510(k) Report

510(k) · Cleared

Cleared

VERIFY® HPU Chemical Indicator and VERIFY® Vaporized VH2O2 Process Indicator Adhesive Label

K172746510(k) AbbreviatedCleared Feb 9, 2018150-day reviewClass II21 CFR 880.2800

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · JOJ Indicator, Physical/Chemical Sterilization Process

Cleared on Feb 9, 2018 as substantially equivalent to K140515 (STERIS Corporation, 2014) after a 150-day review. Its family tree reaches back 4 generations to 2002. 6 later devices descend from it.

Decided
Feb 9, 2018
Review
150 days
Class
Class II
Product code
JOJ
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K994055 (2002). 6 later devices descend from it, 2 citing it directly.

  1. Root · 4 generations backK994055 Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products · 2002 · Recalled
  2. 3 generations backK072510 Verify V-Pro Chemical Indicator, Versions 1 and 2STERIS Corporation · 2007
  3. 2 generations backK091174 Verify V-Pro Chemical Indicator, Version 1A: Verify V-Pro Chemical…STERIS Corporation · 2009
  4. 1 generation backK140515 Verify Hpu Chemical Indicator/ Verify Vaporized VH202 Process…STERIS Corporation · 2014
  5. This deviceK172746 VERIFY® HPU Chemical Indicator and VERIFY® Vaporized VH2O2 Process…STERIS Corporation · 2018
  6. Descendant · cites it directly · 5 of its ownK183295 Celerity HP Chemical Indicator (CI)Steris Corporations · 2019
  7. Descendant · 2 generations on · 4 of its ownK192020 Celerity HP Chemical Indicator, Celerity Vaporized VH2O2 Process…STERIS Corporation · 2019
  8. Descendant · cites it directly · 2 of its ownK231488 Celerity™ HP Chemical Indicator; Celerity™ HP Multivariable Chemical…Steris · 2023
  9. Descendant · 2 generations on · 1 of its ownK240760 CELERITY HP Indicator TapeSteris · 2024
  10. Descendant · 3 generations on · 0 of its ownK242189 CELERITY HP Indicator Tape (PCC077)Steris · 2024

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2019: 22023: 22024: 2
2019high 2 (2019) · low 0 (2020)2026

Most influential descendants

Showing 5 of 6 · See all
  • K183295Celerity HP Chemical Indicator (CI)Steris Corporations · 2019 · cites it directly5descendants
  • K192020Celerity HP Chemical Indicator, Celerity Vaporized VH2O2 Process Indicator Adhesive…STERIS Corporation · 2019 · 2 generations on4descendants
  • K231488Celerity™ HP Chemical Indicator; Celerity™ HP Multivariable Chemical Indicator; VERIFY…Steris · 2023 · cites it directly2descendants
  • K240760CELERITY HP Indicator TapeSteris · 2024 · 2 generations on1descendant
  • K242189CELERITY HP Indicator Tape (PCC077)Steris · 2024 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K994055Sterrad Cyclesure Biological IndicatorAdvanced Sterilization Products20021
−3K072510Verify V-Pro Chemical Indicator, Versions 1 and 2STERIS Corporation2007—
−2K091174Verify V-Pro Chemical Indicator, Version 1A: Verify V-Pro Chemical Indicator…STERIS Corporation2009—
−1K140515Verify Hpu Chemical Indicator/ Verify Vaporized VH202 Process Indicator…STERIS Corporation2014—
0 (this device)K172746VERIFY® HPU Chemical Indicator and VERIFY® Vaporized VH2O2 Process Indicator…STERIS Corporation2018—
+1K183295Celerity HP Chemical Indicator (CI)Steris Corporations2019—
+2K192020Celerity HP Chemical Indicator, Celerity Vaporized VH2O2 Process Indicator…STERIS Corporation2019—
+1K231488Celerity™ HP Chemical Indicator; Celerity™ HP Multivariable Chemical…Steris2023—
+2K240760CELERITY HP Indicator TapeSteris2024—
+3K242189CELERITY HP Indicator Tape (PCC077)Steris2024—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201920
202320
202420
FIG. 1 — Lineage of K172746, 2002–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140515Verify Hpu Chemical Indicator/ Verify Vaporized VH202 Process Indicator Adhesive Label
Summary page 6
…ess indicator (21 CFR 880.2800, Product Code JOJ) Class: II 2. Predicate Device VERIFY® V-PRO Chemical Indicator (K140515) 3. Device Description The VERIFY® V-PRO Chemical Indicator is used in each processing cycle to indicate exposure…
STERIS Corporation2014—

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K183295Celerity HP Chemical Indicator (CI)Steris Corporations20195
K231488Celerity™ HP Chemical Indicator; Celerity™ HP Multivariable Chemical Indicator; VERIFY…Steris20232

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 2 companies; a new generation was cleared every 4.7 years on median.
  • 1 of the 4 ancestors has a recall of any class.
  • Of its descendants, K183295 (Steris Corporations, 2019) has gone furthest, with 5 descendants of its own.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

The VERIFY® HPU Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units when placed within packs to be sterilized to indicate, through a visible change from magenta to yellow, when the device has been exposed to the Lumen, Non Lumen, Flexible or Fast Non Lumen sterilization cycle of a V-PRO® Low Temperature Sterilization System. The VERIFY® Vaporized…

Read the full statement (135 words)

The VERIFY® HPU Chemical Indicator is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units when placed within packs to be sterilized to indicate, through a visible change from magenta to yellow, when the device has been exposed to the Lumen, Non Lumen, Flexible or Fast Non Lumen sterilization cycle of a V-PRO® Low Temperature Sterilization System.

The VERIFY® Vaporized VH2O2 Process Indicator Adhesive Label is a Type 1 vaporized hydrogen peroxide sterilization process indicator. It is designed to distinguish between processed and unprocessed units, when affixed to packs to be sterilized, through a visible color change from magenta to yellow, when the pack has been exposed to the Lumen, Non-Lumen, Flexible or Fast Non Lumen sterilization cycle of a V-PRO® Low Temperature Sterilization System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K172746: VERIFY® HPU Chemical Indicator and VERIFY® Vaporized VH2O2 Process Indicator Adhesive Label.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172746

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026