510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Ohmeda Medical Giraffe Incubator

K020547510(k) TraditionalCleared May 9, 200279-day reviewClass II21 CFR 880.5400Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ohmeda Medical · FMZ Incubator, Neonatal

Cleared on May 9, 2002 as substantially equivalent to K010222 (Ohmeda Medical, 2001) and 2 other predicates after a 79-day review. Its family tree reaches back 1 generation to 1997. 2 of the 9 devices in its family, counting itself, have a Class I recall. 5 later devices descend from it. It has been recalled 4 times.

Decided
May 9, 2002
Review
79 days
Class
Class II
Product code
FMZ
Recalls
4
Predicates
3

Its longest line of descent runs back 1 generation to K954204 (1997). 3 ancestors in all. 5 later devices descend from it, 2 citing it directly. 2 of the 9 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2000: 1 other ancestor2015: 1 descendant cleared, 1 with a Class I recall12007: 1 descendant cleared2022: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared2026: 1 descendant clearedK954204 · Drager Incubator 8000 IC · Drager, Inc. · 19971995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 1 descendant, 1 with a Class I recall13 generations forward: 1 descendant4 generations forward: 1 descendant K954204 · Drager Incubator 8000 IC · Drager, Inc. · 19970
  1. Root · 1 generation backK954204 Drager Incubator 8000 ICDrager, Inc. · 1997
  2. This deviceK020547 Ohmeda Medical Giraffe IncubatorOhmeda Medical · 2002
  3. Descendant · cites it directly · 4 of its ownK072512 Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda…Datex-Ohmeda, Inc. · 2007
  4. Descendant · cites it directly · 3 of its ownK152809 Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Recalled (Class I)
  5. Descendant · 2 generations on · 2 of its ownK213551 Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc. · 2022 · Recalled (Class I)
  6. Descendant · 3 generations on · 1 of its ownK251663 Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025
  7. Descendant · 4 generations on · 0 of its ownK260326 Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation…Datex-Ohmeda, Inc. · 2026

Class I recalls in this family

Most influential descendants

Showing 5 of 5 · See all
  • K072512Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda Uninterruptible…Datex-Ohmeda, Inc. · 2007 · cites it directly4descendants
  • K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · cites it directly3descendants
  • K213551Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on2descendants
  • K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025 · 3 generations on1descendant
  • K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K954204Drager Incubator 8000 ICDrager, Inc.1997—
0 (this device)K020547Ohmeda Medical Giraffe IncubatorOhmeda Medical2002—
+1K072512Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda…Datex-Ohmeda, Inc.2007—
+1K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc.2015Class I
+2K213551Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc.2022Class I
+3K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc.2025—
+4K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
201511
202211
202510
202610
FIG. 1 — Lineage of K020547, 1997–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMZ device

  1. #1 K251663 Giraffe OmniBed Carestation (CS1); Giraffe…score 14.4 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K010222Ohmeda Medical Giraffe Incubator
Summary page 1
…nformation The Giraffe Incubator is substantially equivalent to the following Class II devices: Device Last 510(k) , Number Ohmeda Medical Giraffe Incubator (original) K010222 Drager Incubator 8000 IC K954204 Isolette Infant Incubator Model C2HS K001242 Product Description The Ohmeda Medical Giraffe In…
Ohmeda Medical20013
*K954204Drager Incubator 8000 IC
Summary page 1
…is substantially equivalent to the following Class II devices: Device Last 510(k) , Number Ohmeda Medical Giraffe Incubator (original) K010222 Drager Incubator 8000 IC K954204 Isolette Infant Incubator Model C2HS K001242 Product Description The Ohmeda Medical Giraffe Incubator is an infant bed which pr…
Drager, Inc.1997—
*K001242Isolette Infant Incubator, Model C2HS
Summary page 1
…Class II devices: Device Last 510(k) , Number Ohmeda Medical Giraffe Incubator (original) K010222 Drager Incubator 8000 IC K954204 Isolette Infant Incubator Model C2HS K001242 Product Description The Ohmeda Medical Giraffe Incubator is an infant bed which provides thermal support for infants who are un…
Hill-Rom Air-Shields2000—

Also cleared against K010222: 2 other devices →

What descends from it? (5)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072512Modification to: Ohmeda Medical Giraffe Incubator (Giraffe and Panda Uninterruptible…Datex-Ohmeda, Inc.20074
K152809Giraffe Incubator Carestation CS1 Class IOhmeda Medical, A Division of Datex-Ohmeda, Inc.20153

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 28, 2020Class IIZ-2236-2020Device DesignPotential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
Nov 18, 2016Class IIZ-1505-2017Under Investigation by firmCertain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where…
Apr 25, 2005Not classifiedZ-1047-05OtherNeonatal incubators and beds may unintentionally move from a safe position following an interruption in the power supply.
Jun 13, 2003Not classifiedZ-1064-03OtherFirm's received complaints of medical device malfunction affecting instrument's calibration and alarm systems.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152809Giraffe Incubator Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls
  • K213551Giraffe Omnibed Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K072512 (Datex-Ohmeda, Inc., 2007) has gone furthest, with 4 descendants of its own.
  • 2 other devices also cite its predicate K010222.

How does its category compare?

All FMZ devices

FMZ Incubator, Neonatal: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FMZ

2in 2025

2007: 32008: 12010: 42011: 12014: 12015: 12016: 12017: 12018: 22019: 22022: 12025: 22026 (partial year): 2
2007high 4 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMZ

60in 2010

2007: 81 days median2010: 60 days median
2007high 81 (2007) · low 60 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMZ

2in 2024

2008: 42011: 52013: 1, 1 Class I2016: 12018: 32019: 12023: 8, 8 Class I2024: 2, 2 Class I2026 (partial year): 3
2007high 8 (2023) · low 0 (2007)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · FMZ

250in 2025

2007: 142008: 102009: 82010: 112011: 102012: 122013: 112014: 32015: 92016: 122017: 42018: 182019: 242020: 82021: 152022: 272023: 192024: 1182025: 250
2007high 250 (2025) · low 3 (2014)2025

MAUDE reports naming product code FMZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007381014
20081—410
20090—08
2010460011
20111—510
20120—012
20130—1 (1)11
20141—03
20151—09
20161—112
20171—04
20182—318
20192—124
20200—08
20210—015
20221—027
20230—8 (8)19
20240—2 (2)118
20252—0250
20262—3—

What is it indicated for?

The Giraffe Incubator is an infant incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature controlled environment for the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K020547: Ohmeda Medical Giraffe Incubator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K020547

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026