510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

N Series Patient Monitors

K213799510(k) TraditionalCleared Jul 12, 2022218-day reviewClass II21 CFR 870.1025Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Jul 12, 2022 as substantially equivalent to K202405 (Shenzhen Mindray Bio-Medical Electronics Co., Ltd., 2021) after a 218-day review. Its family tree reaches back 3 generations to 2018. 2 later devices descend from it. It has been recalled 2 times.

Decided
Jul 12, 2022
Review
218 days
Class
Class II
Product code
MHX
Recalls
2
Predicates
1

Its longest line of descent runs back 3 generations to K182075 (2018). 2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2025: 2 descendants clearedReference device: K161531 · IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement Module X2; IntelliVue Patient Monitors MP20, MP30, MP40, MP50, MP60; IntelliVue Patient Monitors MP70, MP80, MP90;IntelliVue Patient Monitors MX400, MX430, MX450, MX500, MX550, MX600;IntelliVue Patient Monitors MX700, MX800 · Philips Medizin Systeme Boeblingen GmbH · 2016 · recalledReference device: K131414 · Carescape Monitor B850 · GE Healthcare Finland Oy · 2013 · recalledReference device: K180881 · HemoSphere Advanced Monitor, HemoSphere Pressure Cable · Edwards Lifesciences, LLC · 2018Reference device: K210906 · IntelliVue Patient Monitor MX750, IntelliVue Patient Monitor MX850, IntelliVue 4-Slot module rack FMX-4 · Philips Medizin Systeme Boeblingen GmbH · 2021 · recalledReference device: K193391 · BeneVision Central Monitoring System · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2020 · recalledReference device: K182979 · IntelliVue Patient Monitors MX100, MX400, MX430, MX450, MX500, MX550, IntelliVue Patient Monitors MX700, MX800, IntelliVue Patient Monitors MP5, MP5SC, Multi-Measurement Module X3 · Philips Medizin Systeme Boeblingen GmbH · 2019K182075 · BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19. BeneVision N22, BeneVision N1) · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2018K192972 · Patient Monitor · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2020K202405 · BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2021201620182020202220242026
  1. Root · 3 generations backK182075 BeneVision N Series Patient Monitors (Including BeneVision N12…Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2018
  2. 2 generations backK192972 Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2020
  3. 1 generation backK202405 BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19…Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2021
  4. This deviceK213799 N Series Patient MonitorsShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2022
  5. Descendant · cites it directly · 0 of its ownK250854 Multi-parameter Patient Monitor (N10); Multi-parameter Patient…Shenzhen Comen Medical Instruments Co.,Ltd · 2025
  6. Descendant · cites it directly · 0 of its ownK243767 SV600, SV800 VentilatorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2025

Most influential descendants

Showing 2 of 2 · See all
  • K250854Multi-parameter Patient Monitor (N10); Multi-parameter Patient Monitor (N12)…Shenzhen Comen Medical Instruments Co.,Ltd · 2025 · cites it directly0descendants
  • K243767SV600, SV800 VentilatorShenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K182075BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision…Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2018—
−2K192972Patient MonitorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2020—
−1K202405BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision…Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2021—
0 (this device)K213799N Series Patient MonitorsShenzhen Mindray Bio-Medical Electronics Co., Ltd.2022—
+1K250854Multi-parameter Patient Monitor (N10); Multi-parameter Patient Monitor (N12)…Shenzhen Comen Medical Instruments Co.,Ltd2025—
+1K243767SV600, SV800 VentilatorShenzhen Mindray Bio-Medical Electronics Co., Ltd.2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202520
FIG. 1 — Lineage of K213799, 2018–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K202405BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22…
Summary page 7
…or electroencephalograph 882.1320, II GXY Cutaneous electrode. Electrode, cutaneous 3. PREDICATE DEVICES • Predicate: K202405 – BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVi…
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2021—
†K161531IntelliVue Patient Monitors MP2, MP5, MP5T, MP5SC, and Multi-Measurement Module X2…Philips Medizin Systeme Boeblingen GmbH20161
†K131414Carescape Monitor B850GE Healthcare Finland Oy20133
†K180881HemoSphere Advanced Monitor, HemoSphere Pressure CableEdwards Lifesciences, LLC2018—
†K210906IntelliVue Patient Monitor MX750, IntelliVue Patient Monitor MX850, IntelliVue 4-Slot…Philips Medizin Systeme Boeblingen GmbH20212
†K193391BeneVision Central Monitoring SystemShenzhen Mindray Bio-Medical Electronics Co., Ltd.20202
†K182979IntelliVue Patient Monitors MX100, MX400, MX430, MX450, MX500, MX550, IntelliVue Patient…Philips Medizin Systeme Boeblingen GmbH2019—

Also cleared against K202405: 1 other device →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K250854Multi-parameter Patient Monitor (N10); Multi-parameter Patient Monitor (N12)…Shenzhen Comen Medical Instruments Co.,Ltd20250
K243767SV600, SV800 VentilatorShenzhen Mindray Bio-Medical Electronics Co., Ltd.20250

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 31, 2026Class IIIZ-0110-2027Labeling designCertain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Aug 15, 2025Class IIZ-2622-2025Software designPotential for activation of an abnormal alarm pause.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 1 company; a new generation was cleared every 1.4 years on median.
  • None of the 3 ancestors has a recall.
  • Of its descendants, K250854 (Shenzhen Comen Medical Instruments Co.,Ltd, 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K202405.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive…

Read the full statement (437 words)

The BeneVision N12/N15/N17/N19/N22 patient monitors are intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration Rate (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Cardiac Output (C.O.), Continuous Cardiac Output (CCO), Mixed/Central Venous Oxygen Saturation (SvO2/ScvO2), Carbon Dioxide (CO2), Oxygen (O2), Anesthetic Gas (AG), Impedance Cardiograph (ICG), Bispectral Index (BIS), Respiration Mechanics (RM), Neuromuscular Transmission Monitoring (NMT), Electroencephalograph (EEG), and Regional Oxygen Saturation (rSO2). The system also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

* BIS, RM, CCO, SvO2/ScvO2, PAWP, NMT monitoring, PNP, and PNC are intended for adult and pediatric patients only. CCO using FloTrac is intended for adult patients only;

* C.O. monitoring and A-Fib are intended for adult patients only;

* ICG monitoring is intended for only adult patients who meet the following requirements: height: 122 to 229cm, weight: 30 to 155kg.

* rSO2 monitoring is intended for use in individuals greater than 2.5kg.

The monitors are to be used in healthcare facilities by clinical professionals or under their guidance. They should only be used by persons who have received adequate training in their use. The BeneVision N12/N15/N17/N19/N22 monitors are not intended for helicopter transport, hospital ambulance, or home use.

The BeneVision N1 Patient Monitor is intended for monitoring, displaying, reviewing, storing, alarming, and transferring of multiple physiological parameters including ECG (3-lead, 5-lead, 6-lead or 12-lead selectable, Arrhythmia Detection, ST Segment Analysis, QT Analysis, and Heart Rate (HR)), Respiration (Resp), Temperature (Temp), Pulse Oxygen Saturation (SpO2), Pulse Rate (PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Pulmonary Artery Wedge Pressure (PAWP), Carbon Dioxide (CO2) and Oxygen (O2). The system also provides an interpretation of resting 12-lead ECG.

All the parameters can be monitored on single adult, pediatric, and neonatal patients except for the following:

• PAWP, PNP, and PNC are intended for adult and pediatric patients only;

• A-Fib is intended for adult patients only;

The BeneVision N1 monitor is to be used in healthcare facilities. It can also be used during patient transport inside and outside of the hospital environment, whereas N1 configured with WMTS technology can be used inside the hospital only. It should be used by clinical professionals or under their guidance. It should only be used by persons who have received adequate training in its use. It is not intended for home use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K213799: N Series Patient Monitors.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K213799

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.