510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Fukuda Denshi Dynascope Model DS-8100N/8100M Patient Monitor

K134046510(k) TraditionalCleared Feb 12, 2015408-day reviewClass II21 CFR 870.1025Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fukuda Denshi USA, Inc. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Feb 12, 2015 as substantially equivalent to K083697 (Fukuda Denshi USA, Inc., 2009) after a 408-day review. Its family tree reaches back 8 generations to 1997. It has been recalled 2 times.

Decided
Feb 12, 2015
Review
408 days
Class
Class II
Product code
MHX
Recalls
2
Predicates
1

Its longest line of descent runs back 8 generations to K962603 (1997). 26 ancestors in all, the oldest K894628 (1990). No later device cites it as a predicate yet.

  1. Root · 8 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 7 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 6 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 5 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 4 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 3 generations backK021090 Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc. · 2002
  7. 2 generations backK081891 Fukuda Denshi Dynascope Model DS-7000/7000M Patient MonitorFukuda Denshi USA, Inc. · 2008
  8. 1 generation backK083697 Fukuda Denshi Dynascope Model DS-7000 Series Patient Monitor, Model…Fukuda Denshi USA, Inc. · 2009
  9. This deviceK134046 Fukuda Denshi Dynascope Model DS-8100N/8100M Patient MonitorFukuda Denshi USA, Inc. · 2015
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−7K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−6K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−5K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−4K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−3K021090Oximax N-550 Pulse Oximeter,Model N-550 Pulse OximeterNellcor Puritan Bennett, Inc.2002—
−2K081891Fukuda Denshi Dynascope Model DS-7000/7000M Patient MonitorFukuda Denshi USA, Inc.2008—
−1K083697Fukuda Denshi Dynascope Model DS-7000 Series Patient Monitor, Model…Fukuda Denshi USA, Inc.2009—
0 (this device)K134046Fukuda Denshi Dynascope Model DS-8100N/8100M Patient MonitorFukuda Denshi USA, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199020
199540
199610
199730
199910
200010
200110
200330
200420
FIG. 1 — Lineage of K134046, 1997–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K083697Fukuda Denshi Dynascope Model DS-7000 Series Patient Monitor, Model…
Summary page 5
…Monitor Legally Marketed Predicate Devices:  Fukuda Denshi DynaScope Model DS-7000 Series Patient Monitor (Model: DS- 7210/7210M), 510(k) # K083697 In addition, function of the DS-8100N/8100M Patient Monitor utilizes technology incorporated into…
Fukuda Denshi USA, Inc.2009—
†K060576Nellcor Oximax N-600X Pulse OximeterNellcor Puritan Bennett, Inc.2006—
†K110028Masimo Radical Y Pulse CO-OximeterMasimo Corporation2011—
†K094012CAPNOSTREAM20 with MICROMEDICO2 ModuleOridion Capnography, Inc.2010—
†K042601Capnostat 5Respironics Novametrix, Inc.20041

Also cleared against K083697: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 16, 2017Class IIZ-0203-2018Component design/selectionFor the Fukuda Denshi patient monitor model DS-8100, in less than 0.003% (17) of the 6227 devices sold worldwide, a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor…
Dec 21, 2015Class IIZ-0661-2016Process controlA possible malfunction where a Short-term Battery Error message has appeared and has been traced back to a bad (damaged) Lithium-Ion capacitor (Short-term Backup battery).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 11 months on median.
  • 4 of the 26 ancestors have a recall of any class.
  • 1 other device also cites its predicate K083697.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

Use of the Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is indicated in those situations where observation of one or more of the following parameters on an individual patient may be required. ECG (waveform, heart rate, ST-Level and ventricular arrhythmias), respiration, non-invasive blood pressure (NIBP), pulse rate, arterial oxygen saturation (SpO2), carboxyhemoglobin saturation (SpCO)*, methemoglobin saturation (SpMet)*, total…

Read the full statement (203 words)

Use of the Fukuda Denshi DynaScope Model DS-8100N/8100M Patient Monitor is indicated in those situations where observation of one or more of the following parameters on an individual patient may be required. ECG (waveform, heart rate, ST-Level and ventricular arrhythmias), respiration, non-invasive blood pressure (NIBP), pulse rate, arterial oxygen saturation (SpO2), carboxyhemoglobin saturation (SpCO)*, methemoglobin saturation (SpMet)*, total hemoglobin concentration (SpHb)*, plethysmograph, temperature, invasive blood pressure (IBP), cardiac output, and carbon dioxide concentration (CO2). *: DS-8100M only

The target populations of the system are adult, pediatric and neonatal patients with the exception of the ST segment, arrhythmia analysis, and SpHb, for which the target populations are adult and pediatric excluding neonates. These observations can include an audible and visual alarm if any of these parameters exceed values that are established by the clinician. The observations may include the individual or comparative trending of one or more of these parameters over a period of up to 24 hours. The DS-8100N/8100M Patient Monitor is indicated in situations where an instantaneous display of waveform, numeric and trended values is desired. The DS-8100N/8100M Patient Monitor is also indicated where a hard copy record of the physiological parameters, the alarms conditions or the trended values may be required.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K134046: Fukuda Denshi Dynascope Model DS-8100N/8100M Patient Monitor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K134046

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.