510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Uroview FD

K161019510(k) TraditionalCleared Aug 17, 2016127-day reviewClass II21 CFR 892.1650Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pausch Medical GmbH · JAA System, X-Ray, Fluoroscopic, Image-Intensified

Cleared on Aug 17, 2016 as substantially equivalent to K101491 (Siemens Medical Solutions USA, Inc., 2010) after a 127-day review. Its family tree reaches back 9 generations to 1980. 2 later devices descend from it. It has been recalled once.

Decided
Aug 17, 2016
Review
127 days
Class
Class II
Product code
JAA
Recalls
1
Predicates
1

Its longest line of descent runs back 9 generations to K801623 (1980). 58 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors2002: 4 other ancestors1999: 3 other ancestors2000: 3 other ancestors1997: 4 other ancestors2004: 2 other ancestors1996: 4 other ancestors1995: 12 other ancestors1993: 3 other ancestors1989: 1 other ancestor1986: 1 other ancestor1992: 2 other ancestors1991: 4 other ancestors1988: 2 other ancestors1981: 1 other ancestor1994: 1 other ancestor2018: 1 descendant cleared2020: 1 descendant clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000×2: 2 generations cleared close together, K052202 (2006), K061054 (2006) K062623 · Axiom Luminos DRF · Siemens Medical Solutions USA, Inc. · 2007 · recalledK101491 · Uroskop Omnia · Siemens Medical Solutions USA, Inc. · 2010 · recalled19801990200020102020
  1. Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 5 generations between
  3. 3 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  4. 2 generations backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  5. 1 generation backK101491 Uroskop OmniaSiemens Medical Solutions USA, Inc. · 2010 · Recalled
  6. This deviceK161019 Uroview FDPausch Medical GmbH · 2016
  7. Descendant · cites it directly · 0 of its ownK200965 XR-MX/1000Karl Storz Endoscopy America, Inc. · 2020
  8. Descendant · cites it directly · 0 of its ownK180651 Uroview FD IIPausch Medical GmbH · 2018
Show all 9 generations
  1. Root · 9 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 8 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 7 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 6 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 5 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 4 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 3 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  8. 2 generations backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  9. 1 generation backK101491 Uroskop OmniaSiemens Medical Solutions USA, Inc. · 2010 · Recalled
  10. This deviceK161019 Uroview FDPausch Medical GmbH · 2016

Most influential descendants

Showing 2 of 2 · See all
  • K200965XR-MX/1000Karl Storz Endoscopy America, Inc. · 2020 · cites it directly0descendants
  • K180651Uroview FD IIPausch Medical GmbH · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−8K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−7K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−6K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−5K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−4K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−3K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.20063
−2K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200720
−1K101491Uroskop OmniaSiemens Medical Solutions USA, Inc.201012
0 (this device)K161019Uroview FDPausch Medical GmbH2016—
+1K200965XR-MX/1000Karl Storz Endoscopy America, Inc.2020—
+1K180651Uroview FD IIPausch Medical GmbH2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198820
198910
199140
199220
199330
199410
1995120
199640
199740
199930
200030
200120
200240
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202010
FIG. 1 — Lineage of K161019, 1980–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAA device

  1. #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101491Uroskop Omnia
Summary page 4
…ubmission of Initial Reports on Diagnostic X-Ray Systems and Their Major Components 5. Legally Marketed Predicate Device UROSKOP Omnia (K101491, 06/30/2010) from Siemens Medical Solutions USA Inc. Page 1 of 8 === Page 5 ==…
Siemens Medical Solutions USA, Inc.201012

Also cleared against K101491: 3 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K200965XR-MX/1000Karl Storz Endoscopy America, Inc.20200
K180651Uroview FD IIPausch Medical GmbH20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 14, 2017Class IIZ-2152-2017Device DesignDuring patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure +…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 36 years and passes through 6 companies; a new generation was cleared every 2.0 years on median.
  • 20 of the 58 ancestors have a recall of any class.
  • Of its descendants, K200965 (Karl Storz Endoscopy America, Inc., 2020) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K101491.

How does its category compare?

All JAA devices

JAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · JAA

2in 2025

2007: 12008: 62010: 12011: 42012: 12013: 62014: 62015: 32016: 22017: 12018: 32019: 62020: 42021: 62022: 12023: 22024: 32025: 22026 (partial year): 1
2007high 6 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAA

89in 2024

2008: 18 days median2011: 48 days median2013: 129 days median2014: 276 days median2015: 197 days median2018: 30 days median2019: 49 days median2020: 39 days median2021: 104 days median2024: 89 days median
2007high 276 (2014) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAA

4in 2025

2007: 352008: 212009: 182010: 62011: 72012: 22013: 32014: 12015: 12016: 72017: 32018: 12019: 22020: 22021: 22022: 52023: 92024: 22025: 4
2007high 35 (2007) · low 1 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAA

33in 2025

2007: 9,5172008: 9,4842009: 12,0402010: 5,8032011: 14,0002012: 11,6842013: 13,7062014: 10,2652015: 2,9222016: 352017: 562018: 362019: 492020: 392021: 352022: 342023: 412024: 252025: 33
2007high 14,000 (2011) · low 25 (2024)2025

MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—359,517
2008618219,484
20090—1812,040
20101—65,803
2011448714,000
20121—211,684
20136129313,706
20146276110,265
2015319712,922
20162—735
20171—356
2018330136
2019649249
2020439239
20216104235
20221—534
20232—941
2024389225
20252—433
20261—0—

What is it indicated for?

The Uroview FD is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray-diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/ urologic treatment table, may be used for urological treatment, planning and diagnostic procedures including but not limited to: • Querying and retrieving patient information and /or images from other modalities. • X-ray…

Read the full statement (178 words)

The Uroview FD is a solid state detector fluoroscopic X-ray system, primarily for urological applications (functional x-ray-diagnostics, endourology and minimal invasive urology/surgery). The system, which includes a radiologic/ urologic treatment table, may be used for urological treatment, planning and diagnostic procedures including but not limited to:

• Querying and retrieving patient information and /or images from other modalities.

• X-ray examination of the urogenital area (e.g. cyctoscopy, kidney, bladder, ureter, urethra) including KUB, IVP, vasovesiculography, reflux-cystogram, cystourethrogram, and micturation cystourethrogram combined with uroflow measurements.

• Ultrasound examinations(in conjunction with a stand-alone ultrasound system) of the kidney, bladder, prostate, scrotum.

• Endourological interventions (e.g. of the urethra, prostate, bladder, sphincter, ostium, kidney and ureter, catheter placement , penile implant placement, transurethral resection of prostate or bladder, alternative treatment of the BPH and brachytherapy).

• Percutaneous interventions (e.g. PCN nephrolithotomy, resection, percutaneous nephrostomy).

• Laparoscopy (e.g. cholecystectomy, nephrectomy, lymph node dissection, abdominal testis detection/correction, varicocele).

• Application of fistula (kidney/bladder).

• Simple procedures (e.g. urethra, testis, phimonis).

• Introcorporeal shock wave lithotripsy.

• Uroflow/urodynamics.

• Pediatric radiological and therapeutic applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K161019: Uroview FD.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161019

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026