510(k) · Cleared
Recalled · Class II · 2016Vantage Titan 3T
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Toshibamedical Systems Corporation · LNH System, Nuclear Magnetic Resonance Imaging
Cleared on Oct 23, 2015 as substantially equivalent to K143008 (Toshibamedical Systems Corporation, 2015) after a 63-day review. Its family tree reaches back 11 generations to 2000. 32 later devices descend from it. It has been recalled once.
- Decided
- Oct 23, 2015
- Review
- 63 days
- Class
- Class II
- Product code
- LNH
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 11 generations to K002531 (2000). 32 later devices descend from it, 2 citing it directly.
- Root · 11 generations backK002531 Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc. · 2000 · Recalled
- 7 generations between
- 3 generations backK120487 Vantage Titan 3TToshibamedical Systems Corporation · 2012
- 2 generations backK132160 Vantage Titan 3TToshibamedical Systems Corporation · 2013
- 1 generation backK143008 Vantage Titan 3TToshibamedical Systems Corporation · 2015
- This deviceK152371 Vantage Titan 3TToshibamedical Systems Corporation · 2015
- Descendant · cites it directly · 31 of its ownK162183 Vantage Galan 3T, MRT-3020, V4.0Toshibamedical Systems Corporation · 2016
- Descendant · 2 generations on · 16 of its ownK170412 Vantage Titan 1.5T, MRT-1510, M-Power GXToshibamedical Systems Corporation · 2017 · Recalled
- Descendant · 3 generations on · 14 of its ownK182282 Vantage Orian 1.5T, MRT-1550, V4.5Canon Medical Systems Corporation · 2018
- Descendant · 2 generations on · 13 of its ownK173382 Vantage Galan 3T, V4.6Toshibamedical Systems Corporation · 2018
- Descendant · 4 generations on · 12 of its ownK193021 Vantage Orian 1.5T, MRT-1550, V6.0Canon Medical Systems Corporation · 2020
Show all 11 generations
- Root · 11 generations backK002531 Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc. · 2000 · Recalled
- 10 generations backK023511 Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2Toshiba America Medical Systems, In.C · 2002 · Recalled
- 9 generations backK032490 Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models…Toshiba America Medical Systems, In.C · 2003 · Recalled
- 8 generations backK060003 Excelart Vantage ZGVToshibamedical Systems Corporation · 2006 · Recalled
- 7 generations backK063361 Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C · 2006 · Recalled
- 6 generations backK080038 Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C · 2008 · Recalled
- 5 generations backK102489 Vantage Titan 3TToshiba Medical Systems Corporation, Japan · 2011 · Recalled
- 4 generations backK113093 Vantage Titan 3TToshibamedical Systems Corporation · 2012 · Recalled
- 3 generations backK120487 Vantage Titan 3TToshibamedical Systems Corporation · 2012
- 2 generations backK132160 Vantage Titan 3TToshibamedical Systems Corporation · 2013
- 1 generation backK143008 Vantage Titan 3TToshibamedical Systems Corporation · 2015
- This deviceK152371 Vantage Titan 3TToshibamedical Systems Corporation · 2015
Showing 5 of 32 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 32 · See all- K162183Vantage Galan 3T, MRT-3020, V4.0Toshibamedical Systems Corporation · 2016 · cites it directly31descendants
- K170412Vantage Titan 1.5T, MRT-1510, M-Power GXToshibamedical Systems Corporation · 2017 · 2 generations on16descendants
- K182282Vantage Orian 1.5T, MRT-1550, V4.5Canon Medical Systems Corporation · 2018 · 3 generations on14descendants
- K173382Vantage Galan 3T, V4.6Toshibamedical Systems Corporation · 2018 · 2 generations on13descendants
- K193021Vantage Orian 1.5T, MRT-1550, V6.0Canon Medical Systems Corporation · 2020 · 4 generations on12descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −11 | K002531 | Excelart with Pianissimo, Model MRT-1501/P3 | Toshiba America Mri, Inc. | 2000 | 4 |
| −10 | K023511 | Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2 | Toshiba America Medical Systems, In.C | 2002 | 2 |
| −9 | K032490 | Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and… | Toshiba America Medical Systems, In.C | 2003 | 3 |
| −8 | K060003 | Excelart Vantage ZGV | Toshibamedical Systems Corporation | 2006 | 1 |
| −7 | K063361 | Excelart Vantage Atlas, Model MRT-1503 | Toshiba America Medical Systems, In.C | 2006 | 1 |
| −6 | K080038 | Vantage Titan, Model MRT-1504/S3 | Toshiba America Medical Systems, In.C | 2008 | 2 |
| −5 | K102489 | Vantage Titan 3T | Toshiba Medical Systems Corporation, Japan | 2011 | 1 |
| −4 | K113093 | Vantage Titan 3T | Toshibamedical Systems Corporation | 2012 | 1 |
| −3 | K120487 | Vantage Titan 3T | Toshibamedical Systems Corporation | 2012 | — |
| −2 | K132160 | Vantage Titan 3T | Toshibamedical Systems Corporation | 2013 | — |
| −1 | K143008 | Vantage Titan 3T | Toshibamedical Systems Corporation | 2015 | — |
| 0 (this device) | K152371 | Vantage Titan 3T | Toshibamedical Systems Corporation | 2015 | — |
| +1 | K162183 | Vantage Galan 3T, MRT-3020, V4.0 | Toshibamedical Systems Corporation | 2016 | — |
| +2 | K170412 | Vantage Titan 1.5T, MRT-1510, M-Power GX | Toshibamedical Systems Corporation | 2017 | 2 |
| +3 | K182282 | Vantage Orian 1.5T, MRT-1550, V4.5 | Canon Medical Systems Corporation | 2018 | — |
| +2 | K173382 | Vantage Galan 3T, V4.6 | Toshibamedical Systems Corporation | 2018 | — |
| +4 | K193021 | Vantage Orian 1.5T, MRT-1550, V6.0 | Canon Medical Systems Corporation | 2020 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2016 | 1 | 0 |
| 2017 | 2 | 0 |
| 2018 | 3 | 0 |
| 2019 | 2 | 0 |
| 2020 | 6 | 0 |
| 2021 | 8 | 0 |
| 2022 | 4 | 0 |
| 2023 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LNH device
- #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K143008 | Vantage Titan 3TSummary page 5…, MRT-3010, V3.50 SOFTWARE 7. DEVICE DESCRIPTION: The Vantage Titan 3T (Model MRT-3010) is a 3 Tesla Magnetic Resonance Imaging (MRI) System and was cleared under K143008. This submission will include the following software functionalities: UTE (Ultra Short TE), MP-RAGE, T2:mEcho, T2*:mEcho, I… | 2015 |
What descends from it? (32)
See all 322 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 14, 2016 | Class II | Z-2135-2016 | Device Design | A potential problem associated to the gradient coil (ASGC) in the Titan 3T MRI System. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 1.2 years on median.
- 8 of the 11 ancestors have a recall of any class.
- Of its descendants, K162183 (Toshibamedical Systems Corporation, 2016) has gone furthest, with 31 descendants of its own.
- 2 other devices also cite its predicate K143008.
How does its category compare?
All LNH devicesLNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · LNH
26in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LNH
122in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LNH
8in 2025
Class I (8 since 2007) Class II/III not classified
Adverse event reports per year · LNH
1,263in 2025
MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 35 | 25 | 114 |
| 2008 | 19 | 53 | 12 | 167 |
| 2009 | 17 | 70 | 7 | 194 |
| 2010 | 12 | 94 | 17 | 165 |
| 2011 | 16 | 81 | 11 | 182 |
| 2012 | 22 | 89 | 5 | 170 |
| 2013 | 19 | 108 | 2 | 107 |
| 2014 | 17 | 108 | 13 | 119 |
| 2015 | 19 | 140 | 12 (2) | 113 |
| 2016 | 20 | 108 | 1 | 135 |
| 2017 | 18 | 98 | 0 | 121 |
| 2018 | 29 | 128 | 71 | 158 |
| 2019 | 26 | 62 | 5 | 158 |
| 2020 | 32 | 121 | 4 | 98 |
| 2021 | 27 | 87 | 41 | 84 |
| 2022 | 30 | 74 | 13 | 138 |
| 2023 | 30 | 124 | 10 (1) | 159 |
| 2024 | 26 | 104 | 4 | 281 |
| 2025 | 26 | 122 | 8 (5) | 1,263 |
| 2026 | 24 | 126 | 43 | — |
What is it indicated for?
Vantage Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The…
Read the full statement (122 words)
Vantage Titan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
• Proton density (PD) (also called hydrogen density)
• Spin-lattice relaxation time (T1)
• Spin-spin relaxation time (T2)
• Flow dynamics
• Chemical Shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K152371 · printed Nov 5, 2026
Cite this page
510(k) Report. “K152371: Vantage Titan 3T.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K152371
Report an extraction error