510(k) Report

510(k) · Cleared

Recalled · Class I · 2021

ROSA ONE Brain application

K200511510(k) TraditionalCleared May 29, 202088-day reviewClass II21 CFR 882.4560Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtech S.A · HAW Neurological Stereotaxic Instrument

Cleared on May 29, 2020 as substantially equivalent to K182417 (Medtech S.A, 2019) after an 88-day review. Its family tree reaches back 9 generations to 2000. 4 of the 29 devices in its family, counting itself, have a Class I recall. 5 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
May 29, 2020
Review
88 days
Class
Class II
Product code
HAW
Recalls
1 (1 Class I)
Predicates
1

Its longest line of descent runs back 9 generations to K002053 (2000). 23 ancestors in all, the oldest K811452 (1981). 5 later devices descend from it, 3 citing it directly. 4 of the 29 devices in this family, counting this one, have a Class I recall.

  1. Root · 9 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 5 generations between
  3. 3 generations backK151359 ROSA BrainMedtech S.A · 2015 · Recalled (Class I)
  4. 2 generations backK172444 ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · Recalled (Class I)
  5. 1 generation backK182417 ROSA ONE Brain ApplicationMedtech S.A · 2019 · Recalled (Class I)
  6. This deviceK200511 ROSA ONE Brain applicationMedtech S.A · 2020
  7. Descendant · cites it directly · 1 of its ownK241333 Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc. · 2025
  8. Descendant · cites it directly · 1 of its ownK214065 ROSA ONE Brain applicationMedtech S.A.S · 2022
  9. Descendant · 2 generations on · 0 of its ownK253856 Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc. · 2026
  10. Descendant · 2 generations on · 0 of its ownK231103 ROSA ONE Brain ApplicationMedtech S.A.S · 2023
  11. Descendant · cites it directly · 0 of its ownK220072 Sinobot X1Sinovation (Beijing) Medical Technology Co., Ltd. · 2023
Show all 9 generations
  1. Root · 9 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 8 generations backK021760 Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc. · 2002
  3. 7 generations backK043536 Navitrack System - S&N Image Free KneeOrthosoft, Inc. · 2005
  4. 6 generations backK060556 Brigit Surgical DeviceZimmer, Inc. · 2006
  5. 5 generations backK092239 Rosa Surgical Device, Model Rosa 1.1Medtech S.A · 2009 · Recalled
  6. 4 generations backK101791 Rosa Surgical DeviceMedtech S.A.S · 2010 · Recalled
  7. 3 generations backK151359 ROSA BrainMedtech S.A · 2015 · Recalled (Class I)
  8. 2 generations backK172444 ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · Recalled (Class I)
  9. 1 generation backK182417 ROSA ONE Brain ApplicationMedtech S.A · 2019 · Recalled (Class I)
  10. This deviceK200511 ROSA ONE Brain applicationMedtech S.A · 2020

Class I recalls in this family

Most influential descendants

Showing 5 of 5 · See all
  • K241333Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc. · 2025 · cites it directly1descendant
  • K214065ROSA ONE Brain applicationMedtech S.A.S · 2022 · cites it directly1descendant
  • K253856Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc. · 2026 · 2 generations on0descendants
  • K231103ROSA ONE Brain ApplicationMedtech S.A.S · 2023 · 2 generations on0descendants
  • K220072Sinobot X1Sinovation (Beijing) Medical Technology Co., Ltd. · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K002053Navitrak System-Optical Option, Model 900.004Orthosoft, Inc.2000—
−8K021760Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc.2002—
−7K043536Navitrack System - S&N Image Free KneeOrthosoft, Inc.2005—
−6K060556Brigit Surgical DeviceZimmer, Inc.2006—
−5K092239Rosa Surgical Device, Model Rosa 1.1Medtech S.A20099
−4K101791Rosa Surgical DeviceMedtech S.A.S201012
−3K151359ROSA BrainMedtech S.A2015Class I
−2K172444ROSA BRAIN (v3.0.0.5)Medtech S.A2018Class I
−1K182417ROSA ONE Brain ApplicationMedtech S.A2019Class I
0 (this device)K200511ROSA ONE Brain applicationMedtech S.A2020—
+1K241333Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc.2025—
+1K214065ROSA ONE Brain applicationMedtech S.A.S2022—
+2K253856Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc.2026—
+2K231103ROSA ONE Brain ApplicationMedtech S.A.S2023—
+1K220072Sinobot X1Sinovation (Beijing) Medical Technology Co., Ltd.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198710
199710
199920
200010
200120
200310
200430
200520

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202320
202510
202610
FIG. 1 — Lineage of K200511, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K182417ROSA ONE Brain Application
Summary page 4
…duct Code: HAW (Brain) 510k #: K200511 III PREDICATE DEVICES ROSA ONE Brain application, manufactured by Medtech S.A., K182417, cleared on February 7, 2019…
Medtech S.A2019Class I

Also cleared against K182417: 2 other devices →

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K241333Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc.20251
K214065ROSA ONE Brain applicationMedtech S.A.S20221
K220072Sinobot X1Sinovation (Beijing) Medical Technology Co., Ltd.20230

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 22, 2021Class IZ-0118-2022Software designThe firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the issue. An incorrect…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K182417ROSA ONE Brain ApplicationMedtech S.A · 2019 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K172444ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K151359ROSA BrainMedtech S.A · 2015 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 20 years and passes through 4 companies; a new generation was cleared every 2.1 years on median.
  • 7 of the 23 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K241333 (Koh Young Technology, Inc., 2025) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K182417.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200511: ROSA ONE Brain application.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K200511

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026