510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Optima XR240amx

K162990510(k) SpecialCleared Nov 15, 201619-day reviewClass II21 CFR 892.1720Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · IZL System, X-Ray, Mobile

Cleared on Nov 15, 2016 as substantially equivalent to K142383 (Ge Medical Systems, LLC, 2014) after a 19-day review. Its family tree reaches back 6 generations to 1994. 3 later devices descend from it. It has been recalled 2 times.

Decided
Nov 15, 2016
Review
19 days
Class
Class II
Product code
IZL
Recalls
2
Predicates
1

Its longest line of descent runs back 6 generations to K931071 (1994). 13 ancestors in all, the oldest K802047 (1980). 3 later devices descend from it, 1 citing it directly.

  1. Root · 6 generations backK931071 Philips ThoravisionPhilips Medical Systems, Inc. · 1994 · Recalled
  2. 5 generations backK982795 Philips Bucky VisionPhilips Medical Systems, Inc. · 1998 · Recalled
  3. 4 generations backK090625 Wireless Portable Detector FD-W17Philips Medical Systems North America Co. · 2009 · Recalled
  4. 3 generations backK102615 Wireless DR Imaging Option-WDR1Ge Medical Systems, LLC · 2010 · Recalled
  5. 2 generations backK103476 BRIVOXR285AMX, Optima XR200AMX, OPTIMAXR220AMXGe Medical Systems, LLC · 2011 · Recalled
  6. 1 generation backK142383 Optima XR200amx, Optima XR220amxGe Medical Systems, LLC · 2014 · Recalled
  7. This deviceK162990 Optima XR240amxGe Medical Systems, LLC · 2016
  8. Descendant · cites it directly · 2 of its ownK173602 Optima XR240amx, AutoGridGe Medical Systems, LLC · 2018 · Recalled
  9. Descendant · 2 generations on · 1 of its ownK182234 Optima XR240amxGe Medical Systems, LLC · 2018 · Recalled
  10. Descendant · 3 generations on · 0 of its ownK210982 AMX NavigateGe Medical Systems, LLC · 2021 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K173602Optima XR240amx, AutoGridGe Medical Systems, LLC · 2018 · cites it directly2descendants
  • K182234Optima XR240amxGe Medical Systems, LLC · 2018 · 2 generations on1descendant
  • K210982AMX NavigateGe Medical Systems, LLC · 2021 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K931071Philips ThoravisionPhilips Medical Systems, Inc.19941
−5K982795Philips Bucky VisionPhilips Medical Systems, Inc.199817
−4K090625Wireless Portable Detector FD-W17Philips Medical Systems North America Co.200928
−3K102615Wireless DR Imaging Option-WDR1Ge Medical Systems, LLC20101
−2K103476BRIVOXR285AMX, Optima XR200AMX, OPTIMAXR220AMXGe Medical Systems, LLC20118
−1K142383Optima XR200amx, Optima XR220amxGe Medical Systems, LLC20148
0 (this device)K162990Optima XR240amxGe Medical Systems, LLC2016—
+1K173602Optima XR240amx, AutoGridGe Medical Systems, LLC20182
+2K182234Optima XR240amxGe Medical Systems, LLC20182
+3K210982AMX NavigateGe Medical Systems, LLC20211

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198610
199410
199710
199810
200210
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201820
202110
FIG. 1 — Lineage of K162990, 1994–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IZL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IZL device

  1. #1 K242015 TOPAZ Mobile X-ray System (Models : TOPAZ-32D…score 19.9 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K142383Optima XR200amx, Optima XR220amx
Summary page 6
…CFR 892.1720 and Product Code: Classification: Class II Product Codes: IZL, MQB Predicate Optima XR200amx and Optima XR220amx (K142383) Device(s): Regulation Name: Mobile X-Ray System Regulation: 21 CFR 892.1720…
Ge Medical Systems, LLC20148
†K161942PerkinElmer, XRpad2 3025 HWC-M Flat Panel DetectorPerkinelmer, Inc.2016—
†K161966PerkinElmer XRpad2 4336 HWC-M Flat Panel DetectorPerkinelmer, Inc.2016—

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173602Optima XR240amx, AutoGridGe Medical Systems, LLC20182

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 5, 2026Class IIZ-2728-2026Process controlGE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the…
Nov 25, 2024Class IIZ-0813-2025Component design/selectionGE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 3 companies; a new generation was cleared every 3.3 years on median.
  • 9 of the 13 ancestors have a recall of any class.
  • Of its descendants, K173602 (Ge Medical Systems, LLC, 2018) has gone furthest, with 2 descendants of its own.

How does its category compare?

All IZL devices

IZL System, X-Ray, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · IZL

11in 2025

2007: 32008: 42009: 32010: 32011: 62012: 32014: 72015: 42016: 62017: 72018: 82019: 92020: 92021: 132022: 162023: 32024: 92025: 112026 (partial year): 6
2007high 16 (2022) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IZL

153in 2025

2007: 30 days median2008: 49 days median2009: 125 days median2010: 65 days median2011: 66 days median2012: 83 days median2014: 104 days median2015: 88 days median2016: 47 days median2017: 45 days median2018: 47 days median2019: 33 days median2020: 43 days median2021: 48 days median2022: 89 days median2023: 55 days median2024: 159 days median2025: 153 days median2026 (partial year): 157 days median
2007high 159 (2024) · low 30 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IZL

47in 2025

2007: 72010: 42011: 62012: 52013: 102014: 42015: 42019: 22020: 12022: 12023: 62024: 52025: 472026 (partial year): 5
2007high 47 (2025) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IZL

27in 2025

2007: 162008: 272009: 352010: 752011: 1182012: 242013: 172014: 512015: 302016: 332017: 352018: 272019: 402020: 482021: 472022: 322023: 772024: 392025: 27
2007high 118 (2011) · low 16 (2007)2025

MAUDE reports naming product code IZL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007330716
2008449027
20093125035
2010365475
20116666118
2012383524
20130—1017
20147104451
2015488430
2016647033
2017745035
2018847027
2019933240
2020943148
20211348047
20221689132
2023355677
20249159539
2025111534727
202661575—

What is it indicated for?

The Optima XR240amx is intended to take exposures utilizing film, computed radiography (CR), or wireless detectors, which are intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR). Optima XR240amx is a self-contained; battery operated mobile radiographic imaging system designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect…

Read the full statement (212 words)

The Optima XR240amx is intended to take exposures utilizing film, computed radiography (CR), or wireless detectors, which are intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).

Optima XR240amx is a self-contained; battery operated mobile radiographic imaging system designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects).

The Optima XR240amx is indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room.

The system is indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position.

This device is not intended for mammographic applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K162990: Optima XR240amx.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162990

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026