510(k) Report

510(k) · Cleared

Recalled · Class III · 2022

VERIFY SixCess Steam Indicators, VERIFY SixCess Flash Indicators, VERIFY Bowie Dick Test Pack, VERIFY SixCess Challenge Pack, VERIFY SixCess FP Challenge Pack

K162758510(k) TraditionalCleared Mar 21, 2017172-day reviewClass II21 CFR 880.2800Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · JOJ Indicator, Physical/Chemical Sterilization Process

Cleared on Mar 21, 2017 as substantially equivalent to K002741 (Albert Browne, Ltd., 2002) and 5 other predicates after a 172-day review. Its family tree reaches back 5 generations to 1994. It has been recalled once.

Decided
Mar 21, 2017
Review
172 days
Class
Class II
Product code
JOJ
Recalls
1
Predicates
6

Its longest line of descent runs back 5 generations to K932129 (1994). 20 ancestors in all, the oldest K801109 (1980). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2009: 1 other ancestor2007: 3 other ancestors2002: 2 other ancestors2008: 1 other ancestor2001: 1 other ancestor1997: 2 other ancestors1990: 2 other ancestors1989: 1 other ancestor1994: 1 other ancestor1980: 1 other ancestorReference device: K965154 · TST Control Integrator for Steam Autoclave · Albert Browne, Ltd. · 1997K932129 · 3M Autoclave Tapes · 3M Company · 1994K992767 · Browne Packaging and Label Steam Process Indicator · Albert Browne, Ltd. · 2000K032801 · Modified Browne Packaging and Label Steam Process Indicator · Albert Browne, Ltd. · 2003K073683 · Verify Challenge Packs - Version 2 · STERIS Corporation · 2008K103053 · Verify 270FP Challenge Pack · STERIS Corporation · 2011 · recalled1980199020002010
  1. Root · 5 generations backK932129 3M Autoclave Tapes3M Company · 1994
  2. 4 generations backK992767 Browne Packaging and Label Steam Process IndicatorAlbert Browne, Ltd. · 2000
  3. 3 generations backK032801 Modified Browne Packaging and Label Steam Process IndicatorAlbert Browne, Ltd. · 2003
  4. 2 generations backK073683 Verify Challenge Packs - Version 2STERIS Corporation · 2008
  5. 1 generation backK103053 Verify 270FP Challenge PackSTERIS Corporation · 2011 · Recalled
  6. This deviceK162758 VERIFY SixCess Steam Indicators, VERIFY SixCess Flash Indicators…STERIS Corporation · 2017
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K9321293M Autoclave Tapes3M Company1994—
−4K992767Browne Packaging and Label Steam Process IndicatorAlbert Browne, Ltd.2000—
−3K032801Modified Browne Packaging and Label Steam Process IndicatorAlbert Browne, Ltd.2003—
−2K073683Verify Challenge Packs - Version 2STERIS Corporation2008—
−1K103053Verify 270FP Challenge PackSTERIS Corporation20111
0 (this device)K162758VERIFY SixCess Steam Indicators, VERIFY SixCess Flash Indicators, VERIFY Bowie…STERIS Corporation2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198910
199020
199410
199720
200110
200220
200730
200810
200910
FIG. 1 — Lineage of K162758, 1994–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K002741Browne TST Single Use Bowie Dick/Integrator Test Pack
Summary page 10
…fication Number: 21 CFR 880.2800 (b) Product Code: JOJ 2. Predicate Devices 510(k) # Device Name K002741 Browne TST Dual Use Test Pack K070461 VERIFY Steam Indicators K071895 VERIFY 275F 3 Indi…
Albert Browne, Ltd.2002—
*K070461Verify Steam Indicators, Models 250F 25, 250F 30, 270F 15, 270F 3-10 and 270F 4
Summary page 10
…Code: JOJ 2. Predicate Devices 510(k) # Device Name K002741 Browne TST Dual Use Test Pack K070461 VERIFY Steam Indicators K071895 VERIFY 275F 3 Indicator K073683 VERIFY Challenge Packs –…
STERIS Corporation20071
*K071895Verify 275 Degree F. 3-10 Indicator; 10 Indicator; 3 Indicator
Summary page 10
…510(k) # Device Name K002741 Browne TST Dual Use Test Pack K070461 VERIFY Steam Indicators K071895 VERIFY 275F 3 Indicator K073683 VERIFY Challenge Packs – Versions 2 K083643 VERIFY 275F…
STERIS Corporation2007—
*K073683Verify Challenge Packs - Version 2
Summary page 10
…K002741 Browne TST Dual Use Test Pack K070461 VERIFY Steam Indicators K071895 VERIFY 275F 3 Indicator K073683 VERIFY Challenge Packs – Versions 2 K083643 VERIFY 275F Gravity Indicator K103053 VERIFY…
STERIS Corporation2008—
*K083643Verify 275F Gravity Indicators
Summary page 10
…K070461 VERIFY Steam Indicators K071895 VERIFY 275F 3 Indicator K073683 VERIFY Challenge Packs – Versions 2 K083643 VERIFY 275F Gravity Indicator K103053 VERIFY 270FP Challenge Pack The proposed devices are identical…
STERIS Corporation2009—
*K103053Verify 270FP Challenge Pack
Summary page 10
…K071895 VERIFY 275F 3 Indicator K073683 VERIFY Challenge Packs – Versions 2 K083643 VERIFY 275F Gravity Indicator K103053 VERIFY 270FP Challenge Pack The proposed devices are identical to the predicates in performance and composition w…
STERIS Corporation20111
†K965154TST Control Integrator for Steam AutoclaveAlbert Browne, Ltd.1997—

Also cleared against K002741: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 10, 2022Class IIIZ-1736-2022Process controlProduct contains incorrect chemical indicators in packaging.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 3 companies; a new generation was cleared every 4.3 years on median.
  • 4 of the 20 ancestors have a recall of any class.
  • 1 other device also cites its predicate K002741.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

The Verify® Steam Indicators are emulating indicators intended for use in steam sterilization. The Verify® Steam Indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The indicator models and their cycle temperatures, types, and times are: MODEL TEMPERATURE STERILIZATION TYPE TIME Verity 250F 30 250°F (121°C) gravity steam 30 minutes Verify 270F 15 270°F (132°C) gravity steam 15 minutes…

Read the full statement (546 words)

The Verify® Steam Indicators are emulating indicators intended for use in steam sterilization. The Verify® Steam Indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The indicator models and their cycle temperatures, types, and times are:

MODEL TEMPERATURE STERILIZATION TYPE TIME

Verity 250F 30 250°F (121°C) gravity steam 30 minutes

Verify 270F 15 270°F (132°C) gravity steam 15 minutes

Verify 270F 3-10 270°F (132°C) gravity flash steam 3 or 10 minutes

Verity 270F 4 270°F (132°C) SFPP, pre-vacuum and Express steam 4 minutes

The Verify® 275F 3 Indicator is a chemical indicator intended for use by health care providers to accompany products being sterilized through a sterilization procedure. The Verify® 275F 3 Indicator is an emulating indicator intended for use in steam sterilization. The indicator changes color from yellow to blue/purple when exposed to 275°F (135°C) pre-vacuum sterilization cycles for 3 minutes as indicated in the following table:

MODEL TEMPERATURE STERILIZATION TYPE TIME

Verify 275F 3 275°F (135°C) Pre-vacuum steam 3 minutes

The Verify® 275F Gravity Indicators are chemical integrators which meet ANSI/AAMI 1140-1:2005 for emulating indicators intended for use in steam sterilization. The Verify® 275F Gravity Indicators change color from yellow to blue/purple when exposed to 275°F (135°C) and to the appropriate cycle type and duration. The Verify® 275F Indicators models and their cycle types and times are:

MODEL TEMPERATURE STERILIZATION TYPE TIME

Verify 275F 3-10 275°F (135°C) Gravity flash steam 3 and 10 minutes

Verify 275F 10 275°F (135°C) Gravity steam 10 minutes

The Verify® Challenge Packs —Version 2 - are test packs consisting of an emulating indicator surrounded by a steam penetration barrier, intended for use in steam sterilization. The Verify® Challenge Packs indicators change color from yellow to blue/purple when exposed to the appropriate cycle temperature, type, and duration. The challenge pack models and their cycle temperatures, types, and times are:

- Verify® 270F 4 - 270°F (132°C) Pre-vacuum, Steam Flush Pressure Pulse (SFPP), 4 minutes

- Verify® 275F 3 - 275°F (135°C) Pre-vacuum 3 minutes

There is a process indicator outside of the packs which undergoes a visual color change when exposed to steam in a temperature range of 250°F (121°C) to 275°F (135°C).

The Verify® 270FP Challenge Pack is a test pack consisting of three emulating indicator inks situated on a test sheet surrounded by a steam penetration barrier, intended for use in SFPP (steam-flush pressure-pulse) and pre-vacuum steam sterilization for load release. The Verify® 270FP Challenge Pack indicators change color from yellow to blue/purple when exposed to the following conditions at 270°F (132°C):

- 4 minute indicator - SFPP and pre-vacuum steam sterilization for 4 minutes.

- 10 minute indicator - Pre-vacuum steam sterilization for 10 minutes.

- 20 minute indicator - Pre-vacuum steam sterilization for 20 minutes.

The Browne TST Dual Use Test Pack is a chemical sterilization process monitor that can be used as either:

- Bowie Dick Test - in an empty chamber to monitor air removal and steam penetration during the vacuum stage of 132°C (270°F), 134°C (273°F) , and 135 °C (275 °F) test cycles.

- Challenge Pack - in a loaded chamber to monitor steam penetration during the hold/plateau stage of 132°C (270°F), 134 °C (273 °F), and 135 °C (275 °F) steam sterilization cycles.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K162758: VERIFY SixCess Steam Indicators, VERIFY SixCess Flash Indicators, VERIFY Bowie Dick Test Pack, VERIFY SixCess Challenge Pack, VERIFY SixCess FP Challenge Pack.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K162758

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026