510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Aquilion Prime SP, TSX-303B/1, v8.4

K172188510(k) TraditionalCleared Oct 6, 201778-day reviewClass II21 CFR 892.1750Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshiba Medical Systems Coroporation · JAK System, X-Ray, Tomography, Computed

Cleared on Oct 6, 2017 as substantially equivalent to K141741 (Toshibamedical Systems Corporation, 2014) and 1 other predicate after a 78-day review. Its family tree reaches back 5 generations to 2010. 3 later devices descend from it. It has been recalled once.

Decided
Oct 6, 2017
Review
78 days
Class
Class II
Product code
JAK
Recalls
1
Predicates
2

Its longest line of descent runs back 5 generations to K093891 (2010). 8 ancestors in all. 3 later devices descend from it, 1 citing it directly.

  1. Root · 5 generations backK093891 Aquilion One Version 4.6 W/Dual Energy Scan System, Model TSX-301A/2Toshibamedical Systems Corporation · 2010 · Recalled
  2. 4 generations backK121553 Aquilion RXLToshibamedical Systems Corporation · 2012 · Recalled
  3. 3 generations backK151833 Aquilion Lightning (TSX-035A/2, V7.0)Toshibamedical Systems Corporation · 2015 · Recalled
  4. 2 generations backK153263 Aquilion Lightning, TSX-035A/4 and /5, V7.0Toshibamedical Systems Corporation · 2016
  5. 1 generation backK170019 Aquilion Lightning, TSX-036A/1, V8.4Toshibamedical Systems Corporation · 2017 · Recalled
  6. This deviceK172188 Aquilion Prime SP, TSX-303B/1, v8.4Toshiba Medical Systems Coroporation · 2017
  7. Descendant · cites it directly · 2 of its ownK192832 Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-iCanon Medical Systems Corporation · 2020 · Recalled
  8. Descendant · 2 generations on · 1 of its ownK203042 Aquilion Exceed LB (TSX-202A/3) V10.6 with AiCE-ICanon Medical Systems Corporation · 2020
  9. Descendant · 3 generations on · 0 of its ownK211828 Aquilion Exceed LBCanon Medical Systems Corporation · 2021

Most influential descendants

Showing 3 of 3 · See all
  • K192832Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-iCanon Medical Systems Corporation · 2020 · cites it directly2descendants
  • K203042Aquilion Exceed LB (TSX-202A/3) V10.6 with AiCE-ICanon Medical Systems Corporation · 2020 · 2 generations on1descendant
  • K211828Aquilion Exceed LBCanon Medical Systems Corporation · 2021 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K093891Aquilion One Version 4.6 W/Dual Energy Scan System, Model TSX-301A/2Toshibamedical Systems Corporation20101
−4K121553Aquilion RXLToshibamedical Systems Corporation20121
−3K151833Aquilion Lightning (TSX-035A/2, V7.0)Toshibamedical Systems Corporation20151
−2K153263Aquilion Lightning, TSX-035A/4 and /5, V7.0Toshibamedical Systems Corporation2016—
−1K170019Aquilion Lightning, TSX-036A/1, V8.4Toshibamedical Systems Corporation20171
0 (this device)K172188Aquilion Prime SP, TSX-303B/1, v8.4Toshiba Medical Systems Coroporation2017—
+1K192832Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-iCanon Medical Systems Corporation20201
+2K203042Aquilion Exceed LB (TSX-202A/3) V10.6 with AiCE-ICanon Medical Systems Corporation2020—
+3K211828Aquilion Exceed LBCanon Medical Systems Corporation2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201210
201310
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202020
202110
FIG. 1 — Lineage of K172188, 2010–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141741Aquilion Prime, V6.00
Summary page 5
…ber Clearance Date Aquilion PRIME, Toshiba America TSX‐303A/A, /B and /F, v6.00 K141741 November 11, 2014 Medical Systems (Primary Predicate Device) Aq…
Toshibamedical Systems Corporation20142
*K170019Aquilion Lightning, TSX-036A/1, V8.4
Summary page 5
…dical Systems (Primary Predicate Device) Aquilion Lightning SP Toshiba America K170019 February 2, 2017 TSX‐036A/1, v8.4 Medical Systems 12. REASON FOR SUBMISSION: Modificat…
Toshibamedical Systems Corporation20171
†K142465Aquilion ONE Vision, v7.0Toshibamedical Systems Corporation20151
†K133324Suresubtraction LungToshiba Medical Systems Corporation, Japan2014—
†K170177Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1Toshibamedical Systems Corporation20171

Also cleared against K141741: 3 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K192832Aquilion Prime SP (TSX-303B8) V10.2 with AiCE-iCanon Medical Systems Corporation20202

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 25, 2019Class IIZ-1724-2020Software designScanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 10 months on median.
  • 6 of the 8 ancestors have a recall of any class.
  • Of its descendants, K192832 (Canon Medical Systems Corporation, 2020) has gone furthest, with 2 descendants of its own.
  • 3 other devices also cite its predicate K141741.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.

The Aquilion Prime SP has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K172188: Aquilion Prime SP, TSX-303B/1, v8.4.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K172188

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026