical, Inc.2010—
+1K160667Anybattery C2 Rechargeable BatteryAnybattery, Inc.2017—
+1K120616Spacelabs Healthcare Qube Compact Monitor (91390Spacelabs Medical, Inc.20129
+1K112962Spacelabs Healthcare Xprezzon Beside Monitor (9913930)Spacelabs Medical, Inc.20117

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201210
201710
FIG. 1 — Lineage of K102422, 1984–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K972282Spacelabs Medical Pcis Patient Monitors
Summary page 1
…The model 91496 was cleared under K050605. Predicate Device Name: Spacelabs Medical PCIS Patient Monitors Devices 510(k) Number: K972282 Manufacturer: Spacelabs Medical Inc. Spacelabs Medical 510(k)…
Spacelabs Medical, Inc.19974

Also cleared against K972282: 2 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K160667Anybattery C2 Rechargeable BatteryAnybattery, Inc.20170
K120616Spacelabs Healthcare Qube Compact Monitor (91390Spacelabs Medical, Inc.20120
K112962Spacelabs Healthcare Xprezzon Beside Monitor (9913930)Spacelabs Medical, Inc.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 26, 2014Class IIZ-2707-2014Environmental controlThe clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional…
Jul 8, 2011Not classifiedZ-3032-2011Device DesignThe clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally.
Mar 24, 2011Not classifiedZ-2134-2011OtherThe circuit that supplies backup power to the modules in the Medical Patient Monitor, Model 91387, is not working. If there is a power interruption, the three-minute backup may not be operational and customers may lose any changes they…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 2 companies; a new generation was cleared every 13.2 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K160667 (Anybattery, Inc., 2017) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K972282.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

Spacelabs Healthcare patient monitors, functioning as either bedside or central monitors; passively display data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the model 91367, 91369, 91370 and 91387, when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and…

Read the full statement (267 words)

Spacelabs Healthcare patient monitors, functioning as either bedside or central monitors; passively display data generated by Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the model 91367, 91369, 91370 and 91387, when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, SpO 2 , temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed.

Spacelabs Healthcare patient monitors are intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature.

Spacelabs Healthcare patient monitors may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitors allow network-based applications to open windows and display information on other networked monitors.

Spacelabs Healthcare patient monitors are also designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders.

Spacelabs Healthcare patient monitors are intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102422: Spacelabs Medical Patient Monitors.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102422

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026