510(k) Report

510(k) · Cleared

Cleared

Vantage Galan 3T, MRT-3020, V6.0 with AiCE Reconstruction Processing Unit for MR

K192574510(k) TraditionalCleared Mar 9, 2020172-day reviewClass II21 CFR 892.1000

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Canon Medical Systems Corporation · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Mar 9, 2020 as substantially equivalent to K181593 (Canon Medical Systems Corporation, 2018) after a 172-day review. Its family tree reaches back 15 generations to 2000. 9 later devices descend from it.

Decided
Mar 9, 2020
Review
172 days
Class
Class II
Product code
LNH
Recalls
None
Predicates
1

Its longest line of descent runs back 15 generations to K002531 (2000). 9 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2021: 4 descendants cleared2022: 2 descendants cleared2023: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant clearedK002531 · Excelart with Pianissimo, Model MRT-1501/P3 · Toshiba America Mri, Inc. · 2000 · recalledK023511 · Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2 · Toshiba America Medical Systems, In.C · 2002 · recalledK032490 · Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and MRT-1503/P2 · Toshiba America Medical Systems, In.C · 2003 · recalledK060003 · Excelart Vantage ZGV · Toshibamedical Systems Corporation · 2006 · recalledK063361 · Excelart Vantage Atlas, Model MRT-1503 · Toshiba America Medical Systems, In.C · 2006 · recalledK080038 · Vantage Titan, Model MRT-1504/S3 · Toshiba America Medical Systems, In.C · 2008 · recalledK102489 · Vantage Titan 3T · Toshiba Medical Systems Corporation, Japan · 2011 · recalledK113093 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2012 · recalledK120487 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2012K132160 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2013K143008 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2015K152371 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2015 · recalledK162183 · Vantage Galan 3T, MRT-3020, V4.0 · Toshibamedical Systems Corporation · 2016K173382 · Vantage Galan 3T, V4.6 · Toshibamedical Systems Corporation · 2018K181593 · Vantage Galan 3T, MRT-3020/A7, V5.0 · Canon Medical Systems Corporation · 2018200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 1 descendant K181593 · Vantage Galan 3T, MRT-3020/A7, V5.0 · Canon Medical Systems Corporation · 2018K173382 · Vantage Galan 3T, V4.6 · Toshibamedical Systems Corporation · 2018K162183 · Vantage Galan 3T, MRT-3020, V4.0 · Toshibamedical Systems Corporation · 2016K152371 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2015 · recalledK143008 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2015K132160 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2013K120487 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2012K113093 · Vantage Titan 3T · Toshibamedical Systems Corporation · 2012 · recalledK102489 · Vantage Titan 3T · Toshiba Medical Systems Corporation, Japan · 2011 · recalledK080038 · Vantage Titan, Model MRT-1504/S3 · Toshiba America Medical Systems, In.C · 2008 · recalledK063361 · Excelart Vantage Atlas, Model MRT-1503 · Toshiba America Medical Systems, In.C · 2006 · recalledK060003 · Excelart Vantage ZGV · Toshibamedical Systems Corporation · 2006 · recalledK032490 · Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and MRT-1503/P2 · Toshiba America Medical Systems, In.C · 2003 · recalledK023511 · Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2 · Toshiba America Medical Systems, In.C · 2002 · recalledK002531 · Excelart with Pianissimo, Model MRT-1501/P3 · Toshiba America Mri, Inc. · 2000 · recalled−15−10−50+5
  1. Root · 15 generations backK002531 Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc. · 2000 · Recalled
  2. 11 generations between
  3. 3 generations backK162183 Vantage Galan 3T, MRT-3020, V4.0Toshibamedical Systems Corporation · 2016
  4. 2 generations backK173382 Vantage Galan 3T, V4.6Toshibamedical Systems Corporation · 2018
  5. 1 generation backK181593 Vantage Galan 3T, MRT-3020/A7, V5.0Canon Medical Systems Corporation · 2018
  6. This deviceK192574 Vantage Galan 3T, MRT-3020, V6.0 with AiCE Reconstruction Processing…Canon Medical Systems Corporation · 2020
  7. Descendant · cites it directly · 6 of its ownK203553 Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing…Canon Medical Systems Corporation · 2021 · Recalled
  8. Descendant · 2 generations on · 5 of its ownK212056 Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing…Canon Medical Systems Corporation · 2021
  9. Descendant · 3 generations on · 4 of its ownK220192 Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing…Canon Medical Systems Corporation · 2022
  10. Descendant · 4 generations on · 3 of its ownK222387 Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing…Canon Medical Systems Corporation · 2022
  11. Descendant · 5 generations on · 2 of its ownK230355 Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing…Canon Medical Systems Corporation · 2023 · Recalled
Show all 15 generations
  1. Root · 15 generations backK002531 Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc. · 2000 · Recalled
  2. 14 generations backK023511 Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2Toshiba America Medical Systems, In.C · 2002 · Recalled
  3. 13 generations backK032490 Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models…Toshiba America Medical Systems, In.C · 2003 · Recalled
  4. 12 generations backK060003 Excelart Vantage ZGVToshibamedical Systems Corporation · 2006 · Recalled
  5. 11 generations backK063361 Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C · 2006 · Recalled
  6. 10 generations backK080038 Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C · 2008 · Recalled
  7. 9 generations backK102489 Vantage Titan 3TToshiba Medical Systems Corporation, Japan · 2011 · Recalled
  8. 8 generations backK113093 Vantage Titan 3TToshibamedical Systems Corporation · 2012 · Recalled
  9. 7 generations backK120487 Vantage Titan 3TToshibamedical Systems Corporation · 2012
  10. 6 generations backK132160 Vantage Titan 3TToshibamedical Systems Corporation · 2013
  11. 5 generations backK143008 Vantage Titan 3TToshibamedical Systems Corporation · 2015
  12. 4 generations backK152371 Vantage Titan 3TToshibamedical Systems Corporation · 2015 · Recalled
  13. 3 generations backK162183 Vantage Galan 3T, MRT-3020, V4.0Toshibamedical Systems Corporation · 2016
  14. 2 generations backK173382 Vantage Galan 3T, V4.6Toshibamedical Systems Corporation · 2018
  15. 1 generation backK181593 Vantage Galan 3T, MRT-3020/A7, V5.0Canon Medical Systems Corporation · 2018
  16. This deviceK192574 Vantage Galan 3T, MRT-3020, V6.0 with AiCE Reconstruction Processing…Canon Medical Systems Corporation · 2020

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2021: 42022: 22023: 12024: 12025: 1
2021high 4 (2021) · low 0 (2026)2026

Most influential descendants

Showing 5 of 9 · See all
  • K203553Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation · 2021 · cites it directly6descendants
  • K212056Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation · 2021 · 2 generations on5descendants
  • K220192Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation · 2022 · 3 generations on4descendants
  • K222387Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation · 2022 · 4 generations on3descendants
  • K230355Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation · 2023 · 5 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−15K002531Excelart with Pianissimo, Model MRT-1501/P3Toshiba America Mri, Inc.20004
−14K023511Excelart with Pianissimo XG/AG Spin Edition, Model MRT-1501/P3, P2Toshiba America Medical Systems, In.C20022
−13K032490Toshiba Excelart Vantage XGV/Agv Mri Diagnostic System, Models MRT-1503/P3 and…Toshiba America Medical Systems, In.C20033
−12K060003Excelart Vantage ZGVToshibamedical Systems Corporation20061
−11K063361Excelart Vantage Atlas, Model MRT-1503Toshiba America Medical Systems, In.C20061
−10K080038Vantage Titan, Model MRT-1504/S3Toshiba America Medical Systems, In.C20082
−9K102489Vantage Titan 3TToshiba Medical Systems Corporation, Japan20111
−8K113093Vantage Titan 3TToshibamedical Systems Corporation20121
−7K120487Vantage Titan 3TToshibamedical Systems Corporation2012—
−6K132160Vantage Titan 3TToshibamedical Systems Corporation2013—
−5K143008Vantage Titan 3TToshibamedical Systems Corporation2015—
−4K152371Vantage Titan 3TToshibamedical Systems Corporation20151
−3K162183Vantage Galan 3T, MRT-3020, V4.0Toshibamedical Systems Corporation2016—
−2K173382Vantage Galan 3T, V4.6Toshibamedical Systems Corporation2018—
−1K181593Vantage Galan 3T, MRT-3020/A7, V5.0Canon Medical Systems Corporation2018—
0 (this device)K192574Vantage Galan 3T, MRT-3020, V6.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation2020—
+1K203553Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation20211
+2K212056Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation2021—
+3K220192Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation2022—
+4K222387Vantage Galan 3T, MRT-3020, V8.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation2022—
+5K230355Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation20232

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202140
202220
202310
202410
202510
FIG. 1 — Lineage of K192574, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K181593Vantage Galan 3T, MRT-3020/A7, V5.0
Summary page 6
…gnetic Resonance Diagnostic Device) 13. PRODUCT CODE 90-LNH 14. PREDICATE DEVICE Predicate Device (system): Vantage Galan 3T, MRT-3020/A7, V5.0 (K181593) Subject Predicate Device Vantage Galan 3T,…
Canon Medical Systems Corporation2018—

Also cleared against K181593: 2 other devices →

What descends from it? (9)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203553Vantage Galan 3T, MRT-3020, V7.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation20216
K203323Vantage Galan 3T, MRT-3020, V6.0 with AiCE Reconstruction Processing Unit for MRCanon Medical Systems Corporation20211

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 5 companies; a new generation was cleared every 1.2 years on median.
  • 9 of the 15 ancestors have a recall of any class.
  • Of its descendants, K203553 (Canon Medical Systems Corporation, 2021) has gone furthest, with 6 descendants of its own.
  • 2 other devices also cite its predicate K181593.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The…

Read the full statement (118 words)

Vantage Galan 3T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.

MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:

*Proton density (PD) (also called hydrogen density)

*Spin-lattice relaxation time (T1)

*Spin-spin relaxation time (T2)

*Flow dynamics

*Chemical Shift

Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K192574: Vantage Galan 3T, MRT-3020, V6.0 with AiCE Reconstruction Processing Unit for MR.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K192574

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026