510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

NOxBOXi Nitric Oxide Delivery System

K171696510(k) TraditionalCleared Oct 2, 2018482-day reviewClass II21 CFR 868.5165Recalls 6 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Praxair Distribution, Inc. · MRN Apparatus, Nitric Oxide Delivery

Cleared on Oct 2, 2018 as substantially equivalent to K092545 (Ino Therapeutics, LLC, 2010) and 1 other predicate after a 482-day review. Its family tree reaches back 5 generations to 2001. 5 of the 14 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 6 times, including 3 Class I recalls.

Decided
Oct 2, 2018
Review
482 days
Class
Class II
Product code
MRN
Recalls
6 (3 Class I)
Predicates
2

Its longest line of descent runs back 5 generations to K003665 (2001). 9 ancestors in all, the oldest K000186 (2000). 4 later devices descend from it, 2 citing it directly. 5 of the 14 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor2003: 1 other ancestor2000: 2 other ancestors2022: 1 descendant cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared, 1 with a Class I recall12024: 1 descendant cleared, 1 with a Class I recall1K003665 · Fadassi Medical FM-1 NO Blender · Fadasis Medical, Inc. · 2001K011874 · Fadasis Medical FM-1 NO Blender · Fadasis Medical, Inc. · 2001K052663 · Ino Therapeutics Inoblender, Model 10004 · Ino Therapeutics · 2005K061901 · Ino Therapeutics Inomax DS (Delivery System), Model 10003 · Ino Therapeutics · 2006 · recalledK092545 · Inomax DS (Delivery System), Model 10003 · Ino Therapeutics, LLC · 2010200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 3 other ancestors 1 generation forward: 2 descendants, 2 with a Class I recall22 generations forward: 1 descendant, 1 with a Class I recall13 generations forward: 1 descendant, 1 with a Class I recall1 K092545 · Inomax DS (Delivery System), Model 10003 · Ino Therapeutics, LLC · 2010K061901 · Ino Therapeutics Inomax DS (Delivery System), Model 10003 · Ino Therapeutics · 2006 · recalledK052663 · Ino Therapeutics Inoblender, Model 10004 · Ino Therapeutics · 2005K011874 · Fadasis Medical FM-1 NO Blender · Fadasis Medical, Inc. · 2001K003665 · Fadassi Medical FM-1 NO Blender · Fadasis Medical, Inc. · 2001−50
  1. Root · 5 generations backK003665 Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  2. 4 generations backK011874 Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc. · 2001
  3. 3 generations backK052663 Ino Therapeutics Inoblender, Model 10004Ino Therapeutics · 2005
  4. 2 generations backK061901 Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics · 2006 · Recalled
  5. 1 generation backK092545 Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC · 2010
  6. This deviceK171696 NOxBOXi Nitric Oxide Delivery SystemPraxair Distribution, Inc. · 2018
  7. Descendant · cites it directly · 3 of its ownK201339 The NOxBOXi Nitric Oxide Delivery SystemPraxair, Inc. · 2020 · Recalled (Class I)
  8. Descendant · cites it directly · 2 of its ownK220898 NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2022 · Recalled (Class I)
  9. Descendant · 2 generations on · 1 of its ownK231823 NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2023 · Recalled (Class I)
  10. Descendant · 3 generations on · 0 of its ownK233251 NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2024 · Recalled (Class I)

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K201339The NOxBOXi Nitric Oxide Delivery SystemPraxair, Inc. · 2020 · cites it directly3descendants
  • K220898NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2022 · cites it directly2descendants
  • K231823NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2023 · 2 generations on1descendant
  • K233251NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K003665Fadassi Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−4K011874Fadasis Medical FM-1 NO BlenderFadasis Medical, Inc.2001—
−3K052663Ino Therapeutics Inoblender, Model 10004Ino Therapeutics2005—
−2K061901Ino Therapeutics Inomax DS (Delivery System), Model 10003Ino Therapeutics20063
−1K092545Inomax DS (Delivery System), Model 10003Ino Therapeutics, LLC2010—
0 (this device)K171696NOxBOXi Nitric Oxide Delivery SystemPraxair Distribution, Inc.2018—
+1K201339The NOxBOXi Nitric Oxide Delivery SystemPraxair, Inc.2020Class I
+1K220898NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc.2022Class I
+2K231823NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc.2023Class I
+3K233251NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc.2024Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200020
200310
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202011
202211
202311
202411
FIG. 1 — Lineage of K171696, 2001–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MRN device

  1. #1 K242374 ULSPIRA TS™ Nitric Oxide Therapy Systemscore 12.6 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092545Inomax DS (Delivery System), Model 10003
Summary page 5
…K171696 - Page 2 of 8 5. Legally Marketed Predicate Device(s) There are two predicate devices chosen for this submission. 1. K092545 – INO Therapeutics’ INOmax DS Delivery System 2. K122689 – INO Therapeutics’ INOblender – Nitric Oxide Administration – Back-…
Ino Therapeutics, LLC2010—
*K122689Inoblender
Summary page 5
…Legally Marketed Predicate Device(s) There are two predicate devices chosen for this submission. 1. K092545 – INO Therapeutics’ INOmax DS Delivery System 2. K122689 – INO Therapeutics’ INOblender – Nitric Oxide Administration – Back-up system. Note that both the functionality of the…
Ino Therapeutics2012—

Also cleared against K092545: 6 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201339The NOxBOXi Nitric Oxide Delivery System Class IPraxair, Inc.20203
K220898NOxBOXi Nitric Oxide Delivery System Class ILinde Gas & Equipment, Inc.20222

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 14, 2026Class IIZ-3306-2026Component design/selectionPotential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors.
Aug 19, 2026Class IIZ-3165-2026Device DesignPotential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation.
Sep 9, 2025Class IZ-0028-2026Software designFluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when…
Sep 9, 2025Class IZ-0029-2026Software designThe unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
Jul 19, 2023Class IIZ-2544-2025Component design/selectionReplacement of the internal Sample Pump due to high failure rates.
May 3, 2023Class IZ-1667-2023Component change controlDevice may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K220898NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2022 · Cites it directly · ◆ in the 2022 descendants bar of FIG. 12 Class I recalls
  • K201339The NOxBOXi Nitric Oxide Delivery SystemPraxair, Inc. · 2020 · Cites it directly · ◆ in the 2020 descendants bar of FIG. 12 Class I recalls
  • K231823NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 12 Class I recalls
  • K233251NOxBOXi Nitric Oxide Delivery SystemLinde Gas & Equipment, Inc. · 2024 · 3 generations on · ◆ in the 2024 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 17 years and passes through 4 companies; a new generation was cleared every 3.2 years on median.
  • 2 of the 9 ancestors have a recall of any class.
  • Of its descendants, K201339 (Praxair, Inc., 2020) has gone furthest, with 3 descendants of its own.
  • 6 other devices also cite its predicate K092545.

How does its category compare?

All MRN devices

MRN Apparatus, Nitric Oxide Delivery: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · MRN

2in 2025

2007: 22008: 22009: 12010: 22011: 22012: 32013: 22015: 12018: 12020: 32022: 12023: 42024: 22025: 2
2007high 4 (2023) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MRN

566in 2023

2012: 117 days median2020: 84 days median2023: 566 days median
2007high 566 (2023) · low 84 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MRN

2in 2025

2010: 22014: 12023: 12025: 2, 2 Class I2026 (partial year): 2
2007high 2 (2010) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · MRN

155in 2025

2007: 32008: 92009: 102010: 1552011: 552012: 242013: 322014: 792015: 1562016: 942017: 1062018: 1382019: 1442020: 1992021: 2212022: 1152023: 992024: 1382025: 155
2007high 221 (2021) · low 3 (2007)2025

MAUDE reports naming product code MRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—03
20082—09
20091—010
20102—2155
20112—055
20123117024
20132—032
20140—179
20151—0156
20160—094
20170—0106
20181—0138
20190—0144
20203840199
20210—0221
20221—0115
20234566199
20242—0138
20252—2 (2)155
20260—2—

What is it indicated for?

The NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO). The NOxBOXi Nitric Oxide Delivery System Includes: • The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode. • Continuous…

Read the full statement (158 words)

The NOxBOXi Nitric Oxide Delivery System is intended for use by healthcare professionals for the delivery and

monitoring of a constant (user set) concentration of nitric oxide (NO) and the monitoring of NO2 and O2 in the inspiratory ventilator lines of a patient undergoing nitric oxide therapy (iNO).

The NOxBOXi Nitric Oxide Delivery System Includes:

• The NOxBOXi head unit, which delivers NO gas while in the intelligent delivery mode.

• Continuous monitoring and alarms for NO, O2, and NO2.

• The integrated NOxMixer, which provides a backup NO delivery capability that is intended to deliver a continuous

flow of NO, mixed with O2, for iNO therapy and provides a continuous treatment option during transit and transfer

conditions within hospitals.

The NOxBOXi Nitric Oxide Delivery System must only be used in accordance with the indications, contraindications,

warnings and precautions described in the nitric oxide drug packaging inserts and labeling (currently neonates). Refer to

this material prior to use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K171696: NOxBOXi Nitric Oxide Delivery System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K171696

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026