Recalled · Class II · 2013Fresenius 2008T Hemodialysis Machine
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Fresenius Medical Care · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Cleared on Oct 21, 2011 as substantially equivalent to K093902 (Fresenius Medical Care North America, 2010) after a 130-day review. Its family tree reaches back 4 generations to 1994. 1 of the 24 devices in its family, counting itself, has a Class I recall. 19 later devices descend from it. It has been recalled once.
- Decided
- Oct 21, 2011
- Review
- 130 days
- Class
- Class II
- Product code
- KDI
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 4 generations to K921456 (1994). 19 later devices descend from it, 1 citing it directly. 1 of the 24 devices in this family, counting this one, has a Class I recall.
- Root · 4 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
- 3 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
- 2 generations backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
- 1 generation backK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010 · Recalled
- This deviceK111639 Fresenius 2008T Hemodialysis MachineFresenius Medical Care · 2011
- Descendant · cites it directly · 18 of its ownK113427 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America · 2011 · Recalled
- Descendant · 2 generations on · 17 of its ownK120505 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente · 2012 · Recalled
- Descendant · 3 generations on · 16 of its ownK140866 Tablo ConsoleHome Dialysis Plus · 2014
- Descendant · 4 generations on · 11 of its ownK150880 Tablo Console with Wireless CapabilityOutset Medical · 2015
- Descendant · 5 generations on · 10 of its ownK160881 Tablo ConsoleOutset Medical, Inc. · 2016
Showing 5 of 19 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K233335 Tablo® Hemodialysis System9 generations on · 2023
Most influential descendants
Showing 5 of 19 · See all- K113427Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America · 2011 · cites it directly18descendants
- K120505Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente · 2012 · 2 generations on17descendants
- K140866Tablo ConsoleHome Dialysis Plus · 2014 · 3 generations on16descendants
- K150880Tablo Console with Wireless CapabilityOutset Medical · 2015 · 4 generations on11descendants
- K160881Tablo ConsoleOutset Medical, Inc. · 2016 · 5 generations on10descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K921456 | Fresenius 3008/FDS-08 Dialysate Delivery System | Fresenius USA, Inc. | 1994 | 3 |
| −3 | K994267 | Fresenius 2008K | Fresenius Medical Care | 2000 | 5 |
| −2 | K080964 | Fresenius Model 2008T Hemodialysis Machine | Fresenius Medical Care North America | 2009 | 4 |
| −1 | K093902 | Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573 | Fresenius Medical Care North America | 2010 | 2 |
| 0 (this device) | K111639 | Fresenius 2008T Hemodialysis Machine | Fresenius Medical Care | 2011 | — |
| +1 | K113427 | Fresenius 2008T Hemodialysis Machine | Fresenius Medical Care North America | 2011 | 4 |
| +2 | K120505 | Fresenius 2008T Hemodialysis Machine | Fresenius Medical Care North America, Design Cente | 2012 | 4 |
| +3 | K140866 | Tablo Console | Home Dialysis Plus | 2014 | — |
| +4 | K150880 | Tablo Console with Wireless Capability | Outset Medical | 2015 | — |
| +5 | K160881 | Tablo Console | Outset Medical, Inc. | 2016 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 2 | 0 |
| 2021 | 2 | 0 |
| 2022 | 2 | 0 |
| 2023 | 2 | 1 |
| 2024 | 4 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KDI device
- #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K093902 | Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Summary page 1…fication Name: Class 11 per § 876.5860 KC. Legally Marketed Predicate Device (unmodified device) Fresenius 2008T Hemodialysis Machine (K093902). L. Device Description The Eresenius 2008T Hemodialysis Machine (K093902) is indicated for acute and chronic… | 2010 |
What descends from it? (19)
See all 191 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K113427 | Fresenius 2008T Hemodialysis Machine | 2011 | 18 |
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 24, 2013 | Class II | Z-1693-2013 | Nonconforming Material/Component | Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber) |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K233335Tablo® Hemodialysis SystemOutset Medical, Inc. · 2023 · 9 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 5.9 years on median.
- 4 of the 4 ancestors have a recall of any class.
- Of its descendants, K113427 (Fresenius Medical Care North America, 2011) has gone furthest, with 18 descendants of its own.
- 1 other device also cites its predicate K093902.
How does its category compare?
All KDI devicesKDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · KDI
9in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KDI
199in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KDI
2in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · KDI
2,786in 2025
MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 11 | 91 | 4 | 916 |
| 2008 | 6 | 148 | 0 | 1,042 |
| 2009 | 4 | 236 | 3 | 522 |
| 2010 | 9 | 186 | 1 | 823 |
| 2011 | 5 | 85 | 4 | 2,270 |
| 2012 | 6 | 153 | 1 | 1,765 |
| 2013 | 11 | 247 | 2 | 2,360 |
| 2014 | 8 | 164 | 3 | 6,154 |
| 2015 | 5 | 69 | 5 | 4,182 |
| 2016 | 5 | 230 | 4 | 766 |
| 2017 | 8 | 188 | 1 | 963 |
| 2018 | 2 | — | 1 | 1,716 |
| 2019 | 4 | 263 | 6 | 1,587 |
| 2020 | 9 | 266 | 0 | 1,633 |
| 2021 | 2 | — | 2 | 1,944 |
| 2022 | 7 | 259 | 17 | 2,180 |
| 2023 | 7 | 157 | 8 (8) | 2,692 |
| 2024 | 7 | 235 | 4 (2) | 2,441 |
| 2025 | 9 | 199 | 2 | 2,786 |
| 2026 | 6 | 154 | 2 | — |
What is it indicated for?
Fresenius 2008T Hemodialysis Machine is indicated for acute and chronic dialysis therapy.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K111639 · printed Nov 6, 2026
Cite this page
510(k) Report. “K111639: Fresenius 2008T Hemodialysis Machine.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111639