510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Revolution CT

K133705510(k) TraditionalCleared Apr 11, 2014129-day reviewClass II21 CFR 892.1750Recalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · JAK System, X-Ray, Tomography, Computed

Cleared on Apr 11, 2014 as substantially equivalent to K120833 (Ge Medical Systems, LLC, 2012) after a 129-day review. Its family tree reaches back 6 generations to 2001. 21 later devices descend from it. It has been recalled 14 times.

Decided
Apr 11, 2014
Review
129 days
Class
Class II
Product code
JAK
Recalls
14
Predicates
1

Its longest line of descent runs back 6 generations to K013561 (2001). 21 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2022: 3 descendants cleared2016: 2 descendants cleared2019: 3 descendants cleared2023: 2 descendants cleared2020: 2 descendants cleared2021: 2 descendants cleared2025: 3 descendants cleared2026: 3 descendants cleared2024: 1 descendant clearedK013561 · Lightspeed 4.0 CT Scanner System · GE Medical Systems · 2001 · recalledK030420 · Lightspeed 5.0 Computed Tomography System · GE Medical Systems · 2003 · recalledK040372 · Lightspeed 7.0 CT Scanner System · Ge Medical Systems, LLC · 2004 · recalledK061817 · Lightspeed 7.1 CT Scanner System · Ge Medical Systems, LLC · 2006 · recalledK081105 · GE Lightspeed CT750 HD (Lightspeed 8.0) · GE Healthcare · 2008 · recalledK120833 · Discovery CT750 HD · Ge Medical Systems, LLC · 2012 · recalled200020052010201520202025
  1. Root · 6 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
  2. 5 generations backK030420 Lightspeed 5.0 Computed Tomography SystemGE Medical Systems · 2003 · Recalled
  3. 4 generations backK040372 Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC · 2004 · Recalled
  4. 3 generations backK061817 Lightspeed 7.1 CT Scanner SystemGe Medical Systems, LLC · 2006 · Recalled
  5. 2 generations backK081105 GE Lightspeed CT750 HD (Lightspeed 8.0)GE Healthcare · 2008 · Recalled
  6. 1 generation backK120833 Discovery CT750 HDGe Medical Systems, LLC · 2012 · Recalled
  7. This deviceK133705 Revolution CTGe Medical Systems, LLC · 2014
  8. Descendant · cites it directly · 15 of its ownK163213 Revolution CTGe Medical Systems, LLC · 2016 · Recalled
  9. Descendant · 2 generations on · 8 of its ownK183202 Deep Learning Image ReconstructionGe Medical Systems, LLC · 2019
  10. Descendant · 2 generations on · 5 of its ownK191777 Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with…Ge Medical Systems, LLC · 2019 · Recalled
  11. Descendant · 3 generations on · 4 of its ownK213715 Revolution Apex, Revolution Apex Elite, Revolution Apex Plus…Ge Medical Systems, LLC · 2021 · Recalled
  12. Descendant · cites it directly · 4 of its ownK161066 SpotLight CTArineta, Ltd. · 2016

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

2016: 22019: 32020: 22021: 22022: 32023: 22024: 12025: 32026: 3
2016high 3 (2019) · low 0 (2017)2026

Most influential descendants

Showing 5 of 21 · See all
  • K163213Revolution CTGe Medical Systems, LLC · 2016 · cites it directly15descendants
  • K183202Deep Learning Image ReconstructionGe Medical Systems, LLC · 2019 · 2 generations on8descendants
  • K191777Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex editionGe Medical Systems, LLC · 2019 · 2 generations on5descendants
  • K213715Revolution Apex, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex…Ge Medical Systems, LLC · 2021 · 3 generations on4descendants
  • K161066SpotLight CTArineta, Ltd. · 2016 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K013561Lightspeed 4.0 CT Scanner SystemGE Medical Systems20015
−5K030420Lightspeed 5.0 Computed Tomography SystemGE Medical Systems20034
−4K040372Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC20045
−3K061817Lightspeed 7.1 CT Scanner SystemGe Medical Systems, LLC20061
−2K081105GE Lightspeed CT750 HD (Lightspeed 8.0)GE Healthcare20085
−1K120833Discovery CT750 HDGe Medical Systems, LLC20121
0 (this device)K133705Revolution CTGe Medical Systems, LLC2014—
+1K163213Revolution CTGe Medical Systems, LLC201616
+2K183202Deep Learning Image ReconstructionGe Medical Systems, LLC2019—
+2K191777Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex…Ge Medical Systems, LLC201918
+3K213715Revolution Apex, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex…Ge Medical Systems, LLC202111
+1K161066SpotLight CTArineta, Ltd.2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201620
201930
202020
202120
202230
202320
202410
202530
202630
FIG. 1 — Lineage of K133705, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120833Discovery CT750 HD
Summary page 2
…ation Submission- Revolution CT Marketed Devices: The Revolution CT is a new CT device built upon the existing technologies of the predicate device Discovery CT75O HD (K120833). It is of comparable type and substantially equivalent to its predicate device Discovery CT750 HID and GE's other currently mark…
Ge Medical Systems, LLC20121

Also cleared against K120833: 2 other devices →

What descends from it? (21)

See all 21

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K163213Revolution CTGe Medical Systems, LLC201615
K161066SpotLight CTArineta, Ltd.20164
K213465CardioGrapheArineta, Ltd.20223

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 26, 2026Class IIZ-2150-2026Software designGE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution…
Jun 30, 2021Class IIZ-0242-2022Radiation Control for Health and Safety ActThe accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will…
Jun 30, 2021Class IIZ-0816-2022Radiation Control for Health and Safety ActThe accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
Jun 30, 2021Class IIZ-2021-2021Process designWhen adding one or more intended scan groups names must contain only letters, numbers (0-9), or a space or the user interface will not update to confirm the intended group(s) has been added to the acquisition.
Mar 19, 2021Class IIZ-2004-2021Radiation Control for Health and Safety ActGE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or…
Dec 17, 2020Class IIZ-1130-2021Process change controlGE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table…
Jan 19, 2019Class IIZ-1951-2019Device DesignAdditional low dose radiation exposure.
Jul 12, 2018Class IIZ-0776-2019Software designOn the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a location that is…
May 24, 2018Class IIZ-2582-2018Component change controlSome CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
May 9, 2016Class IIZ-1915-2016Process controlGEHC has become aware of a potential safety issue in the Revolution CT Patient Table that can result in unintended motion in cases involving large patients. The Revolution CT cradle can descend on its own after activation by the user.…
Apr 20, 2016Class IIZ-1589-2016Process controlAn out of tolerance torque tool was used on three joints during the manufacturing process. If the X-ray Tube becomes loose on the CT gantry during operation, this could result in beam tracking errors, potential scan aborts and/or a…
Oct 23, 2015Class IIZ-0401-2016Software designGE Healthcare has recently become aware of a potential issue with routine head scans on your Revolution CT scanner. A potential hazardous situation can occur during a routine head scan with possible artifacts that may emulate pathology…
Dec 19, 2014Class IIZ-0985-2015Process controlPotential safety issue due to a required torque check on three bolted joints of the Revolution CT scanner. No injuries or illnesses reported.
Nov 12, 2014Class IIZ-0955-2015Radiation Control for Health and Safety ActA required quality control test was not performed during installation associated with the software of the Revolution CT scanner.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
  • 6 of the 6 ancestors have a recall of any class.
  • Of its descendants, K163213 (Ge Medical Systems, LLC, 2016) has gone furthest, with 15 descendants of its own.
  • 2 other devices also cite its predicate K120833.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles. The system has the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc.. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients…

Read the full statement (173 words)

The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles. The system has the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc.. The system may acquire data using Axial, Cine, Helical, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results

The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications.

The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133705: Revolution CT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133705

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.