510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Fort Defiance Automated Steam Sterilizer

K141009510(k) TraditionalCleared Dec 16, 2014242-day reviewClass II21 CFR 880.6880Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fort Defiance Industries, Inc. · FLE Sterilizer, Steam

Cleared on Dec 16, 2014 as substantially equivalent to K030789 (STERIS Corporation, 2003) after a 242-day review. Its family tree reaches back 5 generations to 1982. It has been recalled 3 times.

Decided
Dec 16, 2014
Review
242 days
Class
Class II
Product code
FLE
Recalls
3
Predicates
1

Its longest line of descent runs back 5 generations to K820783 (1982). 6 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 5 generations backK820783 Castle Microcomputer Sterilizer ControlSybron Corp. · 1982
  2. 4 generations backK970907 Castle Series 100 Steam Sterilizers (Straightline)Getinge/Castle, Inc. · 1997
  3. 3 generations backK994314 Castle Series 100HC Steam SterilizersGetinge/Castle, Inc. · 2000
  4. 2 generations backK012573 Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC…Getinge/Castle, Inc. · 2001
  5. 1 generation backK030789 Amsco Century Steam SterilizerSTERIS Corporation · 2003
  6. This deviceK141009 Fort Defiance Automated Steam SterilizerFort Defiance Industries, Inc. · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K820783Castle Microcomputer Sterilizer ControlSybron Corp.1982—
−4K970907Castle Series 100 Steam Sterilizers (Straightline)Getinge/Castle, Inc.1997—
−3K994314Castle Series 100HC Steam SterilizersGetinge/Castle, Inc.2000—
−2K012573Castle 400HC/500HC Series Steam Sterilizer, Models 422HC, 433HC, 522HC, and…Getinge/Castle, Inc.2001—
−1K030789Amsco Century Steam SterilizerSTERIS Corporation2003—
0 (this device)K141009Fort Defiance Automated Steam SterilizerFort Defiance Industries, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
FIG. 1 — Lineage of K141009, 1982–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FLE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FLE device

  1. #1 K250168 Life Ultra 25L Tabletop Mono Chamber Steam…score 12.7 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030789Amsco Century Steam Sterilizer
Summary page 6
…Sterilizer Classification Name: Steam Sterilizer (21 CFR 880.6880) Product Code FLE 2. Predicate Device K030789, Amsco Century Steam Sterilizer, product code [FLE] cleared March 26, 2003 3. Description of Device The Fort Defiance I…
STERIS Corporation2003—

Also cleared against K030789: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 8, 2026Class IIZ-2983-2026Component design/selectionThe Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Aug 19, 2021Class IIZ-0121-2022Device DesignThe rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
May 24, 2018Class IIZ-2411-2018Device DesignDuring long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 32 years and passes through 4 companies; a new generation was cleared every 2.3 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K030789.

How does its category compare?

All FLE devices

FLE Sterilizer, Steam: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FLE

4in 2025

2007: 52008: 12009: 82010: 22011: 52012: 52013: 52014: 22015: 32016: 12017: 32018: 82019: 52020: 22021: 52022: 102023: 72024: 22025: 42026 (partial year): 3
2007high 10 (2022) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FLE

184in 2025

2007: 67 days median2009: 109 days median2011: 141 days median2012: 111 days median2013: 241 days median2015: 302 days median2017: 233 days median2018: 146 days median2019: 192 days median2021: 147 days median2022: 133 days median2023: 88 days median2025: 184 days median2026 (partial year): 85 days median
2007high 302 (2015) · low 67 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FLE

1in 2021

2007: 12008: 12011: 12012: 12013: 22015: 12018: 12019: 22021: 12026 (partial year): 1
2007high 2 (2013) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FLE

71in 2025

2007: 192008: 212009: 192010: 202011: 632012: 692013: 1122014: 1132015: 682016: 762017: 632018: 902019: 912020: 872021: 812022: 702023: 662024: 642025: 71
2007high 113 (2014) · low 19 (2007)2025

MAUDE reports naming product code FLE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007567119
20081—121
20098109019
20102—020
20115141163
20125111169
201352412112
20142—0113
20153302168
20161—076
20173233063
20188146190
20195192291
20202—087
20215147181
202210133070
2023788066
20242—064
20254184071
20263851—

What is it indicated for?

The Fort Defiance Industries, Inc. Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for field use in a variety of austere environments. Cycle: Immediate Use (IUSS) Recommended Use: Unwrapped nonporous (e.g., instruments) and porous items in…

Read the full statement (213 words)

The Fort Defiance Industries, Inc. Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for field use in a variety of austere environments.

Cycle: Immediate Use (IUSS)

Recommended Use: Unwrapped nonporous (e.g., instruments) and porous items in mixed loads

Maximum Load: 36 lbs.

Exposure Temperature: 270/132 °F/°C

Exposure Time (minutes): 4

Dry Time (minutes): 0

Cycle: Textiles

Recommended Use: Textile packs

Maximum Load: 3 textile packs*

Exposure Temperature: 270/132 °F/°C

Exposure Time (minutes): 4

Dry Time (minutes): 5**

Cycle: Wrapped Instruments

Recommended Use: Wrapped instruments/utensils

Maximum Load: 36 lbs.

Exposure Temperature: 270/132 °F/°C

Exposure Time (minutes): 4

Dry Time (minutes): 20**

Cycle: Wrapped Instruments

Recommended Use: Wrapped instruments/utensils

Maximum Load: 36 lbs.

Exposure Temperature: 270/132 °F/°C

Exposure Time (minutes): 10

Dry Time (minutes): 20**

Cycle: Bowie-Dick Test

Recommended Use: Test

Maximum Load: NA

Exposure Temperature: 273/134 °F/°C

Exposure Time (minutes): 3.5

Dry Time (minutes): 0

Cycle: Vacuum Leak Test

Recommended Use: Test

Maximum Load: NA

Exposure Temperature: NA

Exposure Time: Test time: 20

Dry Time (minutes): 0

Notes:

* AAMI standard 16-towel pack (9"x9"x6")

** Dry times default to established standard conditions, but can be manually increased.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141009: Fort Defiance Automated Steam Sterilizer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141009

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026