510(k) Report

510(k) · Cleared

Recalled · Class I · 2021

ROSA ONE Brain Application

K182417510(k) TraditionalCleared Feb 7, 2019155-day reviewClass II21 CFR 882.4560Recalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtech S.A · HAW Neurological Stereotaxic Instrument

Cleared on Feb 7, 2019 as substantially equivalent to K172444 (Medtech S.A, 2018) after a 155-day review. Its family tree reaches back 8 generations to 2000. 4 of the 36 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 3 times, including a Class I recall.

Decided
Feb 7, 2019
Review
155 days
Class
Class II
Product code
HAW
Recalls
3 (1 Class I)
Predicates
1

Its longest line of descent runs back 8 generations to K002053 (2000). 22 ancestors in all, the oldest K811452 (1981). 13 later devices descend from it, 3 citing it directly. 4 of the 36 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 3 other ancestors2000: 1 other ancestor1999: 2 other ancestors2001: 2 other ancestors1997: 1 other ancestor2005: 2 other ancestors1987: 1 other ancestor1981: 1 other ancestor2003: 1 other ancestor2020: 2 descendants cleared, 1 with a Class I recall12025: 3 descendants cleared2021: 1 descendant cleared2024: 1 descendant cleared2022: 1 descendant cleared2023: 2 descendants cleared2026: 3 descendants clearedK002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalledK101791 · Rosa Surgical Device · Medtech S.A.S · 2010 · recalledK151359 · ROSA Brain · Medtech S.A · 2015 · Class I recallK172444 · ROSA BRAIN (v3.0.0.5) · Medtech S.A · 2018 · Class I recall19801990200020102020
Other ancestors, by generationDescendants, by generation 4 generations back: 3 other ancestors5 generations back: 3 other ancestors6 generations back: 6 other ancestors7 generations back: 1 other ancestor8 generations back: 1 other ancestor 1 generation forward: 3 descendants, 1 with a Class I recall12 generations forward: 5 descendants3 generations forward: 4 descendants4 generations forward: 1 descendant K172444 · ROSA BRAIN (v3.0.0.5) · Medtech S.A · 2018 · Class I recallK151359 · ROSA Brain · Medtech S.A · 2015 · Class I recallK101791 · Rosa Surgical Device · Medtech S.A.S · 2010 · recalledK092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalledK060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000−50
  1. Root · 8 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 7 generations backK021760 Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc. · 2002
  3. 6 generations backK043536 Navitrack System - S&N Image Free KneeOrthosoft, Inc. · 2005
  4. 5 generations backK060556 Brigit Surgical DeviceZimmer, Inc. · 2006
  5. 4 generations backK092239 Rosa Surgical Device, Model Rosa 1.1Medtech S.A · 2009 · Recalled
  6. 3 generations backK101791 Rosa Surgical DeviceMedtech S.A.S · 2010 · Recalled
  7. 2 generations backK151359 ROSA BrainMedtech S.A · 2015 · Recalled (Class I)
  8. 1 generation backK172444 ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · Recalled (Class I)
  9. This deviceK182417 ROSA ONE Brain ApplicationMedtech S.A · 2019
  10. Descendant · cites it directly · 5 of its ownK200047 ExcelsiusGPS Cranial 1.0 ModuleGlobus Medical, Inc. · 2020 · Recalled
  11. Descendant · cites it directly · 5 of its ownK200511 ROSA ONE Brain applicationMedtech S.A · 2020 · Recalled (Class I)
  12. Descendant · 2 generations on · 2 of its ownK231850 QUARTEX® Occipito-Cervico-Thoracic Spinal System ExcelsiusGPS®…Globus Medical, Inc. · 2024
  13. Descendant · 3 generations on · 1 of its ownK252327 CMORE® CT System; CMORE® CT System Navigated InstrumentsIcotec AG · 2025
  14. Descendant · 2 generations on · 1 of its ownK241333 Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc. · 2025

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2020: 22021: 12022: 12023: 22024: 12025: 32026: 3
2020high 3 (2025) · low 1 (2021)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K200047ExcelsiusGPS Cranial 1.0 ModuleGlobus Medical, Inc. · 2020 · cites it directly5descendants
  • K200511ROSA ONE Brain applicationMedtech S.A · 2020 · cites it directly5descendants
  • K231850QUARTEX® Occipito-Cervico-Thoracic Spinal System ExcelsiusGPS® InstrumentsGlobus Medical, Inc. · 2024 · 2 generations on2descendants
  • K252327CMORE® CT System; CMORE® CT System Navigated InstrumentsIcotec AG · 2025 · 3 generations on1descendant
  • K241333Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc. · 2025 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K002053Navitrak System-Optical Option, Model 900.004Orthosoft, Inc.2000—
−7K021760Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc.2002—
−6K043536Navitrack System - S&N Image Free KneeOrthosoft, Inc.2005—
−5K060556Brigit Surgical DeviceZimmer, Inc.2006—
−4K092239Rosa Surgical Device, Model Rosa 1.1Medtech S.A20099
−3K101791Rosa Surgical DeviceMedtech S.A.S201012
−2K151359ROSA BrainMedtech S.A2015Class I
−1K172444ROSA BRAIN (v3.0.0.5)Medtech S.A2018Class I
0 (this device)K182417ROSA ONE Brain ApplicationMedtech S.A2019—
+1K200047ExcelsiusGPS Cranial 1.0 ModuleGlobus Medical, Inc.20203
+1K200511ROSA ONE Brain applicationMedtech S.A2020Class I
+2K231850QUARTEX® Occipito-Cervico-Thoracic Spinal System ExcelsiusGPS® InstrumentsGlobus Medical, Inc.2024—
+3K252327CMORE® CT System; CMORE® CT System Navigated InstrumentsIcotec AG2025—
+2K241333Geniant Cranial (Navigated Neurosurgical Positioning Robot)Koh Young Technology, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198710
199710
199920
200010
200120
200310
200430
200520

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202021
202110
202210
202320
202410
202530
202630
FIG. 1 — Lineage of K182417, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K172444ROSA BRAIN (v3.0.0.5)
Summary page 4
…II Product Code: HAW (Brain) 510k #: K182417 III PREDICATE DEVICES ROSA BRAIN, manufactured by Medtech S.A., K172444, cleared on March 28, 2018 1 of 10 === Page 5…
Medtech S.A2018Class I

What descends from it? (13)

See all 13

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K200047ExcelsiusGPS Cranial 1.0 ModuleGlobus Medical, Inc.20205
K200511ROSA ONE Brain application Class IMedtech S.A20205
K242575Stereotaxic Guiding Surgical Devices, NaoTrac SBrain Navi Biotechnology Co., Ltd.20250

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 22, 2021Class IZ-0118-2022Software designThe firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the issue. An incorrect…
Jan 6, 2020Class IIZ-1151-2020Software designSome cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.
Jan 6, 2020Class IIZ-1150-2020Software designSome cross-sectional images from the image acquisitions of the patients head may not be reconstructed/displayed properly in two and three dimension views when using ROSA Brain software, potentially compromising the surgery planning.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K172444ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K151359ROSA BrainMedtech S.A · 2015 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K200511ROSA ONE Brain applicationMedtech S.A · 2020 · Cites it directly · ◆ in the 2020 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 19 years and passes through 4 companies; a new generation was cleared every 2.3 years on median.
  • 6 of the 22 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K200047 (Globus Medical, Inc., 2020) has gone furthest, with 5 descendants of its own.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic neurosurgery may be appropriate.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K182417: ROSA ONE Brain Application.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K182417

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026