510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

epoc Blood Analysis System

K200107510(k) SpecialCleared Jun 5, 2020140-day reviewClass II21 CFR 862.1120Recalls 19

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Epocal, Inc. · CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph

Cleared on Jun 5, 2020 as substantially equivalent to K061597 (Epocal, Inc., 2006) and 4 other predicates after a 140-day review. Its family tree reaches back 3 generations to 2000. It has been recalled 19 times.

Decided
Jun 5, 2020
Review
140 days
Class
Class II
Product code
CHL
Recalls
19
Predicates
5

Its longest line of descent runs back 3 generations to K003261 (2000). 9 ancestors in all, the oldest K001387 (2000). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2018: 1 other ancestor2010: 2 other ancestors2006: 2 other ancestors2000: 1 other ancestorK003261 · Creatinine Plus · Roche Diagnostics Corp. · 2000 · recalledK024098 · Cobas Integra Creatinine Plus Ver.2 Assay · Roche Diagnostics Corp. · 2002 · recalledK113726 · Epoc Chloride Test and Epoc Creatinine Test · Epocal, Inc. · 201220002005201020152020
  1. Root · 3 generations backK003261 Creatinine PlusRoche Diagnostics Corp. · 2000 · Recalled
  2. 2 generations backK024098 Cobas Integra Creatinine Plus Ver.2 AssayRoche Diagnostics Corp. · 2002 · Recalled
  3. 1 generation backK113726 Epoc Chloride Test and Epoc Creatinine TestEpocal, Inc. · 2012
  4. This deviceK200107 epoc Blood Analysis SystemEpocal, Inc. · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K003261Creatinine PlusRoche Diagnostics Corp.20001
−2K024098Cobas Integra Creatinine Plus Ver.2 AssayRoche Diagnostics Corp.20022
−1K113726Epoc Chloride Test and Epoc Creatinine TestEpocal, Inc.2012—
0 (this device)K200107epoc Blood Analysis SystemEpocal, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200620
201020
201810
FIG. 1 — Lineage of K200107, 2000–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CHL? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CHL device

  1. #1 K252207 ABL90 FLEX PLUS System, safeCLINITUBESscore 14.2 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061597Epoc Blood Analysis System
Summary page 6
…e (Lact), Creatinine (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2) in arterial, venous, and capillary blood samples per k061597, k090109, k092849, k093297, k113726 and k171247. This Special 510(k) submission has been prepared to introduce hardw…
Epocal, Inc.20062
*K092849Blood Collection Tube Samples for Use with Point of Care Blood Analyzer
Summary page 6
…ne (Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2) in arterial, venous, and capillary blood samples per k061597, k090109, k092849, k093297, k113726 and k171247. This Special 510(k) submission has been prepared to introduce hardware and software…
Epocal, Inc.2010—
*K093297Epoc Bgem, Epoc Reader, Epoc Host, Epoc Care-Fill Capillary Tube
Summary page 6
…(Crea), Hematocrit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2) in arterial, venous, and capillary blood samples per k061597, k090109, k092849, k093297, k113726 and k171247. This Special 510(k) submission has been prepared to introduce hardware and software upda…
Epocal, Inc.20101
*K113726Epoc Chloride Test and Epoc Creatinine Test
Summary page 6
…rit (Hct), Blood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2) in arterial, venous, and capillary blood samples per k061597, k090109, k092849, k093297, k113726 and k171247. This Special 510(k) submission has been prepared to introduce hardware and software updates to the curr…
Epocal, Inc.2012—
*K171247epoc Blood Urea Nitrogen Test, epoc Total Carbon Dioxide Test
Summary page 6
…lood Urea Nitrogen (BUN) and Total Carbon Dioxide (TCO2) in arterial, venous, and capillary blood samples per k061597, k090109, k092849, k093297, k113726 and k171247. This Special 510(k) submission has been prepared to introduce hardware and software updates to the current epoc® Ho…
Epocal, Inc.2018—

Also cleared against K061597: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 2, 2026Class IIZ-2005-2026Under Investigation by firmUse of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading…
Jul 7, 2025Class IIZ-2272-2025Under Investigation by firmSiemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/L. The maximum bias…
Mar 19, 2025Class IIZ-1684-2025Process controlThe occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Mar 19, 2025Class IIZ-1685-2025Process controlThe occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Nov 22, 2024Class IIZ-0741-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Nov 22, 2024Class IIZ-0734-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Nov 22, 2024Class IIZ-0736-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Nov 22, 2024Class IIZ-0735-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Nov 22, 2024Class IIZ-0737-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Nov 22, 2024Class IIZ-0740-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Nov 22, 2024Class IIZ-0738-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Nov 22, 2024Class IIZ-0739-2025Software designSiemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement…
Sep 24, 2024Class IIZ-0286-2025Under Investigation by firmSiemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10…
Oct 19, 2023Class IIZ-0601-2024Software designSiemens Healthcare Diagnostics Inc. has confirmed a potential issue that could occur under certain conditions, a software problem is causing erroneous results to be printed. When the results are run on the epoc NXS Host and Delete Blood…
Jun 1, 2023Class IIZ-2476-2023Under Investigation by firmThere is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Jun 1, 2023Class IIZ-2477-2023Under Investigation by firmThere is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Jun 1, 2023Class IIZ-2478-2023Under Investigation by firmThere is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Dec 12, 2022Class IIZ-1215-2023Under Investigation by firmpO2 High Recovery in Quality Control When Adjusted for Barometric Pressure
Aug 28, 2020Class IIZ-2970-2020Process controlSporadically inconsistent discrepant (low bias) glucose results on card lot 01-20095-10. Potential exists for a delay in diagnosis of hyperglycemia or unnecessary treatment for hypoglycemia.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 2 companies; a new generation was cleared every 7.7 years on median.
  • 6 of the 9 ancestors have a recall of any class.
  • 1 other device also cites its predicate K061597.

How does its category compare?

All CHL devices

CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · CHL

1in 2025

2007: 12008: 12010: 12011: 42012: 42013: 52014: 32015: 42016: 32017: 12018: 12020: 42021: 22023: 52024: 12025: 12026 (partial year): 1
2007high 5 (2013) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CHL

267in 2023

2011: 108 days median2012: 65 days median2013: 152 days median2014: 68 days median2015: 155 days median2016: 306 days median2020: 100 days median2023: 267 days median
2007high 306 (2016) · low 65 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CHL

13so far in 2026, the most since 2007

2007: 62008: 52009: 62010: 72011: 12012: 92013: 12014: 12015: 32016: 12017: 12018: 32020: 22023: 42024: 92026 (partial year): 13
2007high 13 (2026) · low 0 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CHL

210in 2025

2007: 162008: 402009: 742010: 202011: 502012: 642013: 622014: 932015: 552016: 1182017: 2062018: 1742019: 1412020: 1172021: 1472022: 1382023: 1772024: 1962025: 210
2007high 210 (2025) · low 16 (2007)2025

MAUDE reports naming product code CHL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—616
20081—540
20090—674
20101—720
20114108150
2012465964
20135152162
2014368193
20154155355
201633061118
20171—1206
20181—3174
20190—0141
202041002117
20212—0147
20220—0138
202352674177
20241—9196
20251—0210
20261—13—

What is it indicated for?

The epoc Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous, or capillary whole blood in the laboratory or at the point of care. The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, pO2, Sodium, Potassium, Ionized Calcium, Chloride, Total…

Read the full statement (330 words)

The epoc Blood Analysis System is intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous, or capillary whole blood in the laboratory or at the point of care.

The Blood Gas Electrolyte and Metabolite (BGEM) Test Card panel configuration includes sensors that quantitate pH, pCO2, pO2, Sodium, Potassium, Ionized Calcium, Chloride, Total Carbon Dioxide, Glucose, Lactate, Blood Urea Nitrogen, Creatinine, and Hematocrit.

pH, pCO2, pO2 (blood gases) measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Sodium and Potassium measurements from the epoc Blood Analysis System are used in diagnosis and treatment of diseases involving electrolyte imbalance.

Ionized Calcium measurements from the epoc Blood Analysis System are used in diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.

Chloride measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of electrolyte and metabolic disorders.

Total Carbon Dioxide measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of disorders associated with changes in body acid-base balance.

Glucose measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of carbohydrate metabolism disorders, including diabetes mellitus and idiopathic hypoglycemia, and of pancreatic islet cell tumors.

Lactate measurements from the epoc Blood Analysis System are used to evaluate the acid-base status and are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).

Blood Urea Nitrogen measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal and metabolic diseases.

Creatinine measurements from the epoc Blood Analysis System are used in the diagnosis and treatment of certain renal diseases and in monitoring renal dialysis.

Hematocrit measurements from the epoc Blood Analysis System are used to distinguish normal from abnormal states of blood volume, such as anemia and erythrocytosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K200107: epoc Blood Analysis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K200107

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026