510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-i

K192832510(k) TraditionalCleared Feb 21, 2020142-day reviewClass II21 CFR 892.1750Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Canon Medical Systems Corporation · JAK System, X-Ray, Tomography, Computed

Cleared on Feb 21, 2020 as substantially equivalent to K172188 (Toshiba Medical Systems Coroporation, 2017) after a 142-day review. Its family tree reaches back 6 generations to 2010. 2 later devices descend from it. It has been recalled once.

Decided
Feb 21, 2020
Review
142 days
Class
Class II
Product code
JAK
Recalls
1
Predicates
1

Its longest line of descent runs back 6 generations to K093891 (2010). 9 ancestors in all. 2 later devices descend from it, 1 citing it directly.

  1. Root · 6 generations backK093891 Aquilion One Version 4.6 W/Dual Energy Scan System, Model TSX-301A/2Toshibamedical Systems Corporation · 2010 · Recalled
  2. 5 generations backK121553 Aquilion RXLToshibamedical Systems Corporation · 2012 · Recalled
  3. 4 generations backK151833 Aquilion Lightning (TSX-035A/2, V7.0)Toshibamedical Systems Corporation · 2015 · Recalled
  4. 3 generations backK153263 Aquilion Lightning, TSX-035A/4 and /5, V7.0Toshibamedical Systems Corporation · 2016
  5. 2 generations backK170019 Aquilion Lightning, TSX-036A/1, V8.4Toshibamedical Systems Corporation · 2017 · Recalled
  6. 1 generation backK172188 Aquilion Prime SP, TSX-303B/1, v8.4Toshiba Medical Systems Coroporation · 2017 · Recalled
  7. This deviceK192832 Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-iCanon Medical Systems Corporation · 2020
  8. Descendant · cites it directly · 1 of its ownK203042 Aquilion Exceed LB (TSX-202A/3) V10.6 with AiCE-ICanon Medical Systems Corporation · 2020
  9. Descendant · 2 generations on · 0 of its ownK211828 Aquilion Exceed LBCanon Medical Systems Corporation · 2021

Most influential descendants

Showing 2 of 2 · See all
  • K203042Aquilion Exceed LB (TSX-202A/3) V10.6 with AiCE-ICanon Medical Systems Corporation · 2020 · cites it directly1descendant
  • K211828Aquilion Exceed LBCanon Medical Systems Corporation · 2021 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K093891Aquilion One Version 4.6 W/Dual Energy Scan System, Model TSX-301A/2Toshibamedical Systems Corporation20101
−5K121553Aquilion RXLToshibamedical Systems Corporation20121
−4K151833Aquilion Lightning (TSX-035A/2, V7.0)Toshibamedical Systems Corporation20151
−3K153263Aquilion Lightning, TSX-035A/4 and /5, V7.0Toshibamedical Systems Corporation2016—
−2K170019Aquilion Lightning, TSX-036A/1, V8.4Toshibamedical Systems Corporation20171
−1K172188Aquilion Prime SP, TSX-303B/1, v8.4Toshiba Medical Systems Coroporation20171
0 (this device)K192832Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-iCanon Medical Systems Corporation2020—
+1K203042Aquilion Exceed LB (TSX-202A/3) V10.6 with AiCE-ICanon Medical Systems Corporation2020—
+2K211828Aquilion Exceed LBCanon Medical Systems Corporation2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201210
201310
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202110
FIG. 1 — Lineage of K192832, 2010–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K172188Aquilion Prime SP, TSX-303B/1, v8.4
Summary page 5
…TSX‐303B/1, v8.4 21 CFR System, X‐ray, Medical Tomography K172188 October 6, 2017 (Primary Predicate 892.1750 Tomography, Sys…
Toshiba Medical Systems Coroporation20171
†K183046Aquilion ONE (TSX-305A/6) V8.9 with AiCECanon Medical Systems Corporation2019—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K203042Aquilion Exceed LB (TSX-202A/3) V10.6 with AiCE-ICanon Medical Systems Corporation20201

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 9, 2020Class IIZ-0549-2021Software designA software problem has been identified which could result in the CT Scanner not proceeding to the next actual scan even though automatic start of the next scan is specified. This could result in the CT scan failing resulting in rescanning…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 1.7 years on median.
  • 7 of the 9 ancestors have a recall of any class.
  • Of its descendants, K203042 (Canon Medical Systems Corporation, 2020) has gone furthest, with 1 descendant of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.

The Aquilion Prime SP has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

AiCE (Advanced Intelligent Clear-IQ Engine) is a noise reduction algorithm that improves image quality and reduces image noise by employing Deep Convolutional Neural Network methods for abdomen, pelvis, lung, cardiac, extremities, head and inner ear applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K192832: Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-i.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K192832

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026