510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Optima CT660

K110227510(k) SpecialCleared Jul 20, 2011176-day reviewClass II21 CFR 892.1750Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · JAK System, X-Ray, Tomography, Computed

Cleared on Jul 20, 2011 as substantially equivalent to K082761 (Ge Medical Systems, LLC, 2008) after a 176-day review. Its family tree reaches back 5 generations to 2001. 6 later devices descend from it. It has been recalled 5 times.

Decided
Jul 20, 2011
Review
176 days
Class
Class II
Product code
JAK
Recalls
5
Predicates
1

Its longest line of descent runs back 5 generations to K013561 (2001). 6 later devices descend from it, 1 citing it directly.

  1. Root · 5 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
  2. 4 generations backK030420 Lightspeed 5.0 Computed Tomography SystemGE Medical Systems · 2003 · Recalled
  3. 3 generations backK040372 Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC · 2004 · Recalled
  4. 2 generations backK061817 Lightspeed 7.1 CT Scanner SystemGe Medical Systems, LLC · 2006 · Recalled
  5. 1 generation backK082761 GE Lightspeed 7.2 CT Scanner SystemGe Medical Systems, LLC · 2008 · Recalled
  6. This deviceK110227 Optima CT660Ge Medical Systems, LLC · 2011
  7. Descendant · cites it directly · 5 of its ownK131576 Optima CT660Ge Medical Systems, LLC · 2013 · Recalled
  8. Descendant · 2 generations on · 4 of its ownK192686 Revolution Maxima, Revolution AceGe Hangwei Medical Systems Co., Ltd. · 2019 · Recalled
  9. Descendant · 3 generations on · 3 of its ownK203169 Revolution AscendGe Healthcare Japan Corporation · 2020 · Recalled
  10. Descendant · 4 generations on · 2 of its ownK213938 Revolution AscendGe Healthcare Japan Corporation · 2022 · Recalled
  11. Descendant · 5 generations on · 0 of its ownK233698 True Enhance DLGe Healthcare Japan Corporation · 2024

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2013: 12019: 12020: 12022: 12024: 2
2013high 2 (2024) · low 0 (2014)2026

Most influential descendants

Showing 5 of 6 · See all
  • K131576Optima CT660Ge Medical Systems, LLC · 2013 · cites it directly5descendants
  • K192686Revolution Maxima, Revolution AceGe Hangwei Medical Systems Co., Ltd. · 2019 · 2 generations on4descendants
  • K203169Revolution AscendGe Healthcare Japan Corporation · 2020 · 3 generations on3descendants
  • K213938Revolution AscendGe Healthcare Japan Corporation · 2022 · 4 generations on2descendants
  • K233698True Enhance DLGe Healthcare Japan Corporation · 2024 · 5 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K013561Lightspeed 4.0 CT Scanner SystemGE Medical Systems20015
−4K030420Lightspeed 5.0 Computed Tomography SystemGE Medical Systems20034
−3K040372Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC20045
−2K061817Lightspeed 7.1 CT Scanner SystemGe Medical Systems, LLC20061
−1K082761GE Lightspeed 7.2 CT Scanner SystemGe Medical Systems, LLC20083
0 (this device)K110227Optima CT660Ge Medical Systems, LLC2011—
+1K131576Optima CT660Ge Medical Systems, LLC20138
+2K192686Revolution Maxima, Revolution AceGe Hangwei Medical Systems Co., Ltd.20195
+3K203169Revolution AscendGe Healthcare Japan Corporation20201
+4K213938Revolution AscendGe Healthcare Japan Corporation20222
+5K233698True Enhance DLGe Healthcare Japan Corporation2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201910
202010
202210
202420
FIG. 1 — Lineage of K110227, 2001–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082761GE Lightspeed 7.2 CT Scanner System
Summary page 2
…dified device. The system is labeled as the Optima CT660 however the commercial names used for the configurations ore Optima CT66O or Optima CT660 s. Predicate Device: K082761 - LightSpeed VCT (GE LightSpeed 7.2 CT Scanner System) Device Description: The Optima CT660 and its associated config urations/…
Ge Medical Systems, LLC20083

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131576Optima CT660Ge Medical Systems, LLC20135

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 24, 2018Class IIZ-2583-2018Component change controlSome CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Apr 27, 2018Class IIZ-2873-2018Employee errorGE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
Apr 27, 2018Class IIZ-2871-2018Employee errorGE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.
May 19, 2017Class IIZ-2182-2017Software designIt was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.
Mar 26, 2013Class IIZ-0023-2014Software design (manufacturing process)GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT 710. The DMPR and…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 2.2 years on median.
  • 5 of the 5 ancestors have a recall of any class.
  • Of its descendants, K131576 (Ge Medical Systems, LLC, 2013) has gone furthest, with 5 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The Optima CT660 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data…

Read the full statement (137 words)

The Optima CT660 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.

The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K110227: Optima CT660.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K110227

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026