Recalled · Class I · 2014Hawkeye 4 Option for Dual-Head Variable Angle Gamma Camera
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
GE Medical Systems · KPS System, Tomography, Computed, Emission
Cleared on Sep 27, 2005 as substantially equivalent to K991841 (Ge Medical Systems F.I. Haifa, 1999) after a 21-day review. Its family tree reaches back 3 generations to 1995. 9 of the 20 devices in its family, counting itself, have a Class I recall. 16 later devices descend from it. It has been recalled 8 times, including a Class I recall.
- Decided
- Sep 27, 2005
- Review
- 21 days
- Class
- Class II
- Product code
- KPS
- Recalls
- 8 (1 Class I)
- Predicates
- 1
Its longest line of descent runs back 3 generations to K952674 (1995). 16 later devices descend from it, 3 citing it directly. 9 of the 20 devices in this family, counting this one, have a Class I recall.
- Root · 3 generations backK952674 Transact: Attenuation Correction System for the Apex Cardial Gamma…Elscint, Inc. · 1995 · Recalled
- 2 generations backK980959 Vtransact: Attenuation Correction System for Dual-Head Variable…Ge Medical Systems F.I. Haifa · 1998
- 1 generation backK991841 Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Recalled (Class I)
- This deviceK052434 Hawkeye 4 Option for Dual-Head Variable Angle Gamma CameraGE Medical Systems · 2005
- Descendant · cites it directly · 13 of its ownK093514 Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Recalled (Class I)
- Descendant · 2 generations on · 6 of its ownK153402 Discovery NM/CT 670 CZTGE Healthcare · 2016 · Recalled (Class I)
- Descendant · 3 generations on · 3 of its ownK182484 VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018
- Descendant · 4 generations on · 2 of its ownK190457 VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2019
- Descendant · 2 generations on · 2 of its ownK111445 Discovery NM 630GE Healthcare · 2011 · Recalled (Class I)
Showing 5 of 16 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 16 · See all- K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · cites it directly13descendants
- K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on6descendants
- K182484VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2018 · 3 generations on3descendants
- K190457VERITON CT whole body SPECT/CT systemSpectrum Dynamics Medical, Ltd. · 2019 · 4 generations on2descendants
- K111445Discovery NM 630GE Healthcare · 2011 · 2 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −3 | K952674 | Transact: Attenuation Correction System for the Apex Cardial Gamma Camera | Elscint, Inc. | 1995 | 1 |
| −2 | K980959 | Vtransact: Attenuation Correction System for Dual-Head Variable Angle Gamma… | Ge Medical Systems F.I. Haifa | 1998 | — |
| −1 | K991841 | Hawkeye Option for Dual-Head Variable Angle Gamma Camera | Ge Medical Systems F.I. Haifa | 1999 | Class I |
| 0 (this device) | K052434 | Hawkeye 4 Option for Dual-Head Variable Angle Gamma Camera | GE Medical Systems | 2005 | — |
| +1 | K093514 | Discovery Model NM/CT 670 | Ge Medical Systems F.I. Haifa | 2009 | Class I |
| +2 | K153402 | Discovery NM/CT 670 CZT | GE Healthcare | 2016 | Class I |
| +3 | K182484 | VERITON CT whole body SPECT/CT system | Spectrum Dynamics Medical, Ltd. | 2018 | — |
| +4 | K190457 | VERITON CT whole body SPECT/CT system | Spectrum Dynamics Medical, Ltd. | 2019 | — |
| +2 | K111445 | Discovery NM 630 | GE Healthcare | 2011 | Class I |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 1 | 1 |
| 2009 | 2 | 1 |
| 2011 | 2 | 2 |
| 2012 | 1 | 1 |
| 2016 | 1 | 1 |
| 2018 | 2 | 1 |
| 2019 | 1 | 0 |
| 2021 | 1 | 0 |
| 2023 | 1 | 0 |
| 2025 | 3 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
9 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KPS device
- #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K991841 | Hawkeye Option for Dual-Head Variable Angle Gamma CameraSummary page 2…Head Variable Angle Gamma Camera" , Radiological, 21 CFR 892.1200, 90-KPS Marketed Device: "Hawkeye Option for Dual-Head Variable Angle Gamma Camera" -K991841 Radiological, 21 CFR 892.1200, 90-KPS Device Description: The "Hawkeye 4 Option" is a modification of the "Ha… | 1999 |
What descends from it? (16)
See all 163 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (9 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Oct 14, 2014 | Class II | Z-0185-2015 | Nonconforming Material/Component | GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative… |
| Oct 14, 2014 | Class II | Z-0186-2015 | Nonconforming Material/Component | GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative… |
| Oct 14, 2014 | Class II | Z-0188-2015 | Nonconforming Material/Component | GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative… |
| Oct 14, 2014 | Class II | Z-0189-2015 | Nonconforming Material/Component | GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative… |
| Oct 14, 2014 | Class II | Z-0187-2015 | Nonconforming Material/Component | GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems. This check is being added to prevent a portion of the system… |
| Oct 14, 2014 | Class II | Z-0184-2015 | Nonconforming Material/Component | GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative… |
| Jun 13, 2013 | Class I | Z-1787-2013 | Device Design | GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System. |
| Mar 6, 2012 | Class II | Z-1078-2013 | Process control | Accelerated fatigue of the lateral axis motion subsystem of the Infinia Computed Emission Tomography System could result in mechanical failure causing the detector to slide; may impact patient and operators. |
The 5 latest of 8 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K991841Hawkeye Option for Dual-Head Variable Angle Gamma CameraGe Medical Systems F.I. Haifa · 1999 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
- K080927Brightview VCT Imaging SystemPhilips Medical Systems (Cleveland), Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
- K093514Discovery Model NM/CT 670Ge Medical Systems F.I. Haifa · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 12 Class I recalls
- K121019Optima NM/CT 640GE Healthcare · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls
- K153402Discovery NM/CT 670 CZTGE Healthcare · 2016 · 2 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
- K173816NM/CT 850, NM/CT 860GE Healthcare · 2018 · 2 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
- K111445Discovery NM 630GE Healthcare · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
- K113350Brivo NM 615GE Healthcare · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls
Nearest 5 of 8; all are on the web page.
What does this lineage say?
- The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 2.7 years on median.
- 2 of the 3 ancestors have a recall of any class; 1 has a Class I recall.
- Of its descendants, K093514 (Ge Medical Systems F.I. Haifa, 2009) has gone furthest, with 13 descendants of its own.
- 5 other devices also cite its predicate K991841.
How does its category compare?
All KPS devicesKPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KPS
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KPS
65in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KPS
16in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · KPS
36in 2025
MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 8 | 38 | 22 | 38 |
| 2008 | 17 | 25 | 16 | 11 |
| 2009 | 9 | 42 | 14 | 15 |
| 2010 | 3 | 91 | 4 | 44 |
| 2011 | 7 | 28 | 34 | 56 |
| 2012 | 6 | 94 | 13 | 34 |
| 2013 | 11 | 84 | 13 | 240 |
| 2014 | 11 | 92 | 18 | 33 |
| 2015 | 3 | 112 | 5 | 24 |
| 2016 | 11 | 65 | 16 | 10 |
| 2017 | 11 | 104 | 11 | 31 |
| 2018 | 13 | 59 | 12 (3) | 26 |
| 2019 | 3 | 60 | 5 (5) | 105 |
| 2020 | 12 | 72 | 8 | 22 |
| 2021 | 15 | 66 | 4 | 10 |
| 2022 | 5 | 60 | 8 (6) | 15 |
| 2023 | 12 | 63 | 6 (3) | 22 |
| 2024 | 11 | 59 | 15 | 23 |
| 2025 | 12 | 65 | 16 | 36 |
| 2026 | 6 | 67 | 6 | — |
What is it indicated for?
The Intended use of the device is to produce attenuation- corrected NM Images. The attenuation maps are also displayed with the NM images to facilitate the localization of the NM activity in the patient anatomy.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K052434 · printed Nov 6, 2026
Cite this page
510(k) Report. “K052434: Hawkeye 4 Option for Dual-Head Variable Angle Gamma Camera.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K052434