510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Amia Automated PD System with Sharesource

K151525510(k) TraditionalCleared Oct 9, 2015123-day reviewClass II21 CFR 876.5630Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Deka Research and Development · FKX System, Peritoneal, Automatic Delivery

Cleared on Oct 9, 2015 as substantially equivalent to K124018 (Deka Research & Development Corp., 2013) and 1 other predicate after a 123-day review. Its family tree reaches back 5 generations to 1994. 1 later device descends from it. It has been recalled 8 times.

Decided
Oct 9, 2015
Review
123 days
Class
Class II
Product code
FKX
Recalls
8
Predicates
2

Its longest line of descent runs back 5 generations to K923065 (1994). 10 ancestors in all, the oldest K902149 (1990). 1 later device descends from it, 1 citing it directly.

  1. Root · 5 generations backK923065 Baxter Personal Cycler Autommated Peritoneal DialBaxter Healthcare Corp · 1994 · Recalled
  2. 4 generations backK012988 Homechoice Personal Cycler Peritoneal Dialysis System, Models…Baxter Healthcare Corp · 2001 · Recalled
  3. 3 generations backK043363 Fresenius Liberty CyclerFresenius Medical Care North America · 2005 · Recalled
  4. 2 generations backK103220 Voyager Peritoneal Dialysis SystemDeka Research & Development Corp. · 2011
  5. 1 generation backK124018 Amia Automated PD SystemDeka Research & Development Corp. · 2013 · Recalled
  6. This deviceK151525 Amia Automated PD System with SharesourceDeka Research and Development · 2015
  7. Descendant · cites it directly · 0 of its ownK201867 Homechoice Claria APD System, Sharesource Connectivity Platform for…Baxter Healthcare Corporation · 2020 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K201867Homechoice Claria APD System, Sharesource Connectivity Platform for Use with the…Baxter Healthcare Corporation · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K923065Baxter Personal Cycler Autommated Peritoneal DialBaxter Healthcare Corp19944
−4K012988Homechoice Personal Cycler Peritoneal Dialysis System, Models 5C8310, 5C8302…Baxter Healthcare Corp20019
−3K043363Fresenius Liberty CyclerFresenius Medical Care North America20059
−2K103220Voyager Peritoneal Dialysis SystemDeka Research & Development Corp.2011—
−1K124018Amia Automated PD SystemDeka Research & Development Corp.20131
0 (this device)K151525Amia Automated PD System with SharesourceDeka Research and Development2015—
+1K201867Homechoice Claria APD System, Sharesource Connectivity Platform for Use with…Baxter Healthcare Corporation20202

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199910
200110
200620

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
FIG. 1 — Lineage of K151525, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FKX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FKX device

  1. #1 K243371 Byonyks X-1 APD Cycler; Byonyks Automated PD…score 11.1 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K124018Amia Automated PD System
Summary page 4
…ogy/Urology Predicate Device(s) The Amia Automated PD System is substantially equivalent to the Amia Automated PD System, which was previously cleared under application K124018. The HomeChoice PRO Automated Personal Cycler previously cleared under K102936 is also included as a reference predicate. The Sha…
Deka Research & Development Corp.20131
*K061515Renalsoft
Summary page 4
…ded as a reference predicate. The Sharesource software is substantially equivalent to Baxter Healthcare Corporation’s RenalSoft v.2.0, which was previously cleared under K061515. The PDK accessory for the Amia Automated PD System (K124018) is also included as a reference predicate. Device Description The…
Baxter Healthcare Corp20062
†K102936Homechoice/Homechoice Pro Automated Personal Cycler Peritoneal Dialysis System Model…Baxter Healthcare Corp201111

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201867Homechoice Claria APD System, Sharesource Connectivity Platform for Use with the…Baxter Healthcare Corporation20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 15, 2023Class IIZ-0550-2024Process controlThere is to an increase in complaints related to failure alarms for Wet Cassette Integrity Test occurring on the Amia devices.
Jun 30, 2021Class IIZ-2173-2021Under Investigation by firmBaxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria…
Jun 30, 2021Class IIZ-2174-2021Under Investigation by firmBaxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria…
Dec 11, 2018Class IIZ-0849-2019Software designPotential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy
Dec 11, 2018Class IIZ-0850-2019Software designPotential for the software on Automated PD System cyclers which can cause shortened dwell times during a Cycle-Based therapy
Jul 12, 2018Class IIZ-2597-2018Under Investigation by firmThere were instances where the power cord socket detached from the back of the AMIA cycler when the power cord was unplugged from the socket.
Sep 1, 2017Class IIZ-3133-2017Equipment maintenanceThe firm received increased customer complaints for Missing Red Line, Patient Slow Flow, Solution Slow Flow, and Inadequate Drain alerts on certain lots of the AMIA Automated Peritoneal Dialysis Set with Cassette.
Feb 16, 2016Class IIZ-1565-2016Under Investigation by firmBaxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume.…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 4 companies; a new generation was cleared every 3.3 years on median.
  • 8 of the 10 ancestors have a recall of any class.
  • Of its descendants, K201867 (Baxter Healthcare Corporation, 2020) has gone furthest, with 0 descendants of its own.

How does its category compare?

All FKX devices

FKX System, Peritoneal, Automatic Delivery: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · FKX

2in 2025

2011: 22013: 22014: 12015: 12017: 12018: 22020: 12022: 42025: 2
2007high 4 (2022) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FKX

209in 2022

2022: 209 days median
2007high 209 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FKX

1in 2025

2007: 22010: 52011: 22012: 22013: 12014: 42015: 22016: 12018: 32020: 12023: 22025: 1
2007high 5 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FKX

3,128in 2025

2007: 1472008: 7022009: 3402010: 4,9662011: 11,3042012: 18,0342013: 22,0932014: 28,2892015: 24,6632016: 6,1372017: 3,1272018: 5,9272019: 5,9722020: 4,5982021: 5,0762022: 4,5032023: 3,9132024: 3,7712025: 3,128
2007high 28,289 (2014) · low 147 (2007)2025

MAUDE reports naming product code FKX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—2147
20080—0702
20090—0340
20100—54,966
20112—211,304
20120—218,034
20132—122,093
20141—428,289
20151—224,663
20160—16,137
20171—03,127
20182—35,927
20190—05,972
20201—14,598
20210—05,076
2022420904,503
20230—23,913
20240—03,771
20252—13,128
20260—0—

What is it indicated for?

The Amia Automated PD System is intended for automatic control of dialysate solution exchanges in the treatment of adult renal failure patients undergoing peritoneal dialysis. All therapies using the Amia Automated PD System must be prescribed and performed under the responsibility of a physician who is familiar and well informed about peritoneal dialysis. The Sharesource portal is intended for use by healthcare professionals to remotely communicate new…

Read the full statement (120 words)

The Amia Automated PD System is intended for automatic control of dialysate solution exchanges in the treatment of adult renal failure patients undergoing peritoneal dialysis.

All therapies using the Amia Automated PD System must be prescribed and performed under the responsibility of a physician who is familiar and well informed about peritoneal dialysis.

The Sharesource portal is intended for use by healthcare professionals to remotely communicate new or modified treatment parameters with compatible dialysis instruments and transfer completed treatment data to a central database to aid in the review, analysis, and evaluation of patients’ historical treatment results. This system is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K151525: Amia Automated PD System with Sharesource.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151525

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.