510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Spacelabs Healthcare Qube Compact Monitor (91390

K120616510(k) SpecialCleared Mar 29, 201229-day reviewClass II21 CFR 870.1025Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spacelabs Medical, Inc. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Mar 29, 2012 as substantially equivalent to K102422 (Spacelabs Medical, Inc., 2010) after a 29-day review. Its family tree reaches back 3 generations to 1984. It has been recalled 9 times.

Decided
Mar 29, 2012
Review
29 days
Class
Class II
Product code
MHX
Recalls
9
Predicates
1

Its longest line of descent runs back 3 generations to K842616 (1984). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK842616 Spacelabs Alpha PC Patient ComputerSpacelabs, Inc. · 1984 · Recalled
  2. 2 generations backK972282 Spacelabs Medical Pcis Patient MonitorsSpacelabs Medical, Inc. · 1997 · Recalled
  3. 1 generation backK102422 Spacelabs Medical Patient MonitorsSpacelabs Medical, Inc. · 2010 · Recalled
  4. This deviceK120616 Spacelabs Healthcare Qube Compact Monitor (91390Spacelabs Medical, Inc. · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K842616Spacelabs Alpha PC Patient ComputerSpacelabs, Inc.19842
−2K972282Spacelabs Medical Pcis Patient MonitorsSpacelabs Medical, Inc.19974
−1K102422Spacelabs Medical Patient MonitorsSpacelabs Medical, Inc.20103
0 (this device)K120616Spacelabs Healthcare Qube Compact Monitor (91390Spacelabs Medical, Inc.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
FIG. 1 — Lineage of K120616, 1984–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K102422Spacelabs Medical Patient Monitors
Summary page 1
…edicate Predicate EquivalentDevices: 310(0) Number Manufacturer/ Model Spacelabs H-ealthcare K102422 Spacelabs Medical Qube Compact Monitor Patient Monitors,…
Spacelabs Medical, Inc.20103

Also cleared against K102422: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 5, 2025Class IIZ-0169-2026Device DesignDue to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically…
Sep 5, 2025Class IIZ-0167-2026Device DesignDue to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically…
Sep 5, 2025Class IIZ-0168-2026Device DesignDue to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically…
May 8, 2015Class IIZ-1608-2015Software designThe input power can fail even though the power source at the wall has not been interrupted. In addition, there is the potential for the qube Compact Monitor to fail to turn on. This is a retrospective recall related to a recall initiated…
Jan 28, 2015Class IIZ-1145-2015Device DesignSpacelabs Healthcare qube Bedside Monitors, Model 91390, are recalled because the firm has received multiple reports of qube monitors failing to boot up or returning to factory default configuration settings following power on or reset.
Jun 6, 2014Class IIZ-1852-2014Component design/selectionSpacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing network connection after switching from hardwired Ethernet connection to using the Qube Docking Station without powering down the…
Oct 17, 2013Class IIZ-0188-2014Nonconforming Material/ComponentSpacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors.
Oct 17, 2013Class IIZ-0351-2014Nonconforming Material/ComponentSpacelabs Healthcare qube Compact Patient Monitor, Model 91390, may fail to power ON due to incorrect programmed microcontroller component.
Oct 19, 2012Class IIZ-0403-2013Nonconforming Material/ComponentSpacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 2 companies; a new generation was cleared every 12.9 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K102422.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by Spacelabs Heatthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood…

Read the full statement (258 words)

The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by Spacelabs Heatthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECG, respiration, invasive and noninvasive blood pressure, SpO 2 , temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed.

The Qube is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature.

The Qube may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitors allow networkbased applications to open windows and display information on other networked monitors.

The Qube is also designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders.

The Qube is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K120616: Spacelabs Healthcare Qube Compact Monitor (91390.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K120616

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026