510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Carestream Pacs

K110919510(k) TraditionalCleared Oct 20, 2011202-day reviewClass II21 CFR 892.2050Recalls 50

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Carestream Health, Inc. · LLZ System, Image Processing, Radiological

Cleared on Oct 20, 2011 as substantially equivalent to K083673 (Carestream Health, Inc., 2008) after a 202-day review. Its family tree reaches back 5 generations to 1996. 8 later devices descend from it. It has been recalled 50 times.

Decided
Oct 20, 2011
Review
202 days
Class
Class II
Product code
LLZ
Recalls
50
Predicates
1

Its longest line of descent runs back 5 generations to K954678 (1996). 8 ancestors in all, the oldest K941223 (1994). 8 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1998: 1 other ancestor1994: 1 other ancestor2024: 1 descendant cleared2014: 1 descendant cleared2018: 1 descendant cleared2015: 1 descendant cleared2017: 1 descendant cleared2025: 1 descendant cleared2020: 1 descendant cleared2023: 1 descendant clearedK954678 · Pro Vision · Algotec Systems, Ltd. · 1996K980648 · Pro Vision · Algotec Systems, Ltd. · 1998K992654 · Plug 'N View 3D, Version 1.0 · Voxar Limited · 1999K020546 · Fusion 7D · Mirada Solutions, Ltd. · 2002K083673 · Carestream Pacs, Model Carestream Pacs Vii · Carestream Health, Inc. · 20081990200020102020
Other ancestors, by generationDescendants, by generation 3 generations back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 2 descendants5 generations forward: 1 descendant K083673 · Carestream Pacs, Model Carestream Pacs Vii · Carestream Health, Inc. · 2008K020546 · Fusion 7D · Mirada Solutions, Ltd. · 2002K992654 · Plug 'N View 3D, Version 1.0 · Voxar Limited · 1999K980648 · Pro Vision · Algotec Systems, Ltd. · 1998K954678 · Pro Vision · Algotec Systems, Ltd. · 1996−50+5
  1. Root · 5 generations backK954678 Pro VisionAlgotec Systems, Ltd. · 1996
  2. 4 generations backK980648 Pro VisionAlgotec Systems, Ltd. · 1998
  3. 3 generations backK992654 Plug 'N View 3D, Version 1.0Voxar Limited · 1999
  4. 2 generations backK020546 Fusion 7DMirada Solutions, Ltd. · 2002
  5. 1 generation backK083673 Carestream Pacs, Model Carestream Pacs ViiCarestream Health, Inc. · 2008
  6. This deviceK110919 Carestream PacsCarestream Health, Inc. · 2011
  7. Descendant · cites it directly · 6 of its ownK132824 Carestream Vue Pacs Version 11.4 Carestream Vue Motion ApplicationCarestream Health, Inc. · 2014
  8. Descendant · 2 generations on · 4 of its ownK151774 CARESTREAM Vue MotionCarestream Health, Inc. · 2015 · Recalled
  9. Descendant · 3 generations on · 3 of its ownK170434 Enterprise Imaging XERO Viewer 8.1Agfa Healthcare N.V. · 2017 · Recalled
  10. Descendant · 4 generations on · 1 of its ownK200822 BraidBraid.Health · 2020
  11. Descendant · 4 generations on · 0 of its ownK252199 AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V. · 2025

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2014: 12015: 12017: 12018: 12020: 12023: 12024: 12025: 1
2014high 1 (2014) · low 0 (2016)2026

Most influential descendants

Showing 5 of 8 · See all
  • K132824Carestream Vue Pacs Version 11.4 Carestream Vue Motion ApplicationCarestream Health, Inc. · 2014 · cites it directly6descendants
  • K151774CARESTREAM Vue MotionCarestream Health, Inc. · 2015 · 2 generations on4descendants
  • K170434Enterprise Imaging XERO Viewer 8.1Agfa Healthcare N.V. · 2017 · 3 generations on3descendants
  • K200822BraidBraid.Health · 2020 · 4 generations on1descendant
  • K252199AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V. · 2025 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K954678Pro VisionAlgotec Systems, Ltd.1996—
−4K980648Pro VisionAlgotec Systems, Ltd.1998—
−3K992654Plug 'N View 3D, Version 1.0Voxar Limited1999—
−2K020546Fusion 7DMirada Solutions, Ltd.2002—
−1K083673Carestream Pacs, Model Carestream Pacs ViiCarestream Health, Inc.2008—
0 (this device)K110919Carestream PacsCarestream Health, Inc.2011—
+1K132824Carestream Vue Pacs Version 11.4 Carestream Vue Motion ApplicationCarestream Health, Inc.2014—
+2K151774CARESTREAM Vue MotionCarestream Health, Inc.20151
+3K170434Enterprise Imaging XERO Viewer 8.1Agfa Healthcare N.V.20172
+4K200822BraidBraid.Health2020—
+4K252199AGFA HealthCare Enterprise ImagingAgfa Healthcare N.V.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199810
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201510
201710
201810
202010
202310
202410
202510
FIG. 1 — Lineage of K110919, 1996–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K083673Carestream Pacs, Model Carestream Pacs Vii
Summary page 1
…vice: Carestreamn PACS Manufactured by Carestream Health, Inc. 5 10(k) No. - K083673 (December 30, 2008) TITLE: 510(k) Summary PART t4:6J5396 VERSION # 210 Unrestricted in…
Carestream Health, Inc.2008—

Also cleared against K083673: 1 other device →

What descends from it? (8)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132824Carestream Vue Pacs Version 11.4 Carestream Vue Motion ApplicationCarestream Health, Inc.20146
K240822Image Management V15Philips Medical Systems Technologies20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 10, 2024Class IIZ-0497-2025Software designA software issue affecting Philips Vue PACS may cause incorrect ischemic map and table value calculations when using the export function in the Perfusion Application for a Siemens CT perfusion study, resulting in misdiagnosis due to…
Sep 16, 2013Class IIZ-0248-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0229-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0255-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0250-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0228-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0243-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0214-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0216-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0218-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0222-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0240-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0224-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0259-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0211-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0236-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0238-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0251-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0235-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0257-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0258-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0220-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0237-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0217-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0223-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0225-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0239-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0221-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0242-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0246-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0252-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0226-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0215-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0232-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0230-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0233-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0241-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0212-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0249-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0244-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0253-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0219-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0256-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0254-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0231-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0213-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0227-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0245-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0260-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…
Sep 16, 2013Class IIZ-0234-2014Software designCarestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 2.5 years on median.
  • 1 of the 8 ancestors has a recall of any class.
  • Of its descendants, K132824 (Carestream Health, Inc., 2014) has gone furthest, with 6 descendants of its own.
  • 1 other device also cites its predicate K083673.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems; The system contains interactive tools in order to ease the process of analyzing and comparing three dimensional (31D) images. It is a single system…

Read the full statement (193 words)

The CARESTREAM PACS is an image management system whose intended use is

to provide completely scaleable local and wide area PACS solutions for hospital and

related institutions/sites, which will archive, distribute, retrieve and display images

and data from all hospital modalities and information systems;

The system contains interactive tools in order to ease the process of analyzing and

comparing three dimensional (31D) images. It is a single system that integrates the

review, dictation and reporting tools that creates a productive work environment for

the radiologists and physicians.

The CARESTREAM PACS Lightweight Viewer software program is used for patient

management by the referral community in order to access and display patient data,

medical reports, and medical images from different modalities including CR, DR, CT,

MR, NM and US after the primary reading has been completed on dedicated

diagnostic workstations,

The CARESTREAM PACS Lightweight Viewer provides wireless and portable

access to medical images for referral purposes. It is not intended to be used as, or

to replace, a full diagnostic workstation or system and should be used only when

there is no access to a workstation. This device is not to be used for mammography.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K110919: Carestream Pacs.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K110919

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026