Integra Licox PTO2 Monitor

K131184510(k) TraditionalCleared Aug 30, 2013127-day reviewClass II21 CFR 882.1620Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Integra Lifesciences Corp. · GWM Device, Monitoring, Intracranial Pressure

Cleared on Aug 30, 2013 as substantially equivalent to K002765 (Integra Neurosciences, 2000) and 3 other predicates after a 127-day review. Its family tree reaches back 3 generations to 2000. It has been recalled 5 times.

Decided
Aug 30, 2013
Review
127 days
Class
Class II
Product code
GWM
Recalls
5
Predicates
4

Its longest line of descent runs back 3 generations to K002765 (2000). 13 ancestors in all, the oldest K813459 (1982). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2012: 2 other ancestors1999: 1 other ancestor1996: 1 other ancestor1992: 1 other ancestor1991: 1 other ancestor1989: 1 other ancestor1986: 1 other ancestor1984: 1 other ancestor1982: 1 other ancestorK002765 · Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB Luicox Imc Introducer Kits-IM,IM2,IM3,II1 Licoximc Tempe · Integra Neurosciences · 2000 · recalledK020558 · Licox CC1.G2 Brain Oxygen Catheter-Microprobe · Integra Neurosciences · 2002K040235 · Licox Pmo Brain Monitoring System · Integra LifeSciences Corporation · 2004 · recalled19801985199019952000200520102015
  1. Root · 3 generations backK002765 Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB…Integra Neurosciences · 2000 · Recalled
  2. 2 generations backK020558 Licox CC1.G2 Brain Oxygen Catheter-MicroprobeIntegra Neurosciences · 2002
  3. 1 generation backK040235 Licox Pmo Brain Monitoring SystemIntegra LifeSciences Corporation · 2004 · Recalled
  4. This deviceK131184 Integra Licox PTO2 MonitorIntegra Lifesciences Corp. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K002765Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB Luicox Imc…Integra Neurosciences20003
−2K020558Licox CC1.G2 Brain Oxygen Catheter-MicroprobeIntegra Neurosciences2002—
−1K040235Licox Pmo Brain Monitoring SystemIntegra LifeSciences Corporation200412
0 (this device)K131184Integra Licox PTO2 MonitorIntegra Lifesciences Corp.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198410
198610
198910
199110
199210
199610
199910
201220
FIG. 1 — Lineage of K131184, 2000–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWM device

  1. #1 K252241 Pressio 3 Multi-parameter Neuromonitoring Systemscore 12.9 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002765Licox CMP Monitor-AC3.1, Licox Imc Oxygen Catheters-CC1,CC1.SB Luicox Imc Introducer…
Summary page 1
…GWM -807.92(a)(3) - Legally marketed device(s) to which equivalence is claimed * LICOX Brain Oxygen Monitoring System, CMP® Monitor and IMC® Systems - K002765 * LICOX PMO BRAIN MONITORING SYSTEM- K040235 * INTEGRA CAM INO ICP MONITOR- K 121573 . INTEGRA CAMINO FLEX VENT…
Integra Neurosciences20003
*K040235Licox Pmo Brain Monitoring System
Summary page 1
…device(s) to which equivalence is claimed * LICOX Brain Oxygen Monitoring System, CMP® Monitor and IMC® Systems - K002765 * LICOX PMO BRAIN MONITORING SYSTEM- K040235 * INTEGRA CAM INO ICP MONITOR- K 121573 . INTEGRA CAMINO FLEX VENTRICULAR INTRACRANIAL PRESSURE MONITORING…
Integra LifeSciences Corporation200412
*K121573Integra Camino Icp Monitor
Summary page 1
…* LICOX Brain Oxygen Monitoring System, CMP® Monitor and IMC® Systems - K002765 * LICOX PMO BRAIN MONITORING SYSTEM- K040235 * INTEGRA CAM INO ICP MONITOR- K 121573 . INTEGRA CAMINO FLEX VENTRICULAR INTRACRANIAL PRESSURE MONITORING KIT WITH INTRORA CAMINO FLEX ADAPTER- K121 1…
Intrgra Lifesciences Corporation20123
*K121159Integra Camino Flex Ventricular Intracranial Pressure Monitoring Kit with Intrgra Camino…
Summary page 1
…5 * INTEGRA CAM INO ICP MONITOR- K 121573 . INTEGRA CAMINO FLEX VENTRICULAR INTRACRANIAL PRESSURE MONITORING KIT WITH INTRORA CAMINO FLEX ADAPTER- K121 159 807.92(a)(4) - Device description The Integra®i Licox® PtO2 monitor instrument is a diagnostic tool for continuously monitoring…
Integra Lifescience Corportation2012—

Also cleared against K002765: 8 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 14, 2015Class IIZ-1619-2015Device DesignIntegra LifeSciences Corporation is recalling Integra's Camino ICP Monitor (CAM 02) & Licox PtO2 Monitor ( LCX02 & LCX02R) due to complaints that the audible speaker malfunctions.
Apr 14, 2015Class IIZ-1618-2015Device DesignIntegra LifeSciences Corporation is recalling Integra's Camino ICP Monitor (CAM02) & Licox PtO2 Monitor ( LCX02 & LCX02R) due to complaints that the audible speaker malfunctions.
Apr 14, 2015Class IIZ-1620-2015Device DesignIntegra LifeSciences Corporation is recalling Integra's Camino ICP Monitor (CAM 02) & Licox PtO2 Monitor ( LCX02 & LCX02R) due to complaints that the audible speaker malfunctions.
Aug 18, 2014Class IIZ-0201-2015Device DesignIntegra LifeSciences Corporation received 2 complaints that the USB port on the Licox Pt02 monitor does not consistently provide the user the ability to extract the Pt02 trend data according to the User's Manual for the device.
Apr 16, 2014Class IIZ-1675-2014Under Investigation by firmIntegra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 2.1 years on median.
  • 6 of the 13 ancestors have a recall of any class.
  • 8 other devices also cite its predicate K002765.

How does its category compare?

All GWM devices

GWM Device, Monitoring, Intracranial Pressure: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · GWM

1in 2025

2007: 22008: 22009: 12010: 22011: 32012: 42013: 22014: 22016: 32017: 32018: 12019: 22020: 32021: 22024: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWM

30in 2020

2011: 123 days median2012: 100 days median2016: 192 days median2017: 37 days median2020: 30 days median
2007high 192 (2016) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWM

3in 2025

2008: 12010: 92011: 62012: 12013: 122014: 22019: 12023: 1, 1 Class I2024: 4, 2 Class I2025: 3
2007high 12 (2013) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · GWM

441in 2025

2007: 1262008: 1052009: 1172010: 1042011: 1842012: 1512013: 1202014: 3602015: 4862016: 4192017: 4452018: 3022019: 2602020: 2682021: 2112022: 4242023: 7442024: 6412025: 441
2007high 744 (2023) · low 104 (2010)2025

MAUDE reports naming product code GWM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—0126
20082—1105
20091—0117
20102—9104
201131236184
201241001151
20132—12120
20142—2360
20150—0486
201631920419
20173370445
20181—0302
20192—1260
20203300268
20212—0211
20220—0424
20230—1 (1)744
20242—4 (2)641
20251—3441
20261—0—

What is it indicated for?

The Integra Licox PtO 2 Monitor measures oxygen partial pressure (PtO 2) and temperature in brain tissue and these parameters are used together as an aid in the determination of the perfusion status of cerebral tissue local to sensor placement. Monitor values are relative within an individual, and should not be used as the sole basis for determining a diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice in cases where hypoxia or ischemia are a concern.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131184: Integra Licox PTO2 Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131184

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026