510(k) Report

510(k) · Cleared

Cleared

Fortrex 0.035 OTW PTA Balloon Catheter

K142654510(k) TraditionalCleared Oct 28, 201440-day reviewClass II21 CFR 870.1250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Covidien, LLC · LIT Catheter, Angioplasty, Peripheral, Transluminal

Cleared on Oct 28, 2014 as substantially equivalent to K103322 (Ev3, Inc., 2010) and 2 other predicates after a 40-day review. Its family tree reaches back 7 generations to 2000.

Decided
Oct 28, 2014
Review
40 days
Class
Class II
Product code
LIT
Recalls
None
Predicates
3

Its longest line of descent runs back 7 generations to K001134 (2000). 18 ancestors in all, the oldest K993305 (2000). No later device cites it as a predicate yet.

  1. Root · 7 generations backK001134 Gazelle Balloon Dilatation CatheterBoston Scientific Corp · 2000
  2. 6 generations backK021735 Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc. · 2002 · Recalled
  3. 5 generations backK050389 Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp · 2005
  4. 4 generations backK053116 Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp · 2005
  5. 3 generations backK072283 Dorado Pta Balloon Dilatation CatheterBard Peripheral Vascular, Inc. · 2007 · Recalled
  6. 2 generations backK082579 Evercross .035 Otw Pta Dilatation CatheterEv3, Inc. · 2008 · Recalled
  7. 1 generation backK110319 Evercross 0.035 Otw Pta Dilatation CatheterEv3, Inc. · 2011
  8. This deviceK142654 Fortrex 0.035 OTW PTA Balloon CatheterCovidien, LLC · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K001134Gazelle Balloon Dilatation CatheterBoston Scientific Corp2000—
−6K021735Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc.200223
−5K050389Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp2005—
−4K053116Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp2005—
−3K072283Dorado Pta Balloon Dilatation CatheterBard Peripheral Vascular, Inc.200715
−2K082579Evercross .035 Otw Pta Dilatation CatheterEv3, Inc.20081
−1K110319Evercross 0.035 Otw Pta Dilatation CatheterEv3, Inc.2011—
0 (this device)K142654Fortrex 0.035 OTW PTA Balloon CatheterCovidien, LLC2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200210
200420
200510
200620
200810
201010
201110
FIG. 1 — Lineage of K142654, 2000–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LIT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LIT device

  1. #1 K241040 Amethyst HP PTA OTW 0.035 Catheterscore 13.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103322Evercross 0.035 Otw Pta Dilation Catheter
Summary page 4
…ular 3. Predicate Devices Device Trade Name EverCross™ .035” OTW PTA Dilatation Catheter (Primary Predicate) 510(k) Number K103322, K110319 510(k) Clearance Date December 06, 2010, April 14, 2011 Device Trade Name Mustang™ Balloon Dilatation Catheter…
Ev3, Inc.20101
*K110319Evercross 0.035 Otw Pta Dilatation Catheter
Summary page 4
…. Predicate Devices Device Trade Name EverCross™ .035” OTW PTA Dilatation Catheter (Primary Predicate) 510(k) Number K103322, K110319 510(k) Clearance Date December 06, 2010, April 14, 2011 Device Trade Name Mustang™ Balloon Dilatation Catheter 5…
Ev3, Inc.2011—
*K103751Mustang Balloon Dilatation Catheters
Summary page 4
…(k) Number K103322, K110319 510(k) Clearance Date December 06, 2010, April 14, 2011 Device Trade Name Mustang™ Balloon Dilatation Catheter 510(k) Number K103751 510(k) Clearance Date March 22, 2011 4. Device Description The Fortrex™ 0.035” PTA Balloon Catheter (Fortrex Catheter) is an…
Boston Scientific2011—

Also cleared against K103322: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 14 years and passes through 5 companies; a new generation was cleared every 1.9 years on median.
  • 4 of the 18 ancestors have a recall of any class.
  • 1 other device also cites its predicate K103322.

How does its category compare?

All LIT devices

LIT Catheter, Angioplasty, Peripheral, Transluminal: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · LIT

3in 2025

2007: 62008: 82009: 72010: 132011: 102012: 192013: 142014: 122015: 192016: 112017: 92018: 62019: 42020: 92021: 42022: 82023: 62024: 62025: 32026 (partial year): 4
2007high 19 (2012) · low 3 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LIT

90in 2025

2007: 64 days median2008: 53 days median2009: 29 days median2010: 78 days median2011: 82 days median2012: 29 days median2013: 79 days median2014: 90 days median2015: 112 days median2016: 116 days median2017: 57 days median2018: 118 days median2019: 40 days median2020: 125 days median2021: 61 days median2022: 122 days median2023: 167 days median2024: 43 days median2025: 90 days median2026 (partial year): 205 days median
2007high 205 (2026) · low 29 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LIT

1in 2025

2009: 22010: 12013: 2, 2 Class I2014: 32016: 12018: 22019: 32021: 12024: 22025: 1
2007high 3 (2014) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LIT

2,073in 2025

2007: 5772008: 5132009: 6232010: 8092011: 8572012: 1,2772013: 1,2422014: 1,1782015: 1,2792016: 1,5292017: 2,0812018: 2,1882019: 2,4052020: 2,4432021: 2,4922022: 2,1552023: 1,8072024: 1,9132025: 2,073
2007high 2,492 (2021) · low 513 (2008)2025

MAUDE reports naming product code LIT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076640577
20088530513
20097292623
201013781809
201110820857
2012192901,277
201314792 (2)1,242
2014129031,178
20151911201,279
20161111611,529
201795702,081
2018611822,188
201944032,405
2020912502,443
202146112,492
2022812202,155
2023616701,807
202464321,913
202539012,073
202642050—

What is it indicated for?

The Fortrex™ 0.035” OTW PTA Balloon Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K142654: Fortrex 0.035 OTW PTA Balloon Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K142654

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026