Fresenius Medical Care Renal Therapies Group, LLC · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Cleared on May 30, 2025 as substantially equivalent to K231534 (Fresenius Medical Care, 2024) and 2 other predicates after a 199-day review. Its family tree reaches back 13 generations to 1994. 4 of the 60 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 6 times.
- Decided
- May 30, 2025
- Review
- 199 days
- Class
- Class II
- Product code
- KDI
- Recalls
- 6
- Predicates
- 3
Its longest line of descent runs back 13 generations to K921456 (1994). 58 ancestors in all, the oldest K770691 (1977). 1 later device descends from it, 1 citing it directly. 4 of the 60 devices in this family, counting this one, have a Class I recall.
- Root · 13 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
- 9 generations between
- 3 generations backK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
- 2 generations backK222952 2008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · Recalled (Class I)
- 1 generation backK231534 5008X Hemodialysis SystemFresenius Medical Care · 2024 · Recalled
- This deviceK243505 5008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC · 2025
- Descendant · cites it directly · 0 of its ownK251851 DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC · 2025
Show all 13 generations
- Root · 13 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
- 12 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
- 11 generations backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
- 10 generations backK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010 · Recalled
- 9 generations backK101715 Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente · 2010 · Recalled
- 8 generations backK120017 Fresenius 2008 Hemodiaysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
- 7 generations backK121341 Fresenius 2008T Hemodialysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
- 6 generations backK124035 2008K@Home Hemodialysis MachineFresenius Medical Care North America · 2013
- 5 generations backK150708 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Recalled (Class I)
- 4 generations backK153449 2008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · Recalled
- 3 generations backK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
- 2 generations backK222952 2008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · Recalled (Class I)
- 1 generation backK231534 5008X Hemodialysis SystemFresenius Medical Care · 2024 · Recalled
- This deviceK243505 5008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC · 2025
Class I recalls in this family
Most influential descendants
Showing 1 of 1 · See all- K251851DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC · 2025 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −13 | K921456 | Fresenius 3008/FDS-08 Dialysate Delivery System | Fresenius USA, Inc. | 1994 | 3 |
| −12 | K994267 | Fresenius 2008K | Fresenius Medical Care | 2000 | 5 |
| −11 | K080964 | Fresenius Model 2008T Hemodialysis Machine | Fresenius Medical Care North America | 2009 | 4 |
| −10 | K093902 | Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573 | Fresenius Medical Care North America | 2010 | 2 |
| −9 | K101715 | Fresenius 2008T Hemodialysis Machine Model 190573 | Fresenius Medical Care North America, Design Cente | 2010 | 1 |
| −8 | K120017 | Fresenius 2008 Hemodiaysis Machine with Bibag System | Fresenius Medical Care North America | 2012 | 3 |
| −7 | K121341 | Fresenius 2008T Hemodialysis Machine with Bibag System | Fresenius Medical Care North America | 2012 | 5 |
| −6 | K124035 | 2008K@Home Hemodialysis Machine | Fresenius Medical Care North America | 2013 | — |
| −5 | K150708 | 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T… | Fresenius Medical Care Renal Therapies Group, LLC | 2015 | Class I |
| −4 | K153449 | 2008K2 Hemodialysis Machine | Fresenius Medical Care Renal Therapies Group, LLC | 2016 | 2 |
| −3 | K173972 | 2008T BlueStar Hemodialysis Machine | Fresenius Medical Care Renal Therapies Group, LLC | 2018 | Class I |
| −2 | K222952 | 2008T BlueStar Hemodailysis Machine | Fresenius Medical Care Renal Therapies Group, LLC | 2022 | Class I |
| −1 | K231534 | 5008X Hemodialysis System | Fresenius Medical Care | 2024 | 1 |
| 0 (this device) | K243505 | 5008X Hemodialysis System | Fresenius Medical Care Renal Therapies Group, LLC | 2025 | — |
| +1 | K251851 | DIASAFEplusUS (F00013010) | Fresenius Medical Care Renal Therapies Group, LLC | 2025 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1977 | 1 | 0 |
| 1981 | 1 | 0 |
| 1984 | 1 | 0 |
| 1987 | 2 | 0 |
| 1988 | 2 | 0 |
| 1989 | 2 | 0 |
| 1990 | 1 | 0 |
| 1994 | 2 | 0 |
| 1995 | 3 | 0 |
| 1996 | 1 | 0 |
| 1997 | 3 | 0 |
| 1998 | 1 | 0 |
| 1999 | 2 | 0 |
| 2000 | 2 | 0 |
| 2001 | 2 | 0 |
| 2002 | 3 | 0 |
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2007 | 1 | 0 |
| 2008 | 1 | 1 |
| 2009 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 2 | 0 |
| 2013 | 2 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2017 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2023 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KDI device
- #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K231534 | 5008X Hemodialysis SystemSummary page 6…Marketed Predicate Devices None of the predicate devices have been subject to any design-related recalls. 1.3.1. 5008X Machine Predicate Device The 5008X Machine (K231534) is the legally marketed predicate device. The 2008T BlueStar Hemodialysis Machine (K231125) is the legally marketed secondary pr… | 2024 |
| * | K231125 | 2008T BlueStar Hemodialysis MachineSummary page 6…lated recalls. 1.3.1. 5008X Machine Predicate Device The 5008X Machine (K231534) is the legally marketed predicate device. The 2008T BlueStar Hemodialysis Machine (K231125) is the legally marketed secondary predicate device. 1.3.2. 5008X Bloodlines Predicate Device The 5008X Bloodlines (K231534… | 2023 |
| * | K201207 | CombiSet SMARTECH Hemodialysis Blood Tubing Set with attached Priming Set and Integrated…Summary page 6…1.3.2. 5008X Bloodlines Predicate Device The 5008X Bloodlines (K231534) are the legally marketed predicate devices. The Combiset SMARTECH Hemodialysis Bloodlines (K201207) are the legally marketed secondary predicate devices.… | 2020 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K251851 | DIASAFEplusUS (F00013010) | 2025 | 0 |
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 10, 2026 | Class II | Z-3087-2026 | Component change control | Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia. |
| Aug 10, 2026 | Class II | Z-3086-2026 | Component change control | Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia. |
| Aug 10, 2026 | Class II | Z-3084-2026 | Component change control | Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia. |
| Aug 10, 2026 | Class II | Z-3085-2026 | Component change control | Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia. |
| May 11, 2026 | Class II | Z-2702-2026 | Process control | Potential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem. |
| Oct 25, 2025 | Class II | Z-0936-2026 | Software design | Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled. |
The 5 latest of 6 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · 2 generations back · ◆ on the line of descent in FIG. 18 Class I recalls
- K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · 2 generations back · ◆ on the line of descent in FIG. 18 Class I recalls
- K080807Blood Tubing Sets (Sterile Fluid Path)Nxstage Medical, Inc. · 2008 · 3 generations back · ◆ in the 2008 ancestors bar of FIG. 17 Class I recalls
- K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · 4 generations back · ◆ on the line of descent in FIG. 18 Class I recalls
What does this lineage say?
- The line of descent spans 31 years and passes through 5 companies; a new generation was cleared every 1.3 years on median.
- 23 of the 58 ancestors have a recall of any class; 4 have a Class I recall.
- Of its descendants, K251851 (Fresenius Medical Care Renal Therapies Group, LLC, 2025) has gone furthest, with 0 descendants of its own.
How does its category compare?
All KDI devicesKDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.
FDA decisions per year · KDI
9in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KDI
199in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KDI
2in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · KDI
2,786in 2025
MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 11 | 91 | 4 | 916 |
| 2008 | 6 | 148 | 0 | 1,042 |
| 2009 | 4 | 236 | 3 | 522 |
| 2010 | 9 | 186 | 1 | 823 |
| 2011 | 5 | 85 | 4 | 2,270 |
| 2012 | 6 | 153 | 1 | 1,765 |
| 2013 | 11 | 247 | 2 | 2,360 |
| 2014 | 8 | 164 | 3 | 6,154 |
| 2015 | 5 | 69 | 5 | 4,182 |
| 2016 | 5 | 230 | 4 | 766 |
| 2017 | 8 | 188 | 1 | 963 |
| 2018 | 2 | — | 1 | 1,716 |
| 2019 | 4 | 263 | 6 | 1,587 |
| 2020 | 9 | 266 | 0 | 1,633 |
| 2021 | 2 | — | 2 | 1,944 |
| 2022 | 7 | 259 | 17 | 2,180 |
| 2023 | 7 | 157 | 8 (8) | 2,692 |
| 2024 | 7 | 235 | 4 (2) | 2,441 |
| 2025 | 9 | 199 | 2 | 2,786 |
| 2026 | 6 | 154 | 2 | — |
What is it indicated for?
Machine: The 5008 X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration. The 5008X Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the…
Read the full statement (619 words)
Machine:
The 5008 X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration.
The 5008X Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine.
The bibag System is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag System is intended to be used as one component in the preparation of dialysate according to a physician’s prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module.
The 5008X Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System.
The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
DIASAFEPIlusUS Filter:
The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate and sterile, non- pyrogenic substitution fluid from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).
The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be <200 CFU/mL and <2 EU/mL, respectively.
The DIASAFE®plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE®plusUS.
5008X Tubing Set:
The blood tubing sets are intended to be used only with the Fresenius Medical Care 5008X Hemodialysis Device, which is indicated for intermittent hemodialysis treatment for patients with acute kidney injury or chronic kidney disease in a healthcare facility. Therapy options include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
The HDF blood tubing sets are indicated for use with a prescribed dialyzer. For use with a compatible dialyzer, as per the labeling.
Additional Indications for Use for the blood tubing set equipped with Twister AFRC
The Fresenius Twister Access Flow Reversing Connector (AFRC) is indicated for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices in order to obtain an access flow measurement.
Additional Indications for Use for the blood tubing set equipped with CLiC
The blood tubing set is intended to be used with the Crit-Line Clip Monitor System (CLiC) on the 5008 X Hemodialysis Device.
The CLiC is used to non-invasively measure hematocrit, oxygen saturation, and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (1.e., increases or decreases the rate at which tluid 1s removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
CitraSure Disinfectant:
CitraSure is intended for thermal disinfection of the Fresenius Medical Care 5008X Hemodialysis System when delivered and diluted per the 5008X heat disinfection program.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K243505 · printed Nov 6, 2026
Cite this page
510(k) Report. “K243505: 5008X Hemodialysis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K243505