510(k) Report

510(k) · Cleared

Recalled · Class III · 2022

Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles

K173626510(k) TraditionalCleared Feb 21, 201889-day reviewClass II21 CFR 880.2800Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · FRC Indicator, Biological Sterilization Process

Cleared on Feb 21, 2018 as substantially equivalent to K121593 (3M Company, 2013) after an 89-day review. Its family tree reaches back 4 generations to 1995. It has been recalled once.

Decided
Feb 21, 2018
Review
89 days
Class
Class II
Product code
FRC
Recalls
1
Predicates
1

Its longest line of descent runs back 4 generations to K926364 (1995). 8 ancestors in all, the oldest K900771 (1990). No later device cites it as a predicate yet.

  1. Root · 4 generations backK926364 Attest Brand Rapid Readout Biological Ind. #129X3M Health Care, Ltd. · 1995
  2. 3 generations backK090569 Modification to: Attest Rapid Readout Biological Indicator, Mode 12923M Company · 2009
  3. 2 generations backK121484 3M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest…3M Company · 2012
  4. 1 generation backK121593 3M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest…3M Company · 2013
  5. This deviceK173626 Celerity 20 Steam Process Challenge Device for Dynamic Air Removal…STERIS Corporation · 2018
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K926364Attest Brand Rapid Readout Biological Ind. #129X3M Health Care, Ltd.1995—
−3K090569Modification to: Attest Rapid Readout Biological Indicator, Mode 12923M Company2009—
−2K1214843M Attest (TM) Super Rapid Readout Biological Indicator, 3M Attest (TM)…3M Company2012—
−1K1215933M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest (TM) Super…3M Company2013—
0 (this device)K173626Celerity 20 Steam Process Challenge Device for Dynamic Air Removal CyclesSTERIS Corporation2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199410
200210
201110
FIG. 1 — Lineage of K173626, 1995–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRC device

  1. #1 K250044 Celerity 5 HP Biological Indicator (LCB052)…score 14.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K1215933M Attest (TM) Super Radid Readout Steam Challenge Pack 3M Attest (TM) Super Rapid 5…
Summary page 5
…Sterilization Process Indicator (21 CFR 880.2800, FRC) 2. Predicate Device 3M Attest Super Rapid 5 Steam-Plus Challenge Pack K121593 Reference Device: Verify Assert Steam Process Challenge Device, K162945 3. Description of Device The Celerity 20…
3M Company2013—
†K162945VERIFY ASSERT STEAM Process Challenge DeviceSTERIS Corporation2017—

Also cleared against K121593: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 25, 2022Class IIIZ-1793-2022Environmental controlFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 3 companies; a new generation was cleared every 4.9 years on median.
  • 1 of the 8 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K121593.

How does its category compare?

All FRC devices

FRC Indicator, Biological Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · FRC

6in 2025

2007: 32008: 12009: 62010: 42011: 62012: 92013: 42014: 132015: 32016: 22017: 102018: 182019: 52020: 102021: 52022: 52023: 62024: 82025: 62026 (partial year): 1
2007high 18 (2018) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRC

29in 2025

2007: 83 days median2009: 183 days median2010: 165 days median2011: 118 days median2012: 178 days median2013: 260 days median2014: 174 days median2015: 130 days median2017: 130 days median2018: 97 days median2019: 88 days median2020: 150 days median2021: 107 days median2022: 193 days median2023: 190 days median2024: 105 days median2025: 29 days median
2007high 260 (2013) · low 29 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRC

8so far in 2026, the most since 2007

2009: 12012: 22013: 22022: 42023: 12026 (partial year): 8
2007high 8 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRC

9in 2025

2007: 1002008: 602009: 452010: 572011: 372012: 462013: 392014: 512015: 1822016: 4822017: 1822018: 222019: 302020: 162021: 102022: 92023: 202024: 222025: 9
2007high 482 (2016) · low 9 (2022)2025

MAUDE reports naming product code FRC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073830100
20081—060
20096183145
20104165057
20116118037
20129178246
20134260239
201413174051
201531300182
20162—0482
2017101300182
20181897022
2019588030
202010150016
20215107010
2022519349
20236190120
20248105022
202562909
20261—8—

What is it indicated for?

The Celerity™ 20 Steam Challenge Pack is used for qualification, routine microbial monitoring, and load monitoring of steam sterilizers.

The validated steam sterilization cycles include:

• 270°F (132°C) 4-minute dynamic air removal (Prevac, SFPP)

• 275°F (135°C) 3-minute dynamic air removal (Prevac, SFPP)

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K173626: Celerity 20 Steam Process Challenge Device for Dynamic Air Removal Cycles.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173626

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026