510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Tablo Hemodialysis System, Tablo Cartridge

K190793510(k) TraditionalCleared Dec 12, 2019259-day reviewClass II21 CFR 876.5860Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Outset Medical, Inc. · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Dec 12, 2019 as substantially equivalent to K160881 (Outset Medical, Inc., 2016) and 1 other predicate after a 259-day review. Its family tree reaches back 10 generations to 1994. 2 later devices descend from it. It has been recalled 2 times.

Decided
Dec 12, 2019
Review
259 days
Class
Class II
Product code
KDI
Recalls
2
Predicates
2

Its longest line of descent runs back 10 generations to K921456 (1994). 16 ancestors in all. 2 later devices descend from it, 1 citing it directly.

  1. Root · 10 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 6 generations between
  3. 3 generations backK140866 Tablo ConsoleHome Dialysis Plus · 2014
  4. 2 generations backK150880 Tablo Console with Wireless CapabilityOutset Medical · 2015
  5. 1 generation backK160881 Tablo ConsoleOutset Medical, Inc. · 2016
  6. This deviceK190793 Tablo Hemodialysis System, Tablo CartridgeOutset Medical, Inc. · 2019
  7. Descendant · cites it directly · 1 of its ownK210782 Tablo CartridgeOutset Medical, Inc. · 2021
  8. Descendant · 2 generations on · 0 of its ownK233798 Moda-flx Hemodialysis System and CartridgeDiality, Inc. · 2024
Show all 10 generations
  1. Root · 10 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 9 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 8 generations backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
  4. 7 generations backK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010 · Recalled
  5. 6 generations backK111639 Fresenius 2008T Hemodialysis MachineFresenius Medical Care · 2011 · Recalled
  6. 5 generations backK113427 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America · 2011 · Recalled
  7. 4 generations backK120505 Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente · 2012 · Recalled
  8. 3 generations backK140866 Tablo ConsoleHome Dialysis Plus · 2014
  9. 2 generations backK150880 Tablo Console with Wireless CapabilityOutset Medical · 2015
  10. 1 generation backK160881 Tablo ConsoleOutset Medical, Inc. · 2016
  11. This deviceK190793 Tablo Hemodialysis System, Tablo CartridgeOutset Medical, Inc. · 2019

Most influential descendants

Showing 2 of 2 · See all
  • K210782Tablo CartridgeOutset Medical, Inc. · 2021 · cites it directly1descendant
  • K233798Moda-flx Hemodialysis System and CartridgeDiality, Inc. · 2024 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−9K994267Fresenius 2008KFresenius Medical Care20005
−8K080964Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America20094
−7K093902Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America20102
−6K111639Fresenius 2008T Hemodialysis MachineFresenius Medical Care20111
−5K113427Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America20114
−4K120505Fresenius 2008T Hemodialysis MachineFresenius Medical Care North America, Design Cente20124
−3K140866Tablo ConsoleHome Dialysis Plus2014—
−2K150880Tablo Console with Wireless CapabilityOutset Medical2015—
−1K160881Tablo ConsoleOutset Medical, Inc.2016—
0 (this device)K190793Tablo Hemodialysis System, Tablo CartridgeOutset Medical, Inc.2019—
+1K210782Tablo CartridgeOutset Medical, Inc.2021—
+2K233798Moda-flx Hemodialysis System and CartridgeDiality, Inc.2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200110
200310
200910
201210
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202110
202410
FIG. 1 — Lineage of K190793, 1994–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K160881Tablo Console
Summary page 4
…fication System for Hemodialysis – Product Code: FIP Hemodialysis system and accessories – Product Code: FJK Regulatory Class: II III. PREDICATE DEVICE Tablo Console, K160881; Tablo Cartridge, K140841 These predicates have not been subject to a design related recall. No reference devices were used in…
Outset Medical, Inc.2016—
*K140841Tablo Cartridge
Summary page 4
…alysis – Product Code: FIP Hemodialysis system and accessories – Product Code: FJK Regulatory Class: II III. PREDICATE DEVICE Tablo Console, K160881; Tablo Cartridge, K140841 These predicates have not been subject to a design related recall. No reference devices were used in this submission. === Page…
Home Dialysis Plus2014—

Also cleared against K160881: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K210782Tablo CartridgeOutset Medical, Inc.20211

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 24, 2023Class IIZ-1351-2023Process controlAn observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.
Jan 20, 2022Class IIZ-0813-2022Device DesignDue to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 7 companies; a new generation was cleared every 1.5 years on median.
  • 7 of the 16 ancestors have a recall of any class.
  • Of its descendants, K210782 (Outset Medical, Inc., 2021) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K160881.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

The Tablo® Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician’s prescription, with a trained individual available as needed who is considered competent in the use of the device by the prescribing physician.

The Tablo® Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extra-corporeal access during hemodialysis. The Tablo Cartridge is compatible only with the Tablo Hemodialysis System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K190793: Tablo Hemodialysis System, Tablo Cartridge.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K190793

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026