TruDi NAV Suction Instruments

K180948510(k) TraditionalCleared Jul 20, 2018100-day reviewClass II21 CFR 882.4560Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Acclarent, Inc. · PGW Ear, Nose, And Throat Stereotaxic Instrument

Cleared on Jul 20, 2018 as substantially equivalent to K161555 (Karl Storz Endoscopy America, Inc., 2016) and 1 other predicate after a 100-day review. Its family tree reaches back 5 generations to 1994. 3 of the 21 devices in its family, counting itself, have a Class I recall. It has been recalled 5 times.

Decided
Jul 20, 2018
Review
100 days
Class
Class II
Product code
PGW
Recalls
5
Predicates
2

Its longest line of descent runs back 5 generations to K911783 (1994). 20 ancestors in all, the oldest K871046 (1987). No later device cites it as a predicate yet. 3 of the 21 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2016: 1 other ancestor2012: 1 other ancestor2005: 1 other ancestor2000: 1 other ancestor2002: 1 other ancestor1998: 2 other ancestors, 2 with a Class I recall21996: 2 other ancestors, 1 with a Class I recall12001: 1 other ancestor1999: 1 other ancestor1995: 3 other ancestors1987: 1 other ancestorK911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K133573 · Fiagon Navigation System · Fiagon GmbH · 2014K161555 · KARL STORZ NAV1 Electromagnetic Navigation System · Karl Storz Endoscopy America, Inc. · 201619851990199520002005201020152020
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 3 other ancestors, 2 with a Class I recall24 generations back: 6 other ancestors, 1 with a Class I recall15 generations back: 2 other ancestors K161555 · KARL STORZ NAV1 Electromagnetic Navigation System · Karl Storz Endoscopy America, Inc. · 2016K133573 · Fiagon Navigation System · Fiagon GmbH · 2014K964229 · Regulus Navigator · Compass Intl., Inc. · 1997K960714 · Isg Viewing Wand · I.S.G. Technologies, Inc. · 1996K911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994−50
  1. Root · 5 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 4 generations backK960714 Isg Viewing WandI.S.G. Technologies, Inc. · 1996
  3. 3 generations backK964229 Regulus NavigatorCompass Intl., Inc. · 1997
  4. 2 generations backK133573 Fiagon Navigation SystemFiagon GmbH · 2014
  5. 1 generation backK161555 KARL STORZ NAV1 Electromagnetic Navigation SystemKarl Storz Endoscopy America, Inc. · 2016
  6. This deviceK180948 TruDi NAV Suction InstrumentsAcclarent, Inc. · 2018

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−4K960714Isg Viewing WandI.S.G. Technologies, Inc.1996—
−3K964229Regulus NavigatorCompass Intl., Inc.1997—
−2K133573Fiagon Navigation SystemFiagon GmbH2014—
−1K161555KARL STORZ NAV1 Electromagnetic Navigation SystemKarl Storz Endoscopy America, Inc.2016—
0 (this device)K180948TruDi NAV Suction InstrumentsAcclarent, Inc.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199530
199621
199822
199910
200010
200110
200210
200510
201210
201610
FIG. 1 — Lineage of K180948, 1994–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PGW device

  1. #1 K261540 Cube 4D Navigation System, VirtuEye Proscore 12.9 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161555KARL STORZ NAV1 Electromagnetic Navigation System
Summary page 4
…iew Panel: Ear, Nose, and Throat Product Code: PGW [807.92(a)(3)] Legally Marketed Devices Predicate Device: KARL STORZ EM Navigated Suction Tubes, (K161555) (cleared as part of the KARL STORZ NAV1 Electromagnetic Navigation System under K161…
Karl Storz Endoscopy America, Inc.2016—
*K153555EM ENT Navigated Suctions
Summary page 4
…cleared as part of the KARL STORZ NAV1 Electromagnetic Navigation System under K161555) Reference Device: Medtronic EM ENT Navigated Suctions, (K153555) [807.92(a)(4)] Device Description Device Description: TruDi™ NAV Suction Instruments (0°, 70°, 90°) are electromagnetically-…
Medtronic Navigation, Inc.20168

Also cleared against K161555: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 9, 2024Class IIZ-2507-2024Process controlSpecific lots of TruDi NAV Suction Instruments 0 Degrees were incorrectly calibrated which when used with the TruDi Navigation System, may result in a discrepancy between the actual tip location and the visualized location displayed on…
Nov 12, 2021Class IIZ-0415-2022Process controlSuction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead…
Mar 6, 2019Class IIZ-0824-2020Device DesignThe firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
Mar 6, 2019Class IIZ-0825-2020Device DesignThe firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
Mar 6, 2019Class IIZ-0826-2020Device DesignThe firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 3 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 4 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 24 years and passes through 5 companies; a new generation was cleared every 2.1 years on median.
  • 7 of the 20 ancestors have a recall of any class; 3 have a Class I recall.
  • 1 other device also cites its predicate K161555.

How does its category compare?

All PGW devices

PGW Ear, Nose, And Throat Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2018, this device’s year, is highlighted.

FDA decisions per year · PGW

1in 2025

2014: 42015: 12016: 92017: 42018: 22019: 12020: 82021: 12022: 22023: 22025: 12026 (partial year): 3
2007high 9 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PGW

100in 2020

2014: 116 days median2016: 115 days median2017: 125 days median2020: 100 days median2026 (partial year): 65 days median
2007high 125 (2017) · low 100 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PGW

1in 2025

2019: 32021: 12023: 12025: 1
2007high 3 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PGW

220in 2025

2014: 72015: 142016: 332017: 422018: 962019: 2472020: 1612021: 1902022: 1662023: 2002024: 2742025: 220
2007high 274 (2024) · low 0 (2007)2025

MAUDE reports naming product code PGW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
2014411607
20151—014
20169115033
20174125042
20182—096
20191—3247
202081000161
20211—1190
20222—0166
20232—1200
20240—0274
20251—1220
20263650—

What is it indicated for?

TruDi™ NAV Suction Instruments are intended for use with the TruDi™ Navigation System during surgical procedures in ENT medicine and skull base surgery to provide navigation of the instruments to the targeted anatomy, and evacuation of gases, liquids, and fragments.

TruDi™ NAV Suction Instruments are intended to be used by ENT surgeons or support staff.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K180948: TruDi NAV Suction Instruments.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K180948

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026