510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Invacare® TDX® SP2 Power Wheelchair; TDXSP2 TDX SP2 with LiNX Electronics, TDXSP2-CG TDX SP2 Base for Single Act. System w/ Ultra Low Maxx, TDXSP2-MCG TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx

K170507510(k) TraditionalCleared Jul 12, 2017141-day reviewClass II21 CFR 890.3860Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Invacare Corporation · ITI Wheelchair, Powered

Cleared on Jul 12, 2017 as substantially equivalent to K141783 (Invacare Corporation, 2015) and 1 other predicate after a 141-day review. Its family tree reaches back 6 generations to 1990. 3 later devices descend from it. It has been recalled 2 times.

Decided
Jul 12, 2017
Review
141 days
Class
Class II
Product code
ITI
Recalls
2
Predicates
2

Its longest line of descent runs back 6 generations to K900565 (1990). 11 ancestors in all. 3 later devices descend from it, 2 citing it directly.

Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant K141783 · TDX SP2 Power Wheechair · Invacare Corporation · 2015K023589 · Storm TDX Power Wheelchair · Invacare Corp. · 2002 · recalledK012909 · Xterra GT Power Wheelchair · Invacare Corp. · 2001K991168 · Arrow FWD Power Wheelchair · Invacare Corp. · 1999 · recalledK982064 · Action Storm Ranger II, Action Storm Power 9000 · Invacare Corp. · 1998K900565 · Invacare Power 9000 Series Wheelchair · Invacare Corp. · 1990−50
  1. Root · 6 generations backK900565 Invacare Power 9000 Series WheelchairInvacare Corp. · 1990
  2. 5 generations backK982064 Action Storm Ranger II, Action Storm Power 9000Invacare Corp. · 1998
  3. 4 generations backK991168 Arrow FWD Power WheelchairInvacare Corp. · 1999 · Recalled
  4. 3 generations backK012909 Xterra GT Power WheelchairInvacare Corp. · 2001
  5. 2 generations backK023589 Storm TDX Power WheelchairInvacare Corp. · 2002 · Recalled
  6. 1 generation backK141783 TDX SP2 Power WheechairInvacare Corporation · 2015
  7. This deviceK170507 Invacare® TDX® SP2 Power Wheelchair; TDXSP2 TDX SP2 with LiNX…Invacare Corporation · 2017
  8. Descendant · cites it directly · 1 of its ownK192216 Invacare AVIVA FX Power Wheelchair, Model: IFX-20MPInvacare Corporation · 2020 · Recalled
  9. Descendant · 2 generations on · 0 of its ownK202379 Invacare AVIVA Storm RX Power ChairInvacare Corporation · 2021 · Recalled
  10. Descendant · cites it directly · 0 of its ownK182946 Invacare TDX SP2-HD with Captain's SeatInvacare Corporation · 2018

Most influential descendants

Showing 3 of 3 · See all
  • K192216Invacare AVIVA FX Power Wheelchair, Model: IFX-20MPInvacare Corporation · 2020 · cites it directly1descendant
  • K202379Invacare AVIVA Storm RX Power ChairInvacare Corporation · 2021 · 2 generations on0descendants
  • K182946Invacare TDX SP2-HD with Captain's SeatInvacare Corporation · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K900565Invacare Power 9000 Series WheelchairInvacare Corp.1990—
−5K982064Action Storm Ranger II, Action Storm Power 9000Invacare Corp.1998—
−4K991168Arrow FWD Power WheelchairInvacare Corp.19992
−3K012909Xterra GT Power WheelchairInvacare Corp.2001—
−2K023589Storm TDX Power WheelchairInvacare Corp.20027
−1K141783TDX SP2 Power WheechairInvacare Corporation2015—
0 (this device)K170507Invacare® TDX® SP2 Power Wheelchair; TDXSP2 TDX SP2 with LiNX Electronics…Invacare Corporation2017—
+1K192216Invacare AVIVA FX Power Wheelchair, Model: IFX-20MPInvacare Corporation20201
+2K202379Invacare AVIVA Storm RX Power ChairInvacare Corporation20211
+1K182946Invacare TDX SP2-HD with Captain's SeatInvacare Corporation2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
199910
200210
201510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
202010
202110
FIG. 1 — Lineage of K170507, 1990–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in ITI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ITI device

  1. #1 K220747 Power Wheelchair (N5515B)score 40.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141783TDX SP2 Power Wheechair
Summary page 4
…tion Name: Powered Wheelchair 21 CFR § 890.3860 Regulatory Class: II Product Code: ITI PRIMARY PREDICATE Invacare® TDX® SP2 Power Wheelchair (K141783) DEVICE: SECONDARY PREDICATE Modular Power Positioning System (K150574) DEVICE: REFERENCE DEVICES: F3 Powered Wheelcha…
Invacare Corporation2015—
*K150574Modular Power Positioning System
Summary page 4
…II Product Code: ITI PRIMARY PREDICATE Invacare® TDX® SP2 Power Wheelchair (K141783) DEVICE: SECONDARY PREDICATE Modular Power Positioning System (K150574) DEVICE: REFERENCE DEVICES: F3 Powered Wheelchair (K143180) F5 Powered Wheelchair (K143014)…
Motion Concepts2015—
†K143180F3Permobil AB2015—
†K143014F5Permobil AB2015—
†K143383Q6 Edge EMPride Mobility Product Corporation20157

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192216Invacare AVIVA FX Power Wheelchair, Model: IFX-20MPInvacare Corporation20201
K182946Invacare TDX SP2-HD with Captain's SeatInvacare Corporation20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 30, 2022Class IIZ-1130-2022Device DesignThe battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery…
Nov 12, 2021Class IIZ-0445-2022Software designPower Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 27 years and passes through 2 companies; a new generation was cleared every 2.5 years on median.
  • 3 of the 11 ancestors have a recall of any class.
  • Of its descendants, K192216 (Invacare Corporation, 2020) has gone furthest, with 1 descendant of its own.

How does its category compare?

All ITI devices

ITI Wheelchair, Powered: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · ITI

29in 2025

2007: 182008: 92009: 72010: 62011: 42012: 62013: 52014: 52015: 102016: 62017: 42018: 42019: 32020: 62021: 132022: 162023: 82024: 382025: 292026 (partial year): 36
2007high 38 (2024) · low 3 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ITI

90in 2025

2007: 42 days median2008: 49 days median2009: 63 days median2010: 46 days median2011: 101 days median2012: 155 days median2013: 177 days median2014: 213 days median2015: 174 days median2016: 173 days median2017: 159 days median2018: 117 days median2019: 244 days median2020: 278 days median2021: 209 days median2022: 96 days median2023: 128 days median2024: 125 days median2025: 90 days median2026 (partial year): 108 days median
2007high 278 (2020) · low 42 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ITI

13so far in 2026, the most since 2007

2007: 32008: 32009: 12010: 22012: 12016: 12017: 22021: 32024: 12025: 3, 2 Class I2026 (partial year): 13
2007high 13 (2026) · low 0 (2011)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · ITI

647in 2025

2007: 1232008: 1332009: 1832010: 1572011: 5252012: 2,9192013: 8,7412014: 7,2822015: 4,6852016: 1,5752017: 3012018: 2732019: 2642020: 2052021: 1662022: 1352023: 2232024: 1462025: 647
2007high 8,741 (2013) · low 123 (2007)2025

MAUDE reports naming product code ITI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200718423123
20089493133
20097631183
20106462157
201141010525
2012615512,919
2013517708,741
2014521307,282
20151017404,685
2016617311,575
201741592301
201841170273
201932440264
202062780205
2021132093166
202216960135
202381280223
2024381251146
202529903 (2)647
20263610813—

What is it indicated for?

The indication for use of the Invacare® TDX® SP2 Power Wheelchair is to provide mobility and positioning to persons limited to a sitting position.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K170507: Invacare® TDX® SP2 Power Wheelchair; TDXSP2 TDX SP2 with LiNX Electronics, TDXSP2-CG TDX SP2 Base for Single Act. System w/ Ultra Low Maxx, TDXSP2-MCG TDX SP2 Base for Multiple Actuator System w/ Ultra Low Maxx.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K170507

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026