Breathe Technologies Life2000™ Ventilation System

K170037510(k) TraditionalCleared Aug 11, 2017219-day reviewClass II21 CFR 868.5895Recalls 13 (12 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Breathe Technologies · CBK Ventilator, Continuous, Facility Use

Cleared on Aug 11, 2017 as substantially equivalent to K141943 (Breathe Technologies, 2015) and 1 other predicate after a 219-day review. Its family tree reaches back 8 generations to 1991. 4 of the 43 devices in its family, counting itself, have a Class I recall. It has been recalled 13 times, including 12 Class I recalls.

Decided
Aug 11, 2017
Review
219 days
Class
Class II
Product code
CBK
Recalls
13 (12 Class I)
Predicates
2

Its longest line of descent runs back 8 generations to K902859 (1991). 42 ancestors in all, the oldest K895541 (1990). No later device cites it as a predicate yet. 4 of the 43 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2006: 1 other ancestor2003: 4 other ancestors, 1 with a Class I recall12005: 1 other ancestor2001: 1 other ancestor2000: 2 other ancestors2004: 1 other ancestor1999: 3 other ancestors1998: 5 other ancestors, 1 with a Class I recall11997: 1 other ancestor1996: 3 other ancestors1995: 3 other ancestors1994: 3 other ancestors1992: 2 other ancestors1991: 1 other ancestor1990: 3 other ancestorsReference device: K023354 · Compressor Mini, 115 V, Model 64 81 779 EH81E · Siemens Elema AB · 2003 · recalledK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK960010 · Servo Ventilator 300 Computer Interface · Siemens Elema AB · 1996K970839 · Servo Ventilator 300A · Siemens Elema AB · 1997 · recalledK992788 · Bear Cub 750 PSV · Bear Medical Systems, Inc. · 1999K013642 · AVEA · Bird Products Corp. · 2002 · Class I recallK022674 · Bird Avea Ventilator · Bird Products Corp. · 2002 · recalledK060647 · LTV 1200 Ventilator · Pulmonetic Systems, Inc. · 2006K141943 · Breathe Technologies LIFE2000 Ventilation System · Breathe Technologies · 2015199019952000200520102015
  1. Root · 8 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 7 generations backK960010 Servo Ventilator 300 Computer InterfaceSiemens Elema AB · 1996
  3. 6 generations backK970839 Servo Ventilator 300ASiemens Elema AB · 1997 · Recalled
  4. 5 generations backK992788 Bear Cub 750 PSVBear Medical Systems, Inc. · 1999
  5. 4 generations backK013642 AVEABird Products Corp. · 2002 · Recalled (Class I)
  6. 3 generations backK022674 Bird Avea VentilatorBird Products Corp. · 2002 · Recalled
  7. 2 generations backK060647 LTV 1200 VentilatorPulmonetic Systems, Inc. · 2006
  8. 1 generation backK141943 Breathe Technologies LIFE2000 Ventilation SystemBreathe Technologies · 2015
  9. This deviceK170037 Breathe Technologies Life2000™ Ventilation SystemBreathe Technologies · 2017

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K902859Servo Ventilator 300Siemens Life Support Systems19913
−7K960010Servo Ventilator 300 Computer InterfaceSiemens Elema AB1996—
−6K970839Servo Ventilator 300ASiemens Elema AB19974
−5K992788Bear Cub 750 PSVBear Medical Systems, Inc.1999—
−4K013642AVEABird Products Corp.2002Class I
−3K022674Bird Avea VentilatorBird Products Corp.20025
−2K060647LTV 1200 VentilatorPulmonetic Systems, Inc.2006—
−1K141943Breathe Technologies LIFE2000 Ventilation SystemBreathe Technologies2015—
0 (this device)K170037Breathe Technologies Life2000™ Ventilation SystemBreathe Technologies2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199030
199110
199220
199430
199530
199630
199710
199851
199930
200020
200110
200341
200410
200510
200610
FIG. 1 — Lineage of K170037, 1991–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141943Breathe Technologies LIFE2000 Ventilation System
Summary page 4
…Life2000 Ventilation System Predicate Device Manufacturer: Breathe Technologies, Inc. Predicate Device Premarket Notification # K141943 Predicate Device Common Name: Continuous Ventilator Predicate Device Classification & Name: 21 CFR 868.5895…
Breathe Technologies2015—
*K062019Respcare Hybrid NE Face Mask, Model TBD,; Disposable, Model TBD
Summary page 5
…dicate Device: Hybrid NE Mask Predicate Device Manufacturer: RespCare, Inc. Predicate Device Premarket Notification # K062019 Predicate Device Common Name: Face Mask Accessory to Continuous…
Innomed Technologies, Inc./Respcare, Inc.2006—
†K023354Compressor Mini, 115 V, Model 64 81 779 EH81ESiemens Elema AB20031

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 7, 2025Class IZ-1725-2025Software in the Use EnvironmentA cybersecurity vulnerability was discovered through internal testing.
Apr 7, 2025Class IZ-1726-2025Software in the Use EnvironmentA cybersecurity vulnerability was discovered through internal testing.
Dec 20, 2024Class IZ-0886-2025Process controlThere is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.
Dec 20, 2024Class IZ-0884-2025Process controlThere is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.
Dec 20, 2024Class IZ-0885-2025Process controlThere is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.
Dec 20, 2024Class IZ-0887-2025Process controlThere is an issue with the ventilator battery charger that could lead to an audible and visual alarm. If the alarm is engaged, the ventilator becomes inoperable.
Sep 12, 2024Class IZ-0046-2025Software Design ChangeThe Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.
May 29, 2024Class IZ-2140-2024Under Investigation by firmPotential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery…
Jan 25, 2023Class IZ-1142-2023Under Investigation by firmThere is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Jan 25, 2023Class IZ-1141-2023Under Investigation by firmThere is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Jan 25, 2023Class IZ-1140-2023Under Investigation by firmThere is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Jan 25, 2023Class IZ-1139-2023Under Investigation by firmThere is the potential for patient desaturation events to occur under certain scenarios when the Life2000 ventilation system is connected with an oxygen concentrator.
Nov 12, 2020Class IIZ-0484-2021Process controlUnder certain conditions, there is a risk for fire.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K013642AVEABird Products Corp. · 2002 · 4 generations back · ◆ on the line of descent in FIG. 113 Class I recalls
  • K023366Datex-Ohmeda 7900 Ventilator EnhancementsDatex-Ohmeda, Inc. · 2003 · 5 generations back · ◆ in the 2003 ancestors bar of FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 5 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 6 companies; a new generation was cleared every 2.7 years on median.
  • 12 of the 42 ancestors have a recall of any class; 3 have a Class I recall.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Breathe Technologies Life2000™ Ventilation System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation. The Life2000 Ventilation System consists of the Life2000 Ventilator and the Life2000 Compressor. The System is intended for use by qualified, trained personnel under the direction of a physician. Specifically, the System is applicable for adult patients who require the…

Read the full statement (107 words)

The Breathe Technologies Life2000™ Ventilation System is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

The Life2000 Ventilation System consists of the Life2000 Ventilator and the Life2000 Compressor.

The System is intended for use by qualified, trained personnel under the direction of a physician. Specifically, the System is applicable for adult patients who require the following types of ventilatory support:

• Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).

• Assist/Control mode of ventilation.

The System is suitable for use in home and institutional settings and is not intended for ambulance or air transportation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170037: Breathe Technologies Life2000™ Ventilation System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170037

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026