510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Zoll Propaq MD

K112761510(k) TraditionalCleared Apr 25, 2012216-day reviewClass III21 CFR 870.5310Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zoll Medical Corporation, World Wide Headquarters · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Apr 25, 2012 as substantially equivalent to K100654 (Zoll Medical Corporation, World Wide Headquarters, 2010) and 2 other predicates after a 216-day review. Its family tree reaches back 10 generations to 1997. 2 of the 63 devices in its family, counting itself, have a Class I recall. It has been recalled once.

Decided
Apr 25, 2012
Review
216 days
Class
Class III
Product code
MKJ
Recalls
1
Predicates
3

Its longest line of descent runs back 10 generations to K962603 (1997). 62 ancestors in all, the oldest K893221 (1990). No later device cites it as a predicate yet. 2 of the 63 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2008: 1 other ancestor2003: 5 other ancestors2006: 1 other ancestor2005: 4 other ancestors2004: 3 other ancestors2002: 5 other ancestors, 1 with a Class I recall12001: 2 other ancestors2000: 4 other ancestors1999: 6 other ancestors1996: 3 other ancestors1995: 5 other ancestors1992: 2 other ancestors1990: 2 other ancestors1998: 4 other ancestors1997: 2 other ancestors, 1 with a Class I recall11991: 2 other ancestors1993: 1 other ancestorK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999×2: 2 generations cleared close together, K012891 (2002), K021090 (2002) K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 2005K060065 · CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor · Oridion Capnography, Inc. · 2006 · recalledK100654 · Zoll Propaq MD · Zoll Medical Corporation, World Wide Headquarters · 201019901995200020052010
Other ancestors, by generation 1 generation back: 2 other ancestors2 generations back: 6 other ancestors3 generations back: 12 other ancestors4 generations back: 18 other ancestors5 generations back: 6 other ancestors, 2 with a Class I recall26 generations back: 6 other ancestors7 generations back: 2 other ancestors K100654 · Zoll Propaq MD · Zoll Medical Corporation, World Wide Headquarters · 2010K060065 · CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter Bedside Monitor · Oridion Capnography, Inc. · 2006 · recalledK050844 · 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL, 750C-3MS, 750CM-3MS, 750C-3NL, 750CM-3NL · Cas Medical Systems, Inc. · 2005K033048 · Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL, 740M-2NL, 740-2NN, 740M-2NN, 740-3MS, 740M-3MS · Cas Medical Systems, Inc. · 2004 · recalledK021090 · Oximax N-550 Pulse Oximeter,Model N-550 Pulse Oximeter · Nellcor Puritan Bennett, Inc. · 2002K012891 · Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and Cables (Aka Accessories) · Nellcor Puritan Bennett, Inc. · 2002 · recalledK991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997−10−50
  1. Root · 10 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 6 generations between
  3. 3 generations backK050844 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL…Cas Medical Systems, Inc. · 2005
  4. 2 generations backK060065 CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One…Oridion Capnography, Inc. · 2006 · Recalled
  5. 1 generation backK100654 Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2010
  6. This deviceK112761 Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2012
Show all 10 generations
  1. Root · 10 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 9 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 8 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 7 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 6 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 5 generations backK021090 Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc. · 2002
  7. 4 generations backK033048 Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS…Cas Medical Systems, Inc. · 2004 · Recalled
  8. 3 generations backK050844 750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL…Cas Medical Systems, Inc. · 2005
  9. 2 generations backK060065 CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One…Oridion Capnography, Inc. · 2006 · Recalled
  10. 1 generation backK100654 Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2010
  11. This deviceK112761 Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters · 2012

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−9K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−8K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−7K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−6K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−5K021090Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc.2002—
−4K033048Cas 740 Series Monitor, Models 740-1, 740M-1, 740-2MS, 740M-2MS, 740-2NL…Cas Medical Systems, Inc.20041
−3K050844750C Series Monitor, Models 750C-2MS, 750CM-2MS, 750C-NNL, 750CM-2NL…Cas Medical Systems, Inc.2005—
−2K060065CAPNOSTREAM20 Two Parameter Bedside Monitor & CAPNOSTREAM10 One Parameter…Oridion Capnography, Inc.20063
−1K100654Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters2010—
0 (this device)K112761Zoll Propaq MDZoll Medical Corporation, World Wide Headquarters2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199020
199120
199220
199310
199550
199630
199721
199840
199960
200040
200120
200251
200350
200430
200540
200610
200810
FIG. 1 — Lineage of K112761, 1997–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K100654Zoll Propaq MD
Summary page 2
…ter (DSK). Carbon Dioxide Gas Analyzer (CCK) Oximeter (DQA) Description: The ZOLL Propaq MD device (reviewed and cleared by the agency under application K100654) is being revised with additional software features. The Propaq MDis alight weight,. portable device designed to be used by train…
Zoll Medical Corporation, World Wide Headquarters2010—
*K072923Zoll E Series Defibrillator
Summary page 6
…ntial Equivalence: The proposed ZOLL Propaq MD is substantially equivalent to the features and functions of the predicate units: ZOLL Propaq MD (K100654), ZOLL E Series (K072923) and Inovise Audicor 12-lead Interpretive algorithm (K032 145). Comparison of Technological Characteristics The proposed ZOLL Pi…
Zoll Medical Corporation, World Wide Headquarters2008—
*K032145Audicor Upgrade System
Summary page 6
…ly equivalent to the features and functions of the predicate units: ZOLL Propaq MD (K100654), ZOLL E Series (K072923) and Inovise Audicor 12-lead Interpretive algorithm (K032 145). Comparison of Technological Characteristics The proposed ZOLL Piopaq MD utilizes the same features and functions as the indica…
Inovise Medical, Inc.2003—

Also cleared against K100654: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 5, 2014Class IIZ-1311-2014Software designDefibrillator/Pacemaker/Monitor May Latch in a Continuous Device Reset Loop and cannot be used

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K022732Infant/Child Reduced Energy ElectrodesMedtronic Physio-Control Corp. · 2002 · 5 generations back · ◆ in the 2002 ancestors bar of FIG. 12 Class I recalls
  • K970209Capnocheck PlusBci Intl., Inc. · 1997 · 5 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 6 companies; a new generation was cleared every 1.1 years on median.
  • 14 of the 62 ancestors have a recall of any class; 2 have a Class I recall.
  • 1 other device also cites its predicate K100654.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—

What is it indicated for?

The Propaq MID is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care and the use of the Propaq MD. The Propaq MID is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an…

Read the full statement (1,215 words)

The Propaq MID is intended for use by trained medical personnel who are familiar with basic monitoring, vital sign assessment, emergency cardiac care and the use of the Propaq MD. The Propaq MID is also intended for use by (or on the order of) physicians at the scene of an emergency or in a hospital emergency room, intensive care unit, cardiac care unit, or other similar areas of a hospital. The usage may be in an ambulance or at the scene of an emergency. It is also intended to be used during the transport of patients. The Propaq MID will be used primarily on patients experiencing symptoms of cardiac arrest or in post trauma situation. It may also be used whenever it is required to monitor any of those functions that are included (as options) in the device. The Propaq MD can be used on pediatric patients (as described in the following table) and on adult patients (21 years of age or older) with and without heart dysfunction.

Pediatric Subpopulation Approx. Age Range

Newborn (neonate) Birth to 1 month of age

Infant 1 month to 2 years of age

Child 2 to 12 years of age

Adolescent 12-21 years of age

When the pediatric patient is less than 8 years of age or weighs less than 55 lbs. (25 kg.), use ZOLL pedi-padz@ pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight.

Manual Defibrillation

Use of the Propaq MID in the manual mode for external and internal defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

* Unconsciousness

* Absence of breathing

* Absence of pulse.

This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.

The unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. Qualified medical personnel must decide when synchronized cardioversion is appropriate.

The patient population will range from newborn (neonate) to adult.

Semiautomatic Operation (AED)

The Propaq MVD products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.

They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.

Use of the Propacj MVD in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

*Unconsciousness.

*Absence of breathing.

*Absence of pulse.

When the patient is less than 8 years of age or weighs less that 55 lbs. (25 Kg), use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine patient's exact age or weight.

EGG Monitoring

The Propaq MVD is intended for use to monitor and/or record 3, 5,12-Lead ECG waveform and heart rate, and to alarm when heart rate is above or below limits set by the operator. The patient population will range from newborn (neonate) to adult, with and without heart dysfunction.

External Transcutaneous Pacing

This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endlocardial stimulation.

The purposesof pacing include:

*Resuscitation from standstill or bradycardia of any etiology:

Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidline, digitalis, b-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.

As a standby when standstill or bradycardia might be expected:

Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous therapy that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endlocardial pacing.

*Suppression of tachycardia:

Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.

*Pediatric Pacing:

Pacing can be performed on pediatric patients weighing 33 lbs. (15kg) or less using ZOLL pediatric hands-free therapy electrode pads. Prolonged pacing (in excess of 30 minutes), piarticularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.

Non-Invasive Blood Pressure Monitoring

The Propaq MD is intended for use to make non-invasive measurements of arterial pressure and heart rate, and to alarm if either parameter is outside of the limits set by the user. Measurements are made using an inflatable cuff on the patients arm or leg. The patient population will range from newborn (neonate) to adult.

Temperature Monitoring

The Propaq MID is intended for use to make continuous temperature measurements of rectal, esophageal, or surface temperatures, and to alarm if the temperature is outside of the limits set by the user. The patient population will range from newborn (neonate) to adult.

5p02 Monitoring

The Propaq MID is intended for use to monitor pulse rate and oxygen saturation of arteriolar hemoglobin, and to alarm if either parameter is outside of the limits set by the user. Measurements are made non-invasively at remote sites such as a finger, toe, ear lobe, bridge of nose, etc. The patient population will range from newborn (neonate) to adult.

Respiration Monitoring

The Propaq MID is intended for use to continuously monitor respiration rate and to alarm if the rate falls outside of the range set by the operator. Because the measurement method actually measures respiratory effort, apnea episodes with continued respiratory effort (such as obstructive apnea) may not be detected. It is not intended to be used as an apnea monitor. The patient population will range from newborn (neonate) to adult.

C02 Monitoring

The Propaq MID is intended for use to make continuous noninvasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate. The patient population will range from newborn (neonate) to adult.

Invasive Pressure Monitoring

The Propaq MVD is intended for use to display and make continuous invasive pressure measurements from any compatible pressure transducer. The primary intended uses are arterial blood pressure, central venous pressure and intracranial pressure monitoring. Any contra-indications of the particular transducer selected by the user shall apply. The patient population will range from newborn (neonate) to adult.

12-Lead Analysis

The 12-lead ECG Analysis is intended for use in acquiring, analyzing and reporting ECG data, and to provide interpretation of the data for consideration by caregivers. The interpretations of ECG data offered by the device are only significant when used in conjunction with caregiver overread as well as consideration of all other relevant patient data. The 12-lead ECG Analysis is intended for use on adults (> 18 years of age).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K112761: Zoll Propaq MD.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K112761

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026