510(k) Report

510(k) · Cleared

Recalled · Class I · 2020

ROSA Brain

K151359510(k) TraditionalCleared Dec 18, 2015212-day reviewClass II21 CFR 882.4560Recalls 7 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtech S.A · HAW Neurological Stereotaxic Instrument

Cleared on Dec 18, 2015 as substantially equivalent to K101791 (Medtech S.A.S, 2010) after a 212-day review. Its family tree reaches back 6 generations to 2000. 4 of the 41 devices in its family, counting itself, have a Class I recall. 20 later devices descend from it. It has been recalled 7 times, including a Class I recall.

Decided
Dec 18, 2015
Review
212 days
Class
Class II
Product code
HAW
Recalls
7 (1 Class I)
Predicates
1

Its longest line of descent runs back 6 generations to K002053 (2000). 20 ancestors in all, the oldest K811452 (1981). 20 later devices descend from it, 2 citing it directly. 4 of the 41 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 3 other ancestors2000: 1 other ancestor1999: 2 other ancestors2001: 2 other ancestors1997: 1 other ancestor2005: 2 other ancestors1987: 1 other ancestor1981: 1 other ancestor2003: 1 other ancestor2019: 2 descendants cleared, 1 with a Class I recall12018: 1 descendant cleared, 1 with a Class I recall12023: 3 descendants cleared2021: 3 descendants cleared2025: 4 descendants cleared2020: 2 descendants cleared, 1 with a Class I recall12024: 1 descendant cleared2022: 1 descendant cleared2026: 3 descendants clearedK002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalledK101791 · Rosa Surgical Device · Medtech S.A.S · 2010 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 3 other ancestors3 generations back: 3 other ancestors4 generations back: 6 other ancestors5 generations back: 1 other ancestor6 generations back: 1 other ancestor 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 5 descendants, 1 with a Class I recall13 generations forward: 3 descendants, 1 with a Class I recall14 generations forward: 5 descendants5 generations forward: 4 descendants6 generations forward: 1 descendant K101791 · Rosa Surgical Device · Medtech S.A.S · 2010 · recalledK092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalledK060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000−50+5
  1. Root · 6 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 5 generations backK021760 Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc. · 2002
  3. 4 generations backK043536 Navitrack System - S&N Image Free KneeOrthosoft, Inc. · 2005
  4. 3 generations backK060556 Brigit Surgical DeviceZimmer, Inc. · 2006
  5. 2 generations backK092239 Rosa Surgical Device, Model Rosa 1.1Medtech S.A · 2009 · Recalled
  6. 1 generation backK101791 Rosa Surgical DeviceMedtech S.A.S · 2010 · Recalled
  7. This deviceK151359 ROSA BrainMedtech S.A · 2015
  8. Descendant · cites it directly · 14 of its ownK172444 ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · Recalled (Class I)
  9. Descendant · 2 generations on · 13 of its ownK182417 ROSA ONE Brain ApplicationMedtech S.A · 2019 · Recalled (Class I)
  10. Descendant · 3 generations on · 5 of its ownK200047 ExcelsiusGPS Cranial 1.0 ModuleGlobus Medical, Inc. · 2020 · Recalled
  11. Descendant · 3 generations on · 5 of its ownK200511 ROSA ONE Brain applicationMedtech S.A · 2020 · Recalled (Class I)
  12. Descendant · cites it directly · 4 of its ownK191597 Stealth Autoguide System, Midas Rex Legend Depth Stop SystemMedtronic Navigation, Inc. · 2019 · Recalled

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

2018: 12019: 22020: 22021: 32022: 12023: 32024: 12025: 42026: 3
2018high 4 (2025) · low 1 (2018)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 20 · See all
  • K172444ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · cites it directly14descendants
  • K182417ROSA ONE Brain ApplicationMedtech S.A · 2019 · 2 generations on13descendants
  • K200047ExcelsiusGPS Cranial 1.0 ModuleGlobus Medical, Inc. · 2020 · 3 generations on5descendants
  • K200511ROSA ONE Brain applicationMedtech S.A · 2020 · 3 generations on5descendants
  • K191597Stealth Autoguide System, Midas Rex Legend Depth Stop SystemMedtronic Navigation, Inc. · 2019 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K002053Navitrak System-Optical Option, Model 900.004Orthosoft, Inc.2000—
−5K021760Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc.2002—
−4K043536Navitrack System - S&N Image Free KneeOrthosoft, Inc.2005—
−3K060556Brigit Surgical DeviceZimmer, Inc.2006—
−2K092239Rosa Surgical Device, Model Rosa 1.1Medtech S.A20099
−1K101791Rosa Surgical DeviceMedtech S.A.S201012
0 (this device)K151359ROSA BrainMedtech S.A2015—
+1K172444ROSA BRAIN (v3.0.0.5)Medtech S.A2018Class I
+2K182417ROSA ONE Brain ApplicationMedtech S.A2019Class I
+3K200047ExcelsiusGPS Cranial 1.0 ModuleGlobus Medical, Inc.20203
+3K200511ROSA ONE Brain applicationMedtech S.A2020Class I
+1K191597Stealth Autoguide System, Midas Rex Legend Depth Stop SystemMedtronic Navigation, Inc.20191

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198710
199710
199920
200010
200120
200310
200430
200520

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201811
201921
202021
202130
202210
202330
202410
202540
202630
FIG. 1 — Lineage of K151359, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101791Rosa Surgical Device
Our reading of the summary gave K101797, FDA’s number for Sonicision Cordless Ultrasonic Dissection Device Model… (Covidien, Formerly Valleylab, A Division of Tyco). Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “ROSA Surgical Device”.
Medtech S.A.S201012

Also cleared against K101791: 3 other devices →

What descends from it? (20)

See all 20

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172444ROSA BRAIN (v3.0.0.5) Class IMedtech S.A201814
K191597Stealth Autoguide System, Midas Rex Legend Depth Stop SystemMedtronic Navigation, Inc.20194

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 6, 2020Class IIZ-1151-2020Software designSome cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
Jan 6, 2020Class IIZ-1150-2020Software designSome cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
Sep 10, 2019Class IZ-0172-2020Software in the Use EnvironmentThe instrument holder may be sent on a trajectory that is not within the intended target. If it is not corrected, the associated device may be placed incorrectly.
Feb 27, 2017Class IIZ-1754-2017Component change controlUnapproved change made by the supplier.
Oct 11, 2016Class IIZ-2868-2017Process designComplaint of the head holder connector locking up mechanically when tightened.
Jul 29, 2016Class IIZ-0116-2018Error in labelingNon-conforming product label.
Jul 20, 2016Class IIZ-0167-2018Software designCommunication errors between ROSANNA BRAIN software, MARIO software and the Stubli CS8C controller.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K172444ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · Cites it directly · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
  • K182417ROSA ONE Brain ApplicationMedtech S.A · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
  • K200511ROSA ONE Brain applicationMedtech S.A · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 15 years and passes through 4 companies; a new generation was cleared every 2.4 years on median.
  • 4 of the 20 ancestors have a recall of any class.
  • Of its descendants, K172444 (Medtech S.A, 2018) has gone furthest, with 14 descendants of its own.
  • 3 other devices also cite its predicate K101791.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K151359: ROSA Brain.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151359

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026