510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Wondfo Multi-Drug Urine Test Cup, Panel

K133968510(k) SpecialCleared Feb 25, 201461-day reviewClass II21 CFR 862.3100Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Guangzhou Wondfo Biotech Co., Ltd. · DKZ Enzyme Immunoassay, Amphetamine

Cleared on Feb 25, 2014 as substantially equivalent to K112071 (Guangzhou Wondfo Biotech Co., Ltd., 2011) after a 61-day review. Its family tree reaches back 2 generations to 2002. It has been recalled once.

Decided
Feb 25, 2014
Review
61 days
Class
Class II
Product code
DKZ
Recalls
1
Predicates
1

Its longest line of descent runs back 2 generations to K020771 (2002). No later device cites it as a predicate yet.

  1. Root · 2 generations backK020771 Acon One Step Drug Screen Test CardACON Laboratories, Inc. · 2002
  2. 1 generation backK112071 Wondfo Cocaine Urine Test Model W10-CU, W10-P, Wondfo…Guangzhou Wondfo Biotech Co., Ltd. · 2011
  3. This deviceK133968 Wondfo Multi-Drug Urine Test Cup, PanelGuangzhou Wondfo Biotech Co., Ltd. · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K020771Acon One Step Drug Screen Test CardACON Laboratories, Inc.2002—
−1K112071Wondfo Cocaine Urine Test Model W10-CU, W10-P, Wondfo Methamphetamine Urine…Guangzhou Wondfo Biotech Co., Ltd.2011—
0 (this device)K133968Wondfo Multi-Drug Urine Test Cup, PanelGuangzhou Wondfo Biotech Co., Ltd.2014—
FIG. 1 — Lineage of K133968, 2002–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DKZ device

  1. #1 K233019 AllTest Multi-Drug Rapid Test Cup; AllTest…score 14.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112071Wondfo Cocaine Urine Test Model W10-CU, W10-P, Wondfo Methamphetamine Urine Test Model…
Summary page 3
…New Devices, Predicate devices (K112071) Indication(s) for For the qualitative determination of Amphetamine For the qualitative determination of use…
Guangzhou Wondfo Biotech Co., Ltd.2011—
†K121987Wondfo Amphetamine Urine Test Wondfo Barbiturates Urine Test Wondfo Benzodiazepines…Guangzhou Wondfo Biotech Co., Ltd.2012—
†K112557Stryker Universal Neuro 3 SystemStryker2012—
†K112236Wondfo Methylenedioxymethamphetamine Urine Test, Wondfo Morphine Urine TestGuangzhou Wondfo Biotech Co., Ltd.2011—
†K112310Wondfo Methadone Urine Test,Opiate TestGuangzhou Wondfo Biotech Co., Ltd.2011—
†K112395Wondfo Phencyclidine Urine Test, Tricyclic Antidepressants Urine TestGuangzhou Wondfo Biotech Co., Ltd.2011—
†K113624Wondfo Buprenorphine Urine Test Wondfo Oxycodone Urine TestGuangzhou Wondfo Biotech Co., Ltd.2012—
†K130055Acon Bup One Step Buprenorphine Test StripGuangzhou Wondfo Biotech Co., Ltd.2013—

Also cleared against K112071: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 25, 2022Class IIZ-1589-2022Environmental controlFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 9.4 years on median.
  • None of the 2 ancestors has a recall.
  • 1 other device also cites its predicate K112071.

How does its category compare?

All DKZ devices

DKZ Enzyme Immunoassay, Amphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · DKZ

1in 2024

2008: 22009: 22010: 42011: 62012: 52013: 82014: 82015: 52016: 52018: 32019: 32020: 12023: 22024: 1
2007high 8 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DKZ

180in 2019

2010: 233 days median2011: 123 days median2012: 55 days median2013: 103 days median2014: 128 days median2015: 107 days median2016: 215 days median2018: 128 days median2019: 180 days median
2007high 233 (2010) · low 55 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DKZ

1in 2017

2007: 12008: 32009: 52011: 42013: 12014: 32017: 1
2007high 5 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DKZ

2in 2025

2008: 122009: 12010: 12011: 22013: 12014: 92015: 82016: 82021: 32022: 42023: 72024: 12025: 2
2007high 12 (2008) · low 0 (2007)2025

MAUDE reports naming product code DKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20082—312
20092—51
2010423301
2011612342
201255500
2013810311
2014812839
2015510708
2016521508
20170—10
2018312800
2019318000
20201—00
20210—03
20220—04
20232—07
20241—01
20250—02
20260—0—

What is it indicated for?

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Phencyclidine, Oxycodone, Buprenorphine, Methadone and Notriptyline in human urine at the cutoff concentrations of: Amphetamine (AMP): 1000 ng/mL…

Read the full statement (444 words)

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Phencyclidine, Oxycodone, Buprenorphine, Methadone and Notriptyline in human urine at the cutoff concentrations of:

Amphetamine (AMP): 1000 ng/mL

Secobarbital (BAR): 300 ng/mL

Oxazepam (BZO): 300 ng/mL

Cocaine (COC): 300 ng/mL

Cannabinoids (THC): 50 ng/mL

Methamphetamine (MET): 1000 ng/mL

Methylenedioxymethamphetamine (MDMA): 500 ng/mL

Morphine (MOP): 300 ng/mL

Phencyclidine (PCP): 25 ng/mL

Oxycodone (OXY): 100 ng/mL

Buprenorphine (BUP): 10 ng/mL

Methadone (MTD): 300 ng/mL

Notriptyline (TCA): 1000 ng/mL

Configuration of the Wondfo Multi-Drug Urine Test Cup can consist of any combination of the above listed drug analytes.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The test may yield positive results for the prescription drugs buprenorphine, oxazepam, oxycodone, and secobarbital when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs.

Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Wondfo Multi-Drug Urine Test Cup and Wondfo Multi-Drug Urine Test Panel are competitive binding, lateral flow immunochromatographic assay for qualitative and simultaneous detection of Amphetamine, Secobarbital, Oxazepam, Cocaine, Cannabinoids, Methamphetamine, Methylenedioxymethamphetamine, Morphine, Phencyclidine, Oxycodone, Buprenorphine, Methadone and Notriptyline in human urine at the cutoff concentrations of:

Amphetamine (AMP): 1000 ng/mL

Secobarbital (BAR): 300 ng/mL

Oxazepam (BZO): 300 ng/mL

Cocaine (COC): 300 ng/mL

Cannabinoids (THC): 50 ng/mL

Methamphetamine (MET): 1000 ng/mL

Methylenedioxymethamphetamine (MDMA): 500 ng/mL

Morphine (MOP): 2000 ng/mL

Phencyclidine (PCP): 25 ng/mL

Oxycodone (OXY): 100 ng/mL

Buprenorphine (BUP): 10 ng/mL

Methadone (MTD): 300 ng/mL

Notriptyline (TCA): 1000 ng/mL

Configuration of the Wondfo Multi-Drug Urine Test Panel can consist of any combination of the above listed drug analytes.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

The test may yield positive results for the prescription drugs buprenorphine, oxazepam, oxycodone, and secobarbital when taken at or above prescribed doses. It is not intended to distinguish between prescription use or abuse of these drugs.

Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133968: Wondfo Multi-Drug Urine Test Cup, Panel.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133968

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026